Syringes with exchangeable needles labeled as recallZ-3075-2024 · Sol-Millennium Medical Inc. · initiated May 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sol-Millennium Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
FDA root cause category: Other
What the company is doing
Sol M issued a Voluntary Recall Action Letter to Distributors to its consignees on 05/09/2024 via email. The notice explained the reason for the recall and requested the following: "ACTIONS REQUIRED: 1. Examine your inventory and quarantine the affected product codes and lots subject to this recall action. 2. Distributors, please identify your customers, and notify them at once of this recall action to the distributor level. Such notification may be enhanced by including the enclosed letter (Appendix B) for your customers and Healthcare Providers (HCP s) in the field. 3. Return the enclosed Appendix B with the information via email to [email protected]. The quarantine should be carried out to the wholesale and distributor level. Products in the field beyond the control of the distributor or wholesaler should be recalled from the field and returned to the distributor. Distributors should quarantine the products in inventory and quarantine the products received from customers until the 510(k) submissions are pending or cleared. The returned product should be quarantined in a secure location to avoid distribution based on your quality management procedures. We will provide communication with any updates as we work with the FDA."
Recall details
- Recall number
- Z-3075-2024
- Status
- Open, Classified
- Initiated by the company
- May 9, 2024
- Posted by the FDA
- Sep 6, 2024
- Recalling firm
- Sol-Millennium Medical Inc.
- Product code
- FMF
- Distribution
- US Nationwide, Czech Republic, Belgium, Portugal, Canada, Saudi Arabia, Netherland, France, Switzerland, Latvia, UK, Spain, Germany, Philippines, Australia, Malaysia, Jamaica, Slovenia, Brazil, Taiwan, Denmark, Sweden, Singapore, Romania, Poland, Barbados, Mexico, Guatemala, Chile, Ireland, Italy, Panama
Product description
Syringes with exchangeable needles labeled as: SOL-CARE: 1) SOL-CARE 10ml Luer Lock Syringe w/Safety Needle 21G*1 1/2 (needle aside), Model No. 102115SN, UPC 818392019858; 2) SOL-CARE 10ml Luer Lock Syringe w/Safety Needle 22G*1 1/2 (needle aside), Model No. 102215SN, UPC 818392019865; 3) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*1 (needle aside), Model No. 12510SN, UPC 818392018851; 4) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*5/8 (needle aside), Model No. 12558SN, UPC 818392018950; 5) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 27G*1/2 (needle aside), Model No. 12705SN, UPC 818392018967; 6) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 30G*1/2 (needle aside), Model No. 13005SN, UPC 818392018974; 7) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 21G*1 (needle aside), Model No. 32110SN, UPC 818392019001; 8) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 21G*1 1/2 (needle aside), Model No. 32115SN, UPC 818392019018; 9) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 22G*1 (needle aside), Model No. 32210SN, UPC 818392019025; 10) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 22G*1 1/2 (needle aside), Model No. 32215SN, UPC 818392019032; 11) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 23G*1 (needle aside), Model No. 32310SN, UPC 818392019049; 12) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 25G*1 (needle aside), Model No. 32510SN, UPC 818392019063; 13) SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 25G*5/8 (needle aside), Model
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
a) SOL-M: 1) Model No. 1812010, UPC 818392010268, Lot Numbers: 4202007; 2) Model No. 1812015, UPC 818392010275, Lot Numbers: 4110010; 3) Model No. 1812110, UPC 818392010244, Lot Numbers: 4202008; 4) Model No. 1812115, UPC 818392010251, Lot Numbers: 2108094, 4110087, 4110088, 4303013, 4303015, 4304027, 4311001, 4312014, 4312022; 5) Model No. 1812210, UPC 818392010237, Lot Numbers: 4302040; 6) Model No. 1832010, UPC 818392011586, Lot Numbers: 4110005; 7) Model No. 1832015, UPC 818392010114, Lot Numbers: 1010016, 1012011, 1012019, 1110067, 1112021, 1112056, 1301021, 4110005; 8) Model No. 1832110, UPC 818392010084, Lot Numbers: 5105119, 19041740; 9) Model No. 1832115, UPC 818392010091, Lot Numbers: 4106053, 4202035, 4306035; 10) Model No. 1832210, UPC 818392010039, Lot Numbers: 4110006, 4209012; 11) Model No. 1832215, UPC 818392010077, Lot Numbers: 4102053, 4106153, 4202003, 4202003, 4306034, 4307015, 4307043, 4308047, 4309030, 4309040, 4310033; 12) Model No. 1832310, UPC 818392010107, Lot Numbers: 4102058, 4102059, 4103020, 4103021, 4103028, 4103029, 4106048, 4106049, 4106052, 4110007, 4110053, 1902183, 19021824; 13) Model No. 1832510, UPC 818392010053, Lot Numbers: 4012001, 4102056, 4103014, 4103015, 4103022, 4103023, 4103030, 4103031, 4106050, 4106051, 4109091, 19021825, 19021826, 19041742; 14) Model No. 1832515, UPC 818392010046, Lot Numbers: 4012002, 4102054, 4103018, 4103019; 15) Model No. 1832558, UPC 818392010022, Lot Numbers: 4007045, 40 10033, 4102050, 4102051, 4102052, 4103016, 4103017, 4110011, 4110012; 16) Model No. 1852010, UPC 818392010190, Lot Numbers: EXPIRED 17) Model No. 1852015, UPC 818392010206, Lot Numbers: EXPIRED 18) Model No. 1852115, UPC 818392010183, Lot Numbers: 4103076, 4105110; 19) Model No. 1852215, UPC 818392010169, Lot Numbers: 4001001, 4003019, 4110083, 4110084, 4202004, 4308046, 4308088, 4312097, 19052010; 20) Model No. 110101120011, UPC 818392014341, Lot Numbers: 4212007, 4306027, 4308066, 19030602; 21) Model No. 110101120012, UPC 818392014358, Lot Numbers: 4001021, 4008151, 4104087, 4110036, 4202006, 19030603; 22) Model No. 1832110T, UPC 818392012880, Lot Numbers: 4103002, 4104043, 4105120, 4106152, 4109083, 4110045, 4111040, 4206041, 4206042, 4206043, 4208008, 4208009, 4208010, 4303041, 4304026, 4306046, 4308071, 4312018; 23) Model No. 1832210T, UPC 818392013061, Lot Numbers: 4110046, 4307048, 4308091; 24) Model No. 1832310T, UPC 10818392013075, Lot Numbers: 4110047; 25) Model No. 1852010T, UPC 818392013702, Lot Numbers: 4104083; 26) Model No. 1852110T, UPC 818392013696, Lot Numbers: 4005030, 4007049, 4103080, 4104084, 4105111, 4302024, 4308067, 4310031, 4312019, 4312102, 4912026, 19072314; b) SOL-CARE: 1) Model No. 102115SN, UPC 818392019858, Lot Numbers: 4104022; 2) Model No. 102215SN, UPC 818392019865, Lot Numbers: 4004004, 4303018, 4304057; 3) Model No. 12510SN, UPC 818392018851, Lot Numbers: 4108037, 7106031, 7109036; 4) Model No. 12558SN, UPC 818392018950, Lot Numbers: 4004005, 4104008, 4104011, 4104012, 4106034, 7107051, 7112016; 5) Model No. 12705SN, UPC 818392018967, Lot Numbers: 4004006, 7109007; 6) Model No. 13005SN, UPC 818392018974, Lot Numbers: 4004002; 7) Model No. 32110SN, UPC 818392019001, Lot Numbers: 4004031, 5101401; 8) Model No. 32115SN, UPC 818392019018, Lot Numbers: 4109051, 19072315; 9) Model No. 32210SN, UPC 818392019025, Lot Numbers: 4007046, 4301035, 4312117; 10) Model No. 32215SN, UPC 818392019032, Lot Numbers: 4301036; 11) Model No. 32310SN, UPC 818392019049, Lot Numbers: 3102005, 4004031, 4104013, 4107091, 4108035, 5104011, 5104012; 12) Model No. 32510SN, UPC 818392019063, Lot Numbers: 4007047, 4007051, 4104009, 4105096, 4105100, 4108031, 4108032, 4108036, 7111006; 13) Model No. 32558SN, UPC 818392019070, Lot Numbers: 4007048, 4007052, 4104010, 4104018; 14) Model No. 52115SN, UPC 818392018912, Lot Numbers: 4110017, 4309051; 15) Model No. 52215SN, UPC 818392018936, Lot Numbers: 4004003, 4107173; c) HENRY SCHEIN: 1) Model No. 900-7842, UPC 10304040017200, Lot Numbers: 1012089, 1012091, 1102061, 1103043, 1103044, 1103045, 1103046, 2111063, 2111064, 2203033, 2209013, 2211025, 2211026, 2307003, 2307094, 2309046, 4005032, 4005054, 4007040, 4007041; 2) Model No. 900-7843, UPC 10304040017217, Lot Numbers: 1012090, 1102062, 1103053, 2211027, 2303042, 2307005, 4003020, 4004038, 4007042, 4109001, 4109002, 4910032, 4911007; 3) Model No. 900-7844, UPC 10304040017224, Lot Numbers: 1012090, 1102062, 1103053, 2211027, 2303042, 2307005, 4003020, 4004038, 4007042, 4109001, 4109002, 4910032, 4911007
Devices named in this recall
41 devices in the FDA device registry whose identifier appears in the recall's code information.
- SOL-CARE Syringe with Safety Needle 32310SNSol-Millennium Medical, Inc. · Syringe, piston1 recall5 reports
- Henry Schein 9007842Henry Schein, Inc. · Syringe, piston1 recall2 reports
- SOL-CARE Syringe with Safety Needle 32510SNSol-Millennium Medical, Inc. · Syringe, piston1 recall2 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832110TSol-Millennium Medical, Inc. · Syringe, piston1 recall2 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1852110TSol-Millennium Medical, Inc. · Syringe, piston1 recall2 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832310Sol-Millennium Medical, Inc. · Syringe, piston1 recall1 report
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1812110Sol-Millennium Medical, Inc. · Syringe, piston1 recall1 report
- Henry Schein 9007843Henry Schein, Inc. · Syringe, piston1 recall0 reports
- Henry Schein 9007844Henry Schein, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832558Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832210Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832515Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832510Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832215Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832110Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832115Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe w/Exch Needle 1832015Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1852215Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1852115Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1852010Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe w/Exch Needle 1852015Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1812210Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1812115Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1812010Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1812015Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1832010Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-CARE 5ml Luer Lock Syringe w/Safety Needle 21G*1/2 52115SNSol-Millennium Medical, Inc. · Syringe, Piston1 recall0 reports
- SOL-CARE 5ml Luer Lock Syringe w/Safety Needle 22G*1/2 52215SNSol-Millennium Medical, Inc. · Syringe, Piston1 recall0 reports
- SOL-CARE Syringe with Safety Needle 12558SNSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-CARE Syringe with Safety Needle 12705SNSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-CARE Syringe with Safety Needle 13005SNSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-CARE Syringe with Safety Needle 32110SNSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 21G*1 1/2 32115SNSol-Millennium Medical, Inc. · Syringe, Piston1 recall0 reports
- SOL-CARE 3ml Luer Lock Syringe w/Safety Needle 22G*1 32210SNSol-Millennium Medical, Inc. · Syringe, Piston1 recall0 reports
- SOL-CARE Syringe with Safety Needle 32215SNSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-CARE Syringe with Safety Needle 32558SNSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe w/Exch Needle Tray 1832210TSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe w/Exch Needle Tray 1832310TSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 1852010TSol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 110101120011Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
- SOL-M™ Luer Lock Syringe with Exchangeable Needle 110101120012Sol-Millennium Medical, Inc. · Syringe, piston1 recall0 reports
Product lines
- SOL-M Luer Lock Syringe with Exchangeable NeedleSol-Millennium Medical, Inc. · 22 models · 6 reports · 1 recall
- SOL-CARE Syringe with Safety NeedleSol-Millennium Medical, Inc. · 8 models · 7 reports · 1 recall
- SOL-M Luer Lock Syringe w/Exch NeedleSol-Millennium Medical, Inc. · 2 models · 0 reports · 1 recall
- SOL-M Luer Lock Syringe w/Exch Needle TraySol-Millennium Medical, Inc. · 2 models · 0 reports · 1 recall
Other products in the same recall event
The FDA groups related products into one recall event (event 94968).
- Z-3074-2024 · Open, ClassifiedSyringes with Fixed Needles labeled as: a) SOL-M: 1) SOL-M 1ml TB Syringe w/Fixed Needle 25G*5/8'', Model No. 181025, UPC 818392014884; 2) SOL-M 3ml Luer Lock Syringe w/Blunt …
- Z-3076-2024 · Open, ClassifiedSyringes with Luer Slip Tips labeled as: 1) SOL-M 1ml Slip Tip Syringe w/o Needle, Model No. 180011, UPC 8183923010138; 2) SOL-M 1ml Slip Tip Syringe w/o Needle, Model No. P1800…
- Z-3078-2024 · Open, ClassifiedSyringes with Low Dead Space labeled as: a) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. 180011LDS, UPC 840368511899; b) SOL-M 1ml Slip Tip Syringe w/o need…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page