Outake recallZ-2559-2024 · Merit Medical Systems, Inc. · initiated Jun 11, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA root cause category: No Marketing Application
What the company is doing
On 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that appropriate personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities. 4. Follow the attached instructions for placing the recall labels on the products containing Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled. 5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification. 6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products. 7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers. For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
Recall details
- Recall number
- Z-2559-2024
- Status
- Open, Classified
- Initiated by the company
- Jun 11, 2024
- Posted by the FDA
- Aug 12, 2024
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 3631 kits
- Product code
- ODY
- FDA submissions
- K063048
- Distribution
- U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Product description
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Outake, REF CET200 and CET300, The Catheter Extractor is a catheter accessory device intended for minimally invasive removal of tunneled long term catheters. The Catheter Extractor is indicated for use with tunneled, double lumen hemodialysis catheters with cuff sizes ranging from 10.8 French to 14.5 French.
Code information (models, lots, serial numbers)
Catalog Number: CET200 UDI-DI code: 00884450052587 Lot Numbers: T2651502 T2700122 T2712304 T2748973 T2758209 T2779867 T2833064 T2850758 T2903817 Catalog Number: CET300 UDI-DI code: 00884450045497 T2626011 T2662043 T2701699 T2726893 T2750399 T2758210 T2805965 T2832137 T2849258 T2886083 T2921078
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Outake® 00884450045497Merit Medical Systems, Inc. · Tunneled catheter remover1 recall0 reports
- Outake® 00884450052587Merit Medical Systems, Inc. · Tunneled catheter remover1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94797).
- Z-2557-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Rapid Response Waste Management System Kit, REF BRC-BAL-INT-SL/B …
- Z-2558-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Custom Procedure Kit (Angiography/angioplasty kit), REF: K12T-107…
- Z-2560-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K02 (Custom Syringe Kit), REF: K02-00873A, Inject fluids into, o…
- Z-2561-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF:K04-00791A An intravascular admini…
- Z-2562-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF: K05-00203M K05-00511J K05-00547J K…
- Z-2563-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K08-PI (Custom Fluid Management Kit), REF: K08-02042 K08-02948 K0…
- Z-2564-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K09/K09T (Custom Manifold Kit), REF: K09-00526L K09-00807AH K09-0…
- Z-2565-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K10/K10-PI (Custom Waste Management Kit), REF: K10T-05815 K10T-05…
- Z-2566-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K11 (Custom Pressure Monitoring Kit), REF: K11-00382C To support…
- Z-2567-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K12T/12PI (Custom Pressure Monitoring Kit), REF: K12-03846 K12-04…
- Z-2568-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K22C (Vein Tray Angiography kit), REF: K09T-11796F K12T-01558G K1…
- Z-2569-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: PC (Pericardiocentesis Tray - Fluid drainage tray), REF: PC101/B …
- Z-2570-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: ReSolve (Catheter and Tip Suction kit), REF: RTT14038MB/A The Re…
- Z-2571-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG (Introducer Catheter), REF: FCL-069-00/A FCL-069-0…
- Z-2572-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude Prestige (Introducer Catheter), REF: PLP-2505/A PLP-2514/…
- Z-2573-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Custom Procedure Kit (Angiography/Angioplasty kit), REF: OPT0242-…
- Z-2574-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude SNAP (Introducer Catheter kit), REF: For the introducti…
- Z-2575-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude SNAP-H (Introducer Catheter kit), REF: PLSH-1006/A PLSH-1…
- Z-2576-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Safety Paracentesis Procedure Tray (Paracentesis Tray kit), REF: …
- Z-2577-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG/Worley advanced CSG (Introducer Catheter kit), REF…
- Z-2578-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K09T-0…
- Z-2579-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K05-021…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page