Skip to content

Medline convenience kits containing semi-rigid suction… recallZ-3114-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Jun 27, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

FDA root cause category: Process control

What the company is doing

Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-3114-2024
Status
Open, Classified
Initiated by the company
Jun 27, 2024
Posted by the FDA
Sep 11, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
3117 units
Product code
LRO
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Product description

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1092B; b) VIDEO BRONCHOSCOPY TRAY, Pack Number DYKE1666; c) SUCTION CANISTER KIT, Pack Number DYKM2196; d) SUCTION CANISTER KIT, Pack Number DYKM2196H; e) BRONCH LAB PACK-LF, Pack Number DYNJ24810B; f) THORACIC, Pack Number DYNJ907991F; g) RECTAL, Pack Number DYNJ908108A; h) ROBOTIC, Pack Number DYNJ908109C; i) CERVICAL, Pack Number DYNJ908748D; j) LAMI, Pack Number DYNJ908750D; k) MERCY TIFFIN BASIC-LF, Pack Number DYNJ909731; l) MERCY TIFFIN LITHOTOMY, Pack Number DYNJ909735; m) BASIC GENERAL SET UP, Pack Number DYNJ909945; n) BASIC GENERAL SET UP, Pack Number DYNJ909945A; o) AMB OCULOPLASTICS MUSC, Pack Number DYNJ910289; p) AMB OCULOPLASTICS MUSC, Pack Number DYNJ910289H; q) MTHZ TOTAL HIP, Pack Number DYNJC9311M; r) BRONCH KIT, Pack Number P906877

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

a) DYKE1092B: UDI/DI case 40889942777366, UDI/DI each10889942777365, Lot Numbers: 24BBK418, 24DBH210; b) DYKE1666: UDI/DI case 40193489708487, UDI/DI each10193489708486, Lot Numbers: 24ADB503, 24CDA282, 24CDB279, 24DDB221, 24EDA406; c) DYKM2196: UDI/DI case 40195327236916, UDI/DI each10195327236915, Lot Numbers: 23LBF762, 24BBB541, 24CBI973, 24DBH448; d) DYKM2196H: UDI/DI case 40195327236916, UDI/DI each10195327236915, Lot Numbers: 23LBF762, 24BBB541, 24CBI973, 24DBH448; e) DYNJ24810B: UDI/DI case 40889942440611, UDI/DI each10889942440610, Lot Numbers: 24CBE214; f) DYNJ907991F: UDI/DI case 40195327641468, UDI/DI each10195327641467, Lot Numbers: 24DLA042, 24DLA594; g) DYNJ908108A: UDI/DI case 40195327507153, UDI/DI each10195327507152, Lot Numbers: 24BBK532; h) DYNJ908109C: UDI/DI case 40195327507238, UDI/DI each10195327507237, Lot Numbers: 23LBT867, 24BBA187, 24DBI723, 24FBA058; i) DYNJ908748D: UDI/DI case 40195327549207, UDI/DI each10195327549206, Lot Numbers: 23LBE394, 24CBD863; j) DYNJ908750D: UDI/DI case 40195327549214, UDI/DI each10195327549213, Lot Numbers: 24FBC284; k) DYNJ909731: UDI/DI case 40195327464524, UDI/DI each10195327464523, Lot Numbers: 24ABU492; l) DYNJ909735: UDI/DI case 40195327464685, UDI/DI each10195327464684, Lot Numbers: 23LBH989, 24ABV099, 24CBG700; m) DYNJ909945: UDI/DI case 40195327512669, UDI/DI each10195327512668, Lot Numbers: 24ABV895; n) DYNJ909945A: UDI/DI case 40195327615803, UDI/DI each10195327615802, Lot Numbers: 24DBA530; o) DYNJ910289: UDI/DI case 40195327618675, UDI/DI each10195327618674, Lot Numbers: 24CBC847; p) DYNJ910289H: UDI/DI case 40198459012489, UDI/DI each10198459012488, Lot Numbers: 24CBC847; q) DYNJC9311M: UDI/DI case 40198459012489, UDI/DI each10198459012488, Lot Numbers: 24FDA111; r) P906877: UDI/DI case 40889942002581, UDI/DI each10889942002580, Lot Numbers: 24FDA021

Devices named in this recall

16 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95128).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page