1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10818392013693), received between Dec 6, 2019 and Dec 30, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 9, 2024Class IIOpen, Classified

    Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.

    Z-3075-2024 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Crack1
  • Decoupling1
  • Fluid/Blood Leak1

Patient problems

  • Chemical Exposure1
  • No Clinical Signs, Symptoms or Conditions1
  • Radiation Exposure, Unintended1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 30, 2025Decoupling; Decoupling

    CUSTOMER REPORTED THAT THE 3CC AND 5CC ARE BOTH ARRIVING WITH BARELY ANY NEEDLES ATTACHED, AND THE ONES THAT ARE ATTACHED WILL GET STUCK IN THE VIAL AFTER DRAWING THE DOSE AND THE SYRINGE WILL BE IN HAND.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23921449 ↗
  • MalfunctionDec 6, 2019Crack; Fluid/Blood Leak

    SUS VOLUNTARY EVENT REPORT NUMBER: MW5090432. A VOLUNTARY EVENT FORM WAS RECEIVED FROM THE FDA AT (B)(6) BY POSTAL SERVICE MAIL ON 11/7/2019. PER THE VOLUNTARY EVENT REPORT, CUSTOMER REPORTED TO THE FDA THAT A 5CC SYRINGE HAD A HAIRLINE CRACK ALONG THE BARREL CAUSING LEAKAGE AND CONTAMINATION OF THE RADIOPHARMACEUTICAL. THE INTEGRITY AND STERILITY OF THE PRODUCT WAS COMPROMISED PER THE CUSTOMER. ON 12/3/2019 ADDITIONAL INFORMATION WAS RECEIVED. THE (B)(6) PHARMACIST PROVIDED A PICTURE OF THE CRACKED SYRINGE. THE PHARMACIST STATED THAT TC-99M MACRO AGGREGATED ALBUMIN WAS THE DRUG THAT WAS FILLED IN THE SYRINGE. THE EXACT DRUG LEAKAGE AMOUNT WAS NOT KNOWN. THE END USER WAS (B)(6) HOSPITAL. PER THE PHARMACIST, HOSPITAL PERSONNEL WAS CONTAMINATED WITH RADIOACTIVE MATERIAL, BUT THERE DOES NOT APPEAR TO HAVE BEEN A REPORTED INJURY. (B)(6) PHARMACIST PROVIDED THE CUSTOMER (END USER) INVESTIGATIVE REPORT. SUMMARY OF (B)(6) HOSPITAL INVESTIGATIVE REPORT: DURING A PROCEDURE ON (B)(6) 2019, THE DOCTOR GAVE THE SYRINGE A STRONGER PUSH AND THE FORCE OF THE INJECTION LEAKED THROUGH THE SYRINGE AT A CRACK DEFECT AND DOSE WAS INADVERTENTLY SPRAYED ONTO DOCTOR. THE PROCEDURE WAS ABORTED AFTER PATIE

    Patient problems: Chemical Exposure; Radiation Exposure, Unintended

    MAUDE report 9432985 ↗

Other Syringe, piston devices

All Syringe, piston devices →