Skip to content

Aisys CS2 WITH ET CONTROLRecalls & reported problems

Datex-Ohmeda Inc · AISYS CS2 WITH ET CONTROL · Class 2

All models of Aisys (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Datex-Ohmeda Aisys CS2 is an anesthesia gas machine. The FDA records hold 96 reports, with no deaths and no injuries; many describe low pressure leak test failures, low oxygen or gas flow, and battery failures, often during pre-use checks with no patient involved. There are 2 open recalls: one says the machine can switch to all-oxygen delivery, and the other says a breathing circuit may be connected to the wrong port.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
98
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 98 adverse event reports that list this exact device (DI 00195278588128), received between Mar 28, 2024 and Sep 30, 2026 — classified as 98 malfunctions.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (26 reports).

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K170872); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 4 reportsOct 2025: 11 reportsNov 2025: 6 reportsDec 2025: 7 reportsJan 2026: 7 reportsFeb 2026: 5 reportsMar 2026: 6 reportsApr 2026: 6 reportsMay 2026: 6 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Open, Classified

    Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

    Z-3190-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Jul 12, 2024Open, Classified

    Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

    Z-0083-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion26
  • Mechanical Problem19
  • Leak/Splash7
  • No Flow7
  • Electrical /Electronic Property Problem6
  • Therapeutic or Diagnostic Output Failure5
  • Unexpected Therapeutic Results5
  • Failure to Run on Battery4

Patient problems

  • No Clinical Signs, Symptoms or Conditions98

Recent adverse event reports

The 25 most recent of 98 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949825 ↗
  • MalfunctionSep 28, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926590 ↗
  • MalfunctionSep 28, 2026Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918423 ↗
  • MalfunctionAug 27, 2026Insufficient Flow or Under Infusion

    It was reported that there was a low-pressure leak test failure resulting in insufficient O2 during pre-use checkout. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50479190 ↗
  • MalfunctionJun 29, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25656887 ↗
  • MalfunctionJun 9, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN BATTERY FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448768 ↗
  • MalfunctionMay 27, 2026Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25299483 ↗
  • MalfunctionMay 21, 2026Leak/Splash

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOW PRESSURE LEAK TEST FAILURE DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241388 ↗
  • MalfunctionMay 21, 2026Leak/Splash

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOW PRESSURE LEAK TEST FAILURE DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241358 ↗
  • MalfunctionMay 18, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25203044 ↗
  • MalfunctionMay 14, 2026Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25172115 ↗
  • MalfunctionMay 11, 2026Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN AUXILIARY OXYGEN IS UNAVAILABLE OR LOSS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25141976 ↗
  • MalfunctionApr 28, 2026Unexpected Therapeutic Results

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN UNEXPECTED THERAPEUTIC SYSTEM PERFORMANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25021166 ↗
  • MalfunctionApr 21, 2026Leak/Splash; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LARGE LEAK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942660 ↗
  • MalfunctionApr 17, 2026Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OVER DELIVERY OF PRESSURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24908577 ↗
  • MalfunctionApr 15, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24889480 ↗
  • MalfunctionApr 7, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN PROLONGED INSUFFICIENT OXYGEN (O2) CONCENTRATION OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24802607 ↗
  • MalfunctionApr 6, 2026Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DUE TO A LARGE LEAK DURING A CASE. THE UNIT WAS REPLACED AND CASE COMPLETED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24790471 ↗
  • MalfunctionMar 26, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A SYSTEM MALFUNCTION DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698280 ↗
  • MalfunctionMar 17, 2026Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LEAK THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24617582 ↗
  • MalfunctionMar 12, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN SYSTEM SHUTDOWN DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579477 ↗
  • MalfunctionMar 10, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT AGENT DELIVERY LESS THAT -20% OF SETTING DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24558901 ↗
  • MalfunctionMar 10, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING SERVICE RESULTING IN INSUFFICIENT OXYGEN (O2) CONCENTRATION OR FRESH GAS FLOW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557653 ↗
  • MalfunctionMar 3, 2026No Flow

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF AUXILIARY O2 DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24493552 ↗
  • MalfunctionFeb 19, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24388146 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170872. One submission often covers many models, so check the product and lot details in each record.

  • Z-0412-2019 · Sep 27, 2018 · Terminated
    Aisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRA
  • Z-0413-2019 · Sep 27, 2018 · Terminated
    Touchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X
  • Z-0411-2019 · Sep 27, 2018 · Terminated
    Aisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE K
  • Z-1199-2018 · Oct 4, 2017 · Terminated
    DATEX-OHMEDA AISYS CS2 ALADIN2 CASSETTE Cassettes are intended to be used to deliver anesthetic agent.

Other Gas-machine, anesthesia devices

All Gas-machine, anesthesia devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page