Luminos dRF Max- Intended to visualize anatomical… recallZ-2526-2024 · Siemens Medical Solutions USA, Inc · initiated Jul 2, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued Customer Safety Advisory Notice (CSAN) XP017/24/S to Director of Diagnostic Imaging and/ or Radiology Laboratory on 7/2/24. Letter states reason for recall, health risk and action to take: Please visually inspect the mounting screw on the ceiling tube (Figure 1) or wall tube (Figure 2) of the display suspension as far as this is possible from the floor. In case the visual inspection of the mounting screw does not show any abnormalities, you can continue to use the system. Nevertheless, until the system has been checked by Siemens Healthineers customer service with the planned field action, we recommend: - not to swivel the support arm of the display ceiling/wall suspension to the mechanical limit with great force and - to ensure that there is no one under the display ceiling/wall suspension when positioning the displays. This Customer Safety Advisory Notice (XP017/24/S) is being distributed to all potentially affected customers. Siemens Healthineers is preparing a field safety corrective action that will be provided with Update XP018/24/S starting in the third quarter of 2024. The corrective action will involve check and replacement of the attachment of the support arm for the display ceiling/wall suspension by a service technician in an onsite visit. In instances where the mounting screw is broken, the support arm will also be replaced with update XP019/24/S. These field actions will be provided to you free of charge. Once the corrective actions are available, our customer service team will contact you to schedule an appointment to perform the above safety corrective action(s). Once the updates are available and if you would like to make an earlier appointment, please feel free to contact customer service at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification
Recall details
- Recall number
- Z-2526-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 2, 2024
- Posted by the FDA
- Aug 7, 2024
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 24 units
- Product code
- OWB
- FDA submissions
- K173639
- Distribution
- US Nationwide distribution.
Product description
Luminos dRF Max- Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762471
Code information (models, lots, serial numbers)
UDI: Serial Numbers: Serial Number 8320 8198 8334 8474 8183 8322 8613 8244 8660 8436 8095 8371 8378 7805 5664 5615 8150 8130 8300 8093 8468 7791 8479 7136
Other devices under the same FDA submission
2 devices cleared under K173639 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LUMINOS Lotus Max 11574100Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system2 recalls7 reports
- Multitom Rax 10860700Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94948).
- Z-2525-2024 · Open, ClassifiedLuminos Agile Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image. Model Number: 10762472
- Z-2527-2024 · Open, ClassifiedLUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100
- Z-2528-2024 · Open, ClassifiedMultitom Rax -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 10860700
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page