Skip to content

CPT recallZ-2450-2024 · Zimmer, Inc. · initiated Jul 2, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.

FDA root cause category: Under Investigation by firm

What the company is doing

On July 2, 2024, the firm notified customers through Urgent Medical Device Correction letters. Customers were informed of the increased risk of postoperative periprosthetic femoral fracture (PFF). The firm will be updating the IFU to reflect the risk of PFF, with anticipated release by the end of August 2024. In addition, the firm will begin phasing out the device with a target of December 2024 to stop sales. The latest IFU can be viewed electronically at labeling.zimmerbiomet.com and entering the part number. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday, or email [email protected]. *** UPDATED 9/18/24: Follow-up communication was distributed to impacted US surgeons on September 18, 2024. This updated letter reflected FDA's Medical Device Safety Communication issued on September 17, 2024, which provided additional recommendations on the use of the affected product during the phase out period, including that surgeons should consider using an alternative prosthesis where possible. The FDA Safety Communication can be found here: https://www.fda.gov/medical-devices/safety-communications/zimmer-biomet-cpt-hip-system-femoral-stem-and-increased-risk-thigh-bone-fracture-fda-safety?utm_medium=email&utm_source=govdelivery. The updated letter also includes a plain language communication that may be provided to patients. The firm stated all US surgeons will be transitioned to alternative brands and the CPT Hip System Femoral Stem will be exited in the US by the end of October 2024.

Recall details

Recall number
Z-2450-2024
Status
Open, Classified
Initiated by the company
Jul 2, 2024
Posted by the FDA
Jul 30, 2024
Recalling firm
Zimmer, Inc.
Quantity
5,500 (US); 242,000 (OUS)
Product code
JDI
FDA submissions
K960658
Distribution
Worldwide - US Nationwide distribution.

Product description

CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010 00811400100 00811400110 00811400200 00811400210 00811400218 00811400230 00811400300 00811400310 00811400318 00811400330 00811400400 00811400410 00811400420 00811400423 00811400426 00811400430 00811400500 00811400510 00811400530 00811401218 00811401318 00811404000 00811405000

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Item Number / UDI-DI 00811400000 / 00889024145733 00811400010 / 00889024145740 00811400100 / 00889024145757 00811400110 / 00889024145764 00811400200 / 00889024145771 00811400210 / 00889024145788 00811400218 / 00889024145795 00811400230 / 00889024145801 00811400300 / 00889024145818 00811400310 / 00889024145825 00811400318 / 00889024145832 00811400330 / 00889024145849 00811400400 / 00889024145856 00811400410 / 00889024145863 00811400420 / 00889024145870 00811400423 / 00889024145887 00811400426 / 00889024145894 00811400430 / 00889024145900 00811400500 / 00889024145917 00811400510 / 00889024145924 00811400530 / 00889024145931 00811401218 / 00889024145948 00811401318 / 00889024145955 00811404000 / 00889024145962 00811405000 / 00889024145979

Devices named in this recall

25 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page