Aisys CS2 WITH ET CONTROLRecalls & reported problems
Datex-Ohmeda Inc · AISYS CS2 WITH ET CONTROL · Class 2
All models of Aisys (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datex-Ohmeda Aisys CS2 is an anesthesia gas machine. The FDA records hold 96 reports, with no deaths and no injuries; many describe low pressure leak test failures, low oxygen or gas flow, and battery failures, often during pre-use checks with no patient involved. There are 2 open recalls: one says the machine can switch to all-oxygen delivery, and the other says a breathing circuit may be connected to the wrong port.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 98
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 98 adverse event reports that list this exact device (DI 00195278588128), received between Mar 28, 2024 and Sep 30, 2026 — classified as 98 malfunctions.
The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (26 reports).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K170872); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2026Open, Classified
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Z-3190-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Jul 12, 2024Open, Classified
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
Z-0083-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Flow or Under Infusion26
- Mechanical Problem19
- Leak/Splash7
- No Flow7
- Electrical /Electronic Property Problem6
- Therapeutic or Diagnostic Output Failure5
- Unexpected Therapeutic Results5
- Failure to Run on Battery4
Patient problems
- No Clinical Signs, Symptoms or Conditions98
Recent adverse event reports
The 25 most recent of 98 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949825 ↗MalfunctionSep 28, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926590 ↗MalfunctionSep 28, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50918423 ↗MalfunctionAug 27, 2026Insufficient Flow or Under Infusion
It was reported that there was a low-pressure leak test failure resulting in insufficient O2 during pre-use checkout. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50479190 ↗MalfunctionJun 29, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25656887 ↗MalfunctionJun 9, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN BATTERY FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25448768 ↗MalfunctionMay 27, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25299483 ↗MalfunctionMay 21, 2026Leak/Splash
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOW PRESSURE LEAK TEST FAILURE DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25241388 ↗MalfunctionMay 21, 2026Leak/Splash
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOW PRESSURE LEAK TEST FAILURE DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25241358 ↗MalfunctionMay 18, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25203044 ↗MalfunctionMay 14, 2026Mechanical Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25172115 ↗MalfunctionMay 11, 2026Mechanical Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN AUXILIARY OXYGEN IS UNAVAILABLE OR LOSS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25141976 ↗MalfunctionApr 28, 2026Unexpected Therapeutic Results
IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN UNEXPECTED THERAPEUTIC SYSTEM PERFORMANCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25021166 ↗MalfunctionApr 21, 2026Leak/Splash; Mechanical Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LARGE LEAK. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24942660 ↗MalfunctionApr 17, 2026Mechanical Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OVER DELIVERY OF PRESSURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24908577 ↗MalfunctionApr 15, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24889480 ↗MalfunctionApr 7, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN PROLONGED INSUFFICIENT OXYGEN (O2) CONCENTRATION OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24802607 ↗MalfunctionApr 6, 2026Gas/Air Leak
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DUE TO A LARGE LEAK DURING A CASE. THE UNIT WAS REPLACED AND CASE COMPLETED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790471 ↗MalfunctionMar 26, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A SYSTEM MALFUNCTION DURING A CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24698280 ↗MalfunctionMar 17, 2026Gas/Air Leak
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LEAK THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24617582 ↗MalfunctionMar 12, 2026Unexpected Shutdown
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN SYSTEM SHUTDOWN DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24579477 ↗MalfunctionMar 10, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT AGENT DELIVERY LESS THAT -20% OF SETTING DURING A CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24558901 ↗MalfunctionMar 10, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING SERVICE RESULTING IN INSUFFICIENT OXYGEN (O2) CONCENTRATION OR FRESH GAS FLOW. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24557653 ↗MalfunctionMar 3, 2026No Flow
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF AUXILIARY O2 DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24493552 ↗MalfunctionFeb 19, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24388146 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170872. One submission often covers many models, so check the product and lot details in each record.
- Z-0412-2019 · Sep 27, 2018 · TerminatedAisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRA
- Z-0413-2019 · Sep 27, 2018 · TerminatedTouchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X
- Z-0411-2019 · Sep 27, 2018 · TerminatedAisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE K
- Z-1199-2018 · Oct 4, 2017 · TerminatedDATEX-OHMEDA AISYS CS2 ALADIN2 CASSETTE Cassettes are intended to be used to deliver anesthetic agent.
Other Gas-machine, anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,109 reports
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,204 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,015 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,540 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page