Prelude SNAP recallZ-2574-2024 · Merit Medical Systems, Inc. · initiated Jun 11, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
FDA root cause category: No Marketing Application
What the company is doing
On 06/11/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL NOTICE" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization. In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/. Customer are instructed to: 1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution. 2. Ensure that appropriate personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities. 4. Follow the attached instructions for placing the recall labels on the products containing Jiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled. 5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification. 6.Post this URGENT MEDICAL DEVICE RECALL NOTICE and Jiangsu Shenli Syringe photographs on or near the affected products. 7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers. For questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.
Recall details
- Recall number
- Z-2574-2024
- Status
- Open, Classified
- Initiated by the company
- Jun 11, 2024
- Posted by the FDA
- Aug 12, 2024
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 55305
- Product code
- DYB
- FDA submissions
- K143255
- Distribution
- U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Product description
Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude SNAP (Introducer Catheter kit), REF: For the introduction of various types of pacing leads and catheters to the heart and coronary venous system.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Catalog Number/UDI-DI code/Lot Numbers: PLS-1006/A 00884450467954 I2868407 I2873488 I2884101 I2889284 I2889285 I2889286 I2889287 I2889289 I2889288 I2903937 I2903938 I2903939 I2914399 I2925168 I2925169 PLS-1007/A 00884450467985 I2790541 I2862963 I2868410 I2868412 I2873489 I2882475 PLS-1008/A 00884450467992 I2868415 I2884088 I2889299 I2889300 I2661407S1 I2661408S1 I2925190 PLS-1009.5/A 00884450468036 I2862967 I2873494 I2889301 I2925173 PLS-1009/A 00884450468012 I2790564 I2873496 I2884103 I2889303 I2889306 PLS-1010.5/A 00884450468050 I2862970 PLS-1010/A 00884450468043 I2873497 I2873498 I2873499 I2873500 I2889327 PLS-1011/A 00884450468067 I2889329 PLS-2506/A 00884450468098 I2873501 I2889331 I2914419 PLS-2508/A 00884450468111 I2889334 PLS-2509/A 00884450468128 I2687649S1 I2868417 I2914426 PLS-2510/A 00884450468135 I2873503 I2889336 PLS-2511/A 00884450468159 I2889337 PLSX-1006/A 00884450489550 I2868420 I2873516 I2889355 I2914466 PLSX-1007/A 00884450489567 I2868421 I2873517 I2883775 I2903957 I2932677 PLSX-1008/A 00884450489574 I2889357 PLSX-1009/A 00884450489598 I2914468
Devices named in this recall
17 devices in the FDA device registry whose identifier appears in the recall's code information.
- Prelude SNAP™ 10884450467982Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall19 reports
- Prelude SNAP™ 10884450467951Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall6 reports
- Prelude SNAP™ 10884450467999Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall1 death2 reports
- Prelude SNAP™ 10884450468156Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall2 reports
- Prelude SNAP™ 10884450468019Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468033Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468040Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468057Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468064Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468095Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468118Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468125Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450468132Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450489557Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450489564Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450489571Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
- Prelude SNAP™ 10884450489595Merit Medical Systems, Inc. · INTRODUCER, CATHETER1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94797).
- Z-2557-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Rapid Response Waste Management System Kit, REF BRC-BAL-INT-SL/B …
- Z-2558-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Custom Procedure Kit (Angiography/angioplasty kit), REF: K12T-107…
- Z-2559-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Outake, REF CET200 and CET300, The Catheter Extractor is a cath…
- Z-2560-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K02 (Custom Syringe Kit), REF: K02-00873A, Inject fluids into, o…
- Z-2561-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF:K04-00791A An intravascular admini…
- Z-2562-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K04 (Custom Syringe Kit), REF: K05-00203M K05-00511J K05-00547J K…
- Z-2563-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K08-PI (Custom Fluid Management Kit), REF: K08-02042 K08-02948 K0…
- Z-2564-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K09/K09T (Custom Manifold Kit), REF: K09-00526L K09-00807AH K09-0…
- Z-2565-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K10/K10-PI (Custom Waste Management Kit), REF: K10T-05815 K10T-05…
- Z-2566-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K11 (Custom Pressure Monitoring Kit), REF: K11-00382C To support…
- Z-2567-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K12T/12PI (Custom Pressure Monitoring Kit), REF: K12-03846 K12-04…
- Z-2568-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: K22C (Vein Tray Angiography kit), REF: K09T-11796F K12T-01558G K1…
- Z-2569-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: PC (Pericardiocentesis Tray - Fluid drainage tray), REF: PC101/B …
- Z-2570-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: ReSolve (Catheter and Tip Suction kit), REF: RTT14038MB/A The Re…
- Z-2571-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG (Introducer Catheter), REF: FCL-069-00/A FCL-069-0…
- Z-2572-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude Prestige (Introducer Catheter), REF: PLP-2505/A PLP-2514/…
- Z-2573-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Custom Procedure Kit (Angiography/Angioplasty kit), REF: OPT0242-…
- Z-2575-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Prelude SNAP-H (Introducer Catheter kit), REF: PLSH-1006/A PLSH-1…
- Z-2576-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Safety Paracentesis Procedure Tray (Paracentesis Tray kit), REF: …
- Z-2577-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: SafeSheath CSG/Worley advanced CSG (Introducer Catheter kit), REF…
- Z-2578-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K09T-0…
- Z-2579-2024 · Open, ClassifiedVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes: Customer Procedure Kit (Angiography/Angioplasty kit), REF: K05-021…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page