TruDi NAV Suction recallZ-2507-2024 · Integra LifeSciences Corp. · initiated Jul 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
FDA root cause category: Process control
What the company is doing
On July 9, 2024 URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters were sent to consignees. What Actions Are Required: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 2) confirming receipt of this notice and email it to Sedgwick at [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 5. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Sedgwick. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 6. To return product subject to this recall, photocopy the completed Business Reply Form (BRF), place it in the return kit box received with the product, and affix the pre-paid authorized shipping label included with the return kit. Acclarent will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling 855-215-5039. Keep the original Business Reply Form (BRF) for your records. 7. To receive replacement, customers must return product subject to this recall by August 31, 2024. Any non-affected product that is sent to Sedgwick will not receive replacement nor will the product be returned. DO NOT return impacted product to Acclarent, replacement will not be given, nor will the product be returned. If you have product complaints or adverse events to report regarding t
Recall details
- Recall number
- Z-2507-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 9, 2024
- Posted by the FDA
- Aug 5, 2024
- Recalling firm
- Integra LifeSciences Corp.
- Quantity
- 265 units
- Product code
- PGW
- FDA submissions
- K180948
- Distribution
- The devices are only sold in the US. The following states have received the product: AL, AZ CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, PA, SC, SD, TN, TX, VA, VT, WI.
Product description
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
Code information (models, lots, serial numbers)
UDI/DI: 10705031245877; Lot Codes: 2310160, 2309110, 2309150, 2308140, 2308160, 2308180.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TruDi NAV Suction TDNS000ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall61 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K180948 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- TruDi NAV Suction TDNS070ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument25 reports
- TruDi NAV Suction TDNS090ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument14 reports
- TruDi NAV Cable TDNC001Acclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page