MiniMed 620G Insulin Pump, REF: MMT-1510, MMT-1710, MMT-1750 recallZ-0002-2025 · Medtronic MiniMed, Inc. · initiated Jul 31, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
FDA root cause category: Device Design
What the company is doing
On 7/31/24, Safety Alert notices were sent to customers, healthcare professionals, distributors who were told the following: 1) Always pay attention to alerts and alarms displayed on your pump. - Low Battery Pump alert will display when your pump has up to 10 hours of battery life left. - Replace Battery alert will display when the pump has less than 30 minutes of battery life left. If the battery is not replaced within 10 minutes, a siren will sound and repeat once every minute until the Replace Battery Now alarm is displayed. At that time, the pump will stop insulin delivery. If the battery is not replaced within 10 minutes of this alarm, a siren will sound, and the pump will shut down. 2) If your pump has been dropped, bumped, or has experienced physical impact, pay special attention to any pump alerts and alarms, including low battery alerts, as they may occur earlier than expected. 3) Pump's battery level is on the status bar located on the pump home screen. 4) Refer to the pump user guide for instructions on how to replace the battery. 5) If you notice any significant changes in battery life or need additional troubleshooting assistance, the firm's Technical Support line for further help. 6) Always have extra new AA lithium or alkaline batteries or fully charged NiMH batteries available, along with your other emergency kit supplies. 7) Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/battery-statusacknowledgement In addition, distribution partners and service providers were provided with a patient letter they could use to inform customers/pump users. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 800-646-4633, option 1.
Recall details
- Recall number
- Z-0002-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2024
- Posted by the FDA
- Oct 3, 2024
- Recalling firm
- Medtronic MiniMed, Inc.
- Quantity
- 234,360
- Product code
- OYC
- Distribution
- US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide
Product description
MiniMed 620G Insulin Pump, REF: MMT-1510, MMT-1710, MMT-1750; MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 640G Insulin Pump, REF: MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, MMT-1752; MiniMed 720G Insulin Pump, REF: MMT-1809, MMT-1810, MMT-1817, MMT-1818, MMT-1859, MMT-1860, MMT-1867, MMT-1868 MiniMed 740G Insulin Pump, REF: MMT-1811, MMT-1812, MMT-1861, MMT-1862
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
REF/UDI-DI(GTIN): MMT-1710/00643169568419; MMT-1750/00763000253189, 00763000375461, 00643169721319, 00643169559745, 763000253189; MMT-1714/00763000253059, 00763000074845, 00643169782396, 763000253059, 643169782396; MMT-1715/ 00763000316631, 763000190439, 763000316631; MMT-1754/00763000253066, 00643169976344, 00763000315719, 00763000381912, 763000253066; MMT-1755/643169752726, 643169873841, 763000190446, 763000101527, 763000317591, 00763000317591, 763000367091, 763000179632, 763000317621, 643169878648, 763000157043, 76300031759103, 00643169752726, 76300031759101, 76300031762103, 76300031762101, 00763000101527, 00763000190446; MMT-1711/00643169662568, 00763000317119, 00763000317089, 00763000404550, 00763000367039, 00763000317102, 00643169662544, 00643169739468, 00643169554931, 00763000192136, 00763000068998, 00763000198596, 00643169662575, 00763000065195, 00763000258337, 00643169999763, 00643169739451, 00643169739475, 00643169739444, 00643169739482, 00763000192211, 00643169729353, 00643169554924, 00763000013066, 00643169554948, 00643169554917, 00643169966284, 00643169629813, 00643169662582, 00643169554955, 00643169966277, 00643169966260, 00643169966307, 00643169966291, 00643169662551, 643169890039, 763000367039, 763000317089, 763000192211, 763000317119, 643169739444, 643169554931, 643169739468; MMT-1712/00763000367046, 00763000317157, 00763000317195, 00763000404567, 00763000192143, 00763000317188, 00763000317140, 00763000065201, 00763000317164, 00643169739512, 00763000069001, 00763000258344, 00643169662636, 00643169999770, 00643169577664, 00763000155346, 00763000198602, 00763000192228, 00643169729360, 00643169739499, 00643169739505, 00643169577688, 00643169739536, 00643169662612, 00643169739529, 00643169629820, 00643169577701, 00643169662599, 00643169577695, 00643169966338, 00643169966352, 00643169966314, 00643169966321, 00643169577671, 00643169662629, 00643169662605, 00643169966345, 763000317195, 763000317140, 763000192228, 643169739512, 643169577688, 643169999770, 763000317157, 763000367046, 643169739505; MMT-1751/00763000194246, 00763000319489, 00763000319359, 00763000319410, 00763000319458, 00763000319519, 00643169521179, 00763000324698, 00763000319366, 00763000194338, 00763000319434, 00763000319533, 00763000324681, 00643169999787, 00763000319380, 00763000319427, 00763000319526, 00763000402686, 00763000319397, 00763000015534, 00763000319342, 00763000402693, 00763000192198, 00763000319373, 00763000319496, 00763000194291, 00643169672239, 00763000002855, 00763000402709, 00763000194352, 00763000015558, 00763000002787, 00763000253288, 00763000165185, 00763000165123, 00643169682849, 00763000015510, 00763000319441, 00643169672253, 00763000319403, 00763000165109, 00763000015480, 00643169683143, 00763000194314, 00643169672246, 00763000194475, 00763000165130, 00763000165154, 00763000083441, 00763000165093, 00763000165116, 00763000165147, 00763000074814, 00763000194345, 00763000194277, 00763000165222, 00763000194307, 00763000165215, 00763000002824, 00643169682245, 00643169682603, 00763000194239, 00763000194253, 00763000194321, 00763000194222, 00643169682269, 00763000002817, 00643169682061, 00763000194260, 00763000015572, 00763000002831, 00643169682337, 00763000015497, 00763000002763, 00643169682085, 00763000015589, 00643169585270, 00763000015503, 00763000015596, 00643169683129, 00643169672277, 00643169522046, 00763000015541, 00763000015565, 00643169682313, 00643169643680, 00643169643413, 00643169682535, 00643169682801, 00643169681996, 00643169521995, 00643169521445, 00643169585294, 00643169682054, 00643169682047, 00643169672260, 00763000002770, 00763000002848, 00763000002879, 00763000002886, 00763000002756, 00643169522060, 00643169682870, 00763000002794, 00763000015459, 00763000002800, 00643169521247, 00643169585256, 00763000002862, 00643169683075, 00643169682351, 00643169522022, 00643169629653, 00643169643710, 00643169682894, 00643169682641, 00643169682009, 00643169682108, 00643169681972, 00643169682627, 00643169683181, 00643169682580, 00643169683167, 00643169682566, 00643169682832, 00643169682023, 00643169682016, 00643169682252, 00643169682917, 00643169682375, 00643169681989, 00643169682078, 00643169683136, 00643169682276, 00643169682030, 00643169682511, 00643169682788, 00643169683051, 00643169682856, 00643169682795, 00643169644274, 00643169643444, 00643169682306, 00643169643482, 00643169683068, 00643169682818, 00643169521520, 00643169643758, 00643169629691, 00643169629677, 00643169643666, 00643169683082, 00643169683105, 00643169682290, 00643169521797, 00643169682825, 00643169682542, 00643169682559, 00643169683099, 00643169521209, 00643169682283, 00643169643956, 00643169643772, 00643169629660, 00643169682528, 00643169521728, 00643169629684, 00643169521544, 00643169644205, 00643169643529, 00643169643505, 00643169521155, 00643169644021, 00643169643468, 00643169644298, 00643169644335, 00643169643987, 00643169644007, 00643169644229, 00643169522084, 00643169521261, 00643169521810, 00643169643390, 00643169521421, 00643169644250, 00643169521568, 00643169643963, 00643169643932, 00643169644045, 00643169521438, 00643169521506, 00643169643734, 00643169521162, 00643169643703, 00643169644236, 00643169643741, 00643169521704, 00643169521476, 00643169644281, 00643169521988, 00643169644243, 00643169521285, 00643169521773, 00643169521971, 00643169522107, 00643169643949, 00643169643697, 00643169521834, 00643169643420, 00643169521223, 00643169643512, 00643169522008, 00643169585287, 00643169521216, 00643169521186, 00643169522015, 00643169521759, 00643169521193, 00643169522053, 00643169521742, 00643169521230, 00643169521780, 00643169521513, 00643169521766, 00643169521469, 00643169585263, 00643169521254, 00643169522077, 00643169521537, 00643169521803, 00643169521711, 00643169643123, 00643169520967, 00643169643253, 00643169520882, 00643169520905, 00643169643208, 00643169643192, 00643169520950, 00643169643178, 00643169643130, 00643169643215, 00643169643154, 00643169643161, 00643169643147, 00643169520974, 00643169520912, 00643169520929, 00643169521018, 00643169520998, 00643169643185, 00643169520943, 00643169520936, 00643169520981, 00643169520899, 00643169643239, 00643169643222, 00643169668829, 00643169574410, 00643169574441, 00643169574434, 00643169574427, 00643169574458, 00643169521735, 00643169521452, 00763000316525; MMT-1752/00763000102593, 00763000318307, 00763000393809, 00643169682474, 00643169682382, 00763000194666, 00763000324704, 00763000194444, 00763000165277, 00763000194383, 00763000319571, 00763000165246, 00763000115968, 00643169521339, 00763000319540, 00763000319274, 00763000194468, 00763000319564, 00763000192204, 00763000318291, 00643169682412, 00763000194451, 00763000165307, 00643169682436, 00763000194390, 00643169643574, 00763000194376, 00643169643581, 00643169521346, 00763000165345, 00763000319618, 00763000319311, 00763000319267, 00763000194369, 00763000165284, 00763000319243, 00763000165338, 00763000319304, 00763000165260, 00763000074821, 00763000194413, 00763000319595, 00763000194406, 00643169682429, 00763000319328, 00763000194420, 00763000319298, 00763000165314, 00763000319557, 00763000319229, 00643169682498, 00763000319588, 00643169999800, 00763000194437, 00643169643628, 00643169521384, 00643169721289, 00643169521292, 00643169643642, 00643169682405, 00643169682443, 00643169521360, 00643169596443, 00643169682450, 00643169596405, 00643169643536, 00643169596382, 00643169521315, 00643169521407, 00643169643567, 00643169643598, 00643169643550, 00643169522305, 00643169521322, 00643169521353, 00643169643260, 00643169643352, 00643169643895, 00643169521612, 00643169521667, 00643169521896, 00643169644120, 00643169644113, 00643169644076, 00643169643314, 00643169643307, 00643169521025, 00643169643802, 00643169644397, 00643169522152, 00643169521117, 00643169644434, 00643169644342, 00643169643871, 00643169644380, 00643169522114, 00643169521049, 00643169521063, 00643169552388, 00643169521872, 00643169521087, 00643169521629, 00643169521575, 00643169522169, 00643169522176, 00643169522145, 00643169643321, 00643169521681, 00643169522138, 00643169521070, 00643169521889, 00643169521841, 00643169521865, 00643169521131, 00643169522206, 00643169521933, 00643169643840, 00643169521605, 00643169521957, 00643169643857, 00643169643864, 00643169644373, 00643169643291, 00643169644182, 00643169643826, 00643169643338, 00643169644168, 00643169644458, 00643169643918, 00643169643833, 00643169644106, 00643169643376, 00643169644410, 00643169644403, 00643169682115, 00643169682153, 00643169683013, 00643169682924, 00643169682962, 00643169682160, 00643169682658, 00643169682696, 00643169683198, 00643169683242, 00643169683280, 00643169683303, 00643169682221, 00643169682207, 00643169682146, 00643169682979, 00643169682184, 00643169682177, 00643169682764, 00643169682702, 00643169682740, 00643169682672, 00643169683235, 00643169683228, 00643169682139, 00643169682955, 00643169683211, 00643169683037, 00643169682948, 00643169683266, 00643169682993, 00643169682689, 00643169682719, 00643169682214, 00643169683259, 00643169682986, 00643169682191, 00763000002633, 00763000002749, 00643169966246, 00763000002671, 00763000002664, 00763000002725, 00643169966253, 00643169668874, 00763000002732, 00763000002695, 00763000002688, 00763000002657, 00643169999817, 00763000002701, 00763000002718, 00763000002640, 00763000015381, 00763000015367, 00763000015435, 00763000015336, 00763000015374, 00763000015398, 00763000015411, 00763000015428, 00763000015350, 00763000015343, 00763000015404, 00763000015442, 00763000069025, 00763000319281, 00763000347536, 00643169521902, 00643169521636, 00643169521056, 00643169522183, 00643169521094, 00643169522381, 00643169522268, 00643169522343, 00643169521100, 00643169521919, 00643169521599, 00643169521643, 00763000319236, 763000255633; MMT-1809/00763000484644, 00763000541330, 00763000769413; MMT-1810/00763000484651, 00763000541347, 00763000769420, 00763000496302; MMT-1859/00763000505196, 00763000505257, 00763000505219, 00763000505318, 00763000505226, 00763000505288, 00763000505271, 00763000505233, 00763000505295; MMT-1860/00763000505387, 00763000505332, 00763000505370, 00763000531171, 00763000505394, 00763000505400, 00763000854768, 00763000505356, 00763000505349, 00763000505325; MMT-1867/00763000638016; MMT-1811/00763000629984, 00763000769451; MMT-1812/00763000484637, 00763000629991, 00763000769468; MMT-1861/00763000504922, 00763000504977, 00763000559557, 00763000504915, 00763000505028, 00763000504960, 00763000504984, 00763000504892, 00763000504861, 00763000504991, 00763000504953, 00763000504908, 00763000504946, 00763000505042, 00763000505004, 00763000505059, 00763000851224, 00763000851217, 00763000504939, 00763000504878; MMT-1862/00763000505080, 00763000505127, 00763000505110, 00763000505158, 00763000505172, 00763000505073, 00763000505066, 00763000505134, 00763000505103, 00763000505189, 00763000505141, 00763000505097, 00763000505165. All serial numbers.
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls8 deaths781 reports
- MiniMed® 630G MMT-1755KRMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control5 recalls13 reports
- MiniMed® 630G MMT-1755KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend5 recalls0 reports
- MiniMed® 630G MMT-1755KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls44 reports
- MiniMed®630G MMT-1755KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls11 reports
- MiniMed® 630G MMT-1755KMKMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls0 reports
- MiniMed® 630G MMT-1755KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls0 reports
- MiniMed™ 630G MMT-1755KRMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control4 recalls0 reports
- MiniMed® 630G MMT-1755KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend3 recalls0 reports
- MiniMed™ 630G MMT-1755KMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls0 reports
- MiniMed™ 630G MMT-1755KRMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95121).
- Z-0001-2025 · Open, ClassifiedMiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851
- Z-0003-2025 · Open, ClassifiedMiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page