Skip to content

PE Centrifuge, REF: B36365, a component of the Power Express recallZ-2979-2024 · Beckman Coulter Biomedical GmbH · initiated Jul 17, 2024

Class II Open, Classified All recalls posted in September 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Biomedical GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On July 18, 2024, Beckman Coulter issued a "Urgent Medical Device Recall" notification to affected consignees. Beckman Coulter asked consignees to take the following actions: 1. follow your laboratory procedure to handle biohazard materials and contact your Beckman Coulter representative for servicing your Power Express system. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. 3. respond within 10 days in one of the following ways: . Electronically, if you received this communication via email. . Manually, complete and return the enclosed Response Form. 4. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Recall details

Recall number
Z-2979-2024
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jul 17, 2024
Posted by the FDA
Sep 5, 2024
Recalling firm
Beckman Coulter Biomedical GmbH
Product code
CDD
FDA submissions
K140496
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Bahrain, Canada, China, Colombia, Egypt, France, Germany, Italy, Jordan, Korea, Kuwait, Lebanon, Malaysia, Panama, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, United Arab Amirates, United Kingdom, Vietnam.

Product description

PE Centrifuge, REF: B36365, a component of the Power Express

Code information (models, lots, serial numbers)

UDI: (01)15099590673550/ All serial numbers Lots

Other devices under the same FDA submission

1 device cleared under K140496 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95064).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page