Power ExpressRecalls & reported problems
Beckman Coulter, Inc. · Power Express · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15099590672324), received between Sep 1, 2021 and Sep 1, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Imprecise or Erratic Results” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K140496); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Imprecise or Erratic Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2021Imprecise or Erratic Results; Imprecise or Erratic Results
THE CUSTOMER REPORTED THAT A STAT RACK IN BAY 7 OF DYNAMIC INLET WAS MISIDENTIFIED AS CENTRIFUGE BYPASS RACK IN POWER EXPRESS SYSTEM. AS A RESULT, THREE SAMPLE TUBES WERE ROUTED ON THE LINE AND WERE SORTED TO THE CONNECTED INSTRUMENTS WHICH GENERATED RESULTS, VALIDATED, AND HOSTED TO LIS (LABORATORY INFORMATION SYSTEM). THE UNCENTRIFUGED SAMPLE ROUTED TO DXC 800 INSTRUMENT. THE CUSTOMER REPORTED RECEIVING ERRONEOUS PATIENT RESULTS FOR MULTIPLE ANALYTES FOR THE CMP (COMPREHENSIVE METABOLIC PROFILE). THE SAMPLE WAS WHOLE BLOOD AND AN ERRONEOUS RESULTS WERE GENERATED FOR NA (SODIUM), K (POTASSIUM), CL (CHLORIDE), CO2 (CARBON DIOXIDE), BUN (BLOOD UREA NITROGEN), GLU (GLUCOSE), CREA (CREATININE), CALC (CALCIUM), TP (TOTAL PROTEIN), ALB (ALBUMIN), TBIL (TOTAL BILIRUBIN), TRIG (TRIGLYCERIDES), CHOLESTEROL ALP (ALKALINE PHOSPHATASE), ALT (ALANINE AMINOTRANSFERASE) AND AST (ASPARTATE AMINOTRANSFERASE). THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LAB. THERE WAS NO CHANGE IN PATIENT TREATMENT IN ASSOCIATION WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12411371 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140496. One submission often covers many models, so check the product and lot details in each record.
- Z-2978-2024 · Jul 17, 2024 · Open, ClassifiedPower Express, REF B90918
- Z-2980-2024 · Jul 17, 2024 · Open, ClassifiedPE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
- Z-2979-2024 · Jul 17, 2024 · Open, ClassifiedPE Centrifuge, REF: B36365, a component of the Power Express
- Z-1741-2020 · May 17, 2019 · TerminatedBeckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.
Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,197 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall911 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death694 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports