Adverse event reports
100
Deaths reported
0
Injuries reported
11
Recalls
1

We found 3 models of TruDi NAV Suction by Acclarent, Inc. in the FDA's device registry (GUDID). Together they are named in 100 adverse event reports in the FDA's MAUDE database, received between Mar 18, 2019 and Dec 10, 2025 — classified as 11 injuries, 89 malfunctions.

The most frequently reported specific device problem is “Incorrect Interpretation of Signal” (82 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 7 reportsDec 2023: 3 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 8 reportsMay 2024: 5 reportsJun 2024: 1 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 9 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

TruDi NAV Suction recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 9, 2024Open, Classified

    Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.

    Z-2507-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported TruDi NAV Suction problems

Device problems

  • Incorrect Interpretation of Signal82
  • Patient Device Interaction Problem6
  • Adverse Event Without Identified Device or Use Problem5
  • Calibration Problem5
  • Display or Visual Feedback Problem4
  • Communication or Transmission Problem1
  • Connection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions86
  • Cerebrospinal Fluid Leakage6
  • No Consequences Or Impact To Patient5
  • Hemorrhage/Blood Loss/Bleeding3
  • Hypoxia2

TruDi NAV Suction models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent TruDi NAV Suction adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Acclarent, Inc. products

All Acclarent, Inc. devices →

Sources and updates