Medline convenience kits containing semi-rigid suction… recallZ-3121-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Jun 27, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
FDA root cause category: Process control
What the company is doing
Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-3121-2024
- Status
- Open, Classified
- Initiated by the company
- Jun 27, 2024
- Posted by the FDA
- Sep 11, 2024
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 1076 units
- Product code
- OJH
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Product description
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) LAMINECTOMY CDS, Pack Number CDS981510L; b) ORTHO TRAUMA-LF, Pack Number DYNJ900720D; c) ORTHO TRAUMA-LF, Pack Number DYNJ900720F; d) KNEE ARTHROSCOPY, Pack Number DYNJ908090C; e) LAMINECTOMY, Pack Number DYNJ908091B; f) PODIATRY, Pack Number DYNJ908092C; g) ANTERIOR TOTAL HIP, Pack Number DYNJ908095B; h) CRANI, Pack Number DYNJ908097A; i) HIP PINNING, Pack Number DYNJ908102B; j) (76) LOWER EXTREMITY, Pack Number DYNJ908103B; k) SHOULDER ARTHROSCOPY, Pack Number DYNJ908110B; m) HAND-LF, Pack Number DYNJ908425; n) TOTAL JOINT, Pack Number DYNJ908660A; o) MERCY TIFFIN HAND, Pack Number DYNJ909734; p) MERCY TIFFIN TOTAL KNEE, Pack Number DYNJ909738; q) MTHZ KNEE ARTHROSCOPY, Pack Number DYNJ909740; r) ACL-WHSC, Pack Number DYNJ910074A; s) EXTREMITY-WHSC, Pack Number DYNJ910077; t) EXTREMITY-WHSC, Pack Number DYNJ910077A; u) KNEE ARTHROSCOPY-WHSC, Pack Number DYNJ910080A; v) MICRODISC, Pack Number DYNJ910370; w) MTHZ TOTAL HIP, Pack Number DYNJC9311L
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
a) CDS981510L: UDI/DI case 40195327195435, UDI/DI each 10195327195434, Lot Numbers: 23LDB386, 23LDB793; b) DYNJ900720D: UDI/DI case 40193489951593, UDI/DI each 10193489951592, Lot Numbers:23LBT897; c) DYNJ900720F: UDI/DI case 40195327635337, UDI/DI each 10195327635336, Lot Numbers: 24CBC838, 24EBH235; d) DYNJ908090C: UDI/DI case 40195327507245, UDI/DI each 10195327507244, Lot Numbers: 24CBH612, 24FBA023; e) DYNJ908091B: UDI/DI case 40195327507023, UDI/DI each 10195327507022, Lot Numbers: 24CBK626; f) DYNJ908092C: UDI/DI case 40195327507030, UDI/DI each 10195327507039, Lot Numbers: 23LBU932, 24ABT063, 24CBP714; g) DYNJ908095B: UDI/ DI case 40195327507184, UDI/DI each 10195327507183, Lot Numbers: 24ABO437, 24ABW989, 24EBH236; h) DYNJ908097A: UDI/DI case 40195327507078, UDI/DI each 10195327507077, Lot Numbers:23LBV779, 24CBU878; i) DYNJ908102B: UDI/DI case 40195327507191, UDI/DI each 10195327507190, Lot Numbers: 24BBI407; j) DYNJ908103B: UDI/DI case 40195327507108, UDI/DI each 10195327507107, Lot Numbers: 24CBO655; k) DYNJ908110B: UDI/DI case 40195327507160, UDI/DI each 10195327507169, Lot Numbers: 24ABS361, 24CBE409; m) DYNJ908425: UDI/DI case 40195327072613, UDI/DI each 10195327072612, Lot Numbers: 24ABL204; n) DYNJ908660A: UDI/DI case 40195327219674, UDI/DI each 10195327219673, Lot Numbers: 23LMC113, 24AME301, 24AMJ018, 24CMH075; o) DYNJ909734: UDI/DI case 40195327464708, UDI/DI each 10195327464707, Lot Numbers: 24ABA222, 24CBO720; p) DYNJ909738: UDI/DI case 40195327464739, UDI/DI each 10195327464738, Lot Numbers: 24ABY029, 4EBH232; q) DYNJ909740: UDI/DI case 40195327464753, UDI/DI each 10195327464752, Lot Numbers: 24BBK598; r) DYNJ910074A: UDI/DI case 40195327639281, UDI/DI each 10195327639280, Lot Numbers: 24CBA353; s) DYNJ910077: UDI/DI case 40195327568383, UDI/DI each 10195327568382, Lot Numbers: 23LBS081; t) DYNJ910077A: UDI/DI case 40195327639342, UDI/DI each 10195327639341, Lot Numbers: 24CBM378, 24DBC165; u) DYNJ910080A: UDI/DI case 40195327639304, UDI/DI each 10195327639303, Lot Numbers: 24BBS375; v) DYNJ910370: UDI/DI case 40195327654536, UDI/DI each 10195327654535, Lot Numbers: 24DDB322, 24EDA091; w) DYNJC9311L: UDI/DI case 40195327464777, UDI/DI each 10195327464776, Lot Numbers: 24CDC175
Devices named in this recall
22 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ900720DMedline Industries, Inc. · Orthopedic tray4 recalls0 reports
- Medline Industries, Inc. DYNJ908097AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908425Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. CDS981510LMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908660AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ909734Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ909738Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ909740Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJC9311LMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908091BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908092CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908103BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908110BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908095BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908102BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ908090CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ910077Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ900720FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ910074AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ910080AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ910077AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ910370Medline Industries, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95128).
- Z-3105-2024 · Open, ClassifiedMedline Semi-rigid suction liners labeled as follows: a) SEMI RIGID SUCTION LINER, REF OR515; b) SEMI RIGID SUCTION LINER WITH TUBING, REF OR526K; c) SEMI RIGID SUCTION LINER WIT…
- Z-3106-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA0311C; b) ANESTHESIA CIRCUIT-LF, Pack Number DYNJAA03…
- Z-3107-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I; b) ENDO KIT, Pack Number DYKE1558B;…
- Z-3108-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: DENTAL, Pack Number DYNJ909451A
- Z-3109-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) BAPTIST FLOYD BRONCHOSCOPY, Pack Number DYKE1456D; b) ENT MINOR, Pack Number DYNJ908099B; c) …
- Z-3110-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR PROCEDURE, Pack Number CDS981509C; b) GENERAL LAPAROSCOPY CDS, Pack Number CDS981512J; …
- Z-3111-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO BRONCH PACK, Pack Number DYKE1355; b) ENDO BASIC KIT, Pack Number DYKE1356; c) ENDOSCOPY…
- Z-3112-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1350D; b) ENDO KIT, Pack Number DYKE1514B
- Z-3113-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR BREAST, Pack Number DYNJ908104B; b) PLASTIC MAJOR, Pack Number DYNJ908106B; c) ABDOMI…
- Z-3114-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1092B; b) VIDEO BRONCHOSCOPY TRAY, Pack Number DYKE1666; c) SUCTIO…
- Z-3116-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: ANES TURNOVER KIT-LF, Pack Number DYNJAA10338J
- Z-3117-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: KIT NEURO SHUNT ST LOUIS CHILD, Pack Number DYNJ907829B
- Z-3118-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: MTHZ GYN LAPAROSCOPY, Pack Number DYNJ909739
- Z-3119-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: MERCY TIFFIN C-SECTION, Pack Number DYNJ909732
- Z-3120-2024 · Open, ClassifiedMedline convenience kits containing semi-rigid suction liners labeled as follows: a) AMB VITRECTOMY, Pack Number DYNJ59921B; b) AMB VITRECTOMY, Pack Number DYNJ59921BH; c) AMB …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.