Aisys recallZ-0083-2025 · Datex-Ohmeda, Inc. · initiated Jul 12, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datex-Ohmeda, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
FDA root cause category: Device Design
What the company is doing
GE Healthcare notified consignees on about 07/12/2024 via letter. Consignees were instructed to ensure all potential users complete the Preoperative Checkout procedure in its entirety, as detailed in the URM, make sure that the breathing circuit is correctly connected, only connect an auxiliary manual breathing circuit to the ACGO port, place the included addendum with the URM and inspect all potentially affected devices for the presence of covers applied to the ACGO Port and ACGO Switch. Additionally, customers were requested to complete and return the response form.
Recall details
- Recall number
- Z-0083-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 12, 2024
- Posted by the FDA
- Oct 18, 2024
- Recalling firm
- Datex-Ohmeda, Inc.
- Quantity
- 92 units
- Product code
- QSF
- FDA submissions
- P210018
- Distribution
- Worldwide distribution - US Nationwide and the countries of Aland Islands, ALBANIA, Algeria, Argentina, Australia, Austria, Azerbaijan, BAHAMAS, Bahrain, Bangladesh, BARBADOS, BELARUS, Belgium, Benin, Bermuda, Bolivia, BONAIRE, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Canada, Chile, CHINA, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, ETHIOPIA, Falkland Islands, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gilbrator, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, ICELAND, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Jamaica, JAPAN, Jordan, Kazakhstan, KENYA, Korea (Republic Of), Korea, Republic of, Kuwait, Laos, LATVIA, Lebanon, Libya, LITHUANIA, Macedonia, Malaysia, Maldives, Malta, MAURITIUS, Mexico, Moldova, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, PAKISTAN, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, PUERTO RICO, Qatar, REUNION, ROMANIA, Russia, San Marino, Saudi Arabia, Senegal, Serbia, SINGAPORE, Slovakia, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, THAILAND, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Uganda, UKRAINE, United Arab Emirates, United Kingdom, Uruguay, UZBEKISTAN, Venezuela, Vietnam, Yemen.
Product description
Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
Code information (models, lots, serial numbers)
GTIN: 00195278588128. Model Number: 1011-9055-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Aisys CS2 WITH ET CONTROLDatex-Ohmeda Inc · Gas-machine, anesthesia2 recalls98 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95045).
- Z-0074-2025 · Open, ClassifiedAespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory suppo…
- Z-0075-2025 · Open, ClassifiedAespire 100. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals t…
- Z-0076-2025 · Open, ClassifiedAespire 7900. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals …
- Z-0077-2025 · Open, ClassifiedAespire View. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals …
- Z-0078-2025 · Open, ClassifiedAvance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals traine…
- Z-0079-2025 · Open, ClassifiedAmingo. OUS only.
- Z-0080-2025 · Open, ClassifiedAvance CS2 and Avance CS2 Pro. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by me…
- Z-0081-2025 · Open, ClassifiedAisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained…
- Z-0082-2025 · Open, ClassifiedAisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals tra…
- Z-0084-2025 · Open, ClassifiedAuxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, howev…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page