MiniMed 630GRecalls & reported problems across all models
Medtronic Minimed, Inc. · 23 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
6 open recalls name at least one MiniMed 630G model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 65,972
- Deaths reported
- 232
- Injuries reported
- 17,368
- Recalls
- 6
We found 23 models of MiniMed 630G by Medtronic Minimed, Inc. in the FDA's device registry (GUDID). Together they are named in 65,972 adverse event reports in the FDA's MAUDE database, received between May 2, 2017 and Jul 3, 2025 — classified as 232 deaths, 17,368 injuries, 48,372 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (14,961 reports).
6 FDA recall records name at least one of these models by its device identifier, 6 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
MiniMed 630G recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Feb 13, 2026Open, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1748-2026 · Root cause: Device Design · names 9 of these models · Recall details · FDA record ↗
- Feb 13, 2026Open, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1740-2026 · Root cause: Device Design · names 10 of these models · Recall details · FDA record ↗
- Jul 31, 2024Open, Classified
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Z-0002-2025 · Root cause: Device Design · names 12 of these models · Recall details · FDA record ↗
- Jul 31, 2024Open, Classified
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Z-0001-2025 · Root cause: Device Design · names 21 of these models · Recall details · FDA record ↗
- Sep 20, 2022Open, Classified
Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
Z-0193-2023 · Root cause: Component change control · names 7 of these models · Recall details · FDA record ↗
- May 23, 2022Open, Classified
Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
Z-1353-2022 · Root cause: Device Design · names 10 of these models · Recall details · FDA record ↗
Most reported MiniMed 630G problems
Device problems
- Patient Device Interaction Problem14,961
- Adverse Event Without Identified Device or Use Problem10,613
- No Display/Image8,287
- No Apparent Adverse Event7,156
- Material Integrity Problem5,671
- Device Displays Incorrect Message5,604
- Break5,214
- Power Problem4,533
Patient problems
- No Consequences Or Impact To Patient34,371
- Hyperglycemia18,223
- No Clinical Signs, Symptoms or Conditions9,010
- Hypoglycemia4,964
- Elevated Ketones/ Diabetic Ketoacidosis1,024
- Nausea404
- Vomiting365
- Blood Loss252
MiniMed 630G models
23 models, most reported first. Each model has its own page with its reports, recalls and device details.
- MiniMed® 630G MMT-1715KArtificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- MiniMed® 630G MMT-1715KMArtificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KAutomated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed®630G MMT-1715KArtificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
- MiniMed® 630G MMT-1715KLAutomated insulin dosing , threshold suspend2 recalls10 deaths2,492 reports
- MiniMed® 630G MMT-1715KLArtificial pancreas device system, threshold suspend4 recalls4 deaths1,294 reports
- MiniMed® 630G MMT-1715KRAutomated insulin dosing , threshold suspend9 deaths1,195 reports
- MiniMed® 630G MMT-1715KAutomated insulin dosing device system, single hormonal control3 recalls8 deaths781 reports
- MiniMed® 630G MMT-1715KLAutomated insulin dosing device system, single hormonal control2 recalls266 reports
- MiniMed® 630G MMT-1715KLAutomated insulin dosing , threshold suspend3 recalls191 reports
- MiniMed® 630G MMT-1755KRAutomated insulin dosing device system, single hormonal control5 recalls13 reports
- MiniMed® 630G MMT-1755KMArtificial pancreas device system, threshold suspend5 recalls0 reports
- MiniMed® 630G MMT-1755KArtificial pancreas device system, threshold suspend4 recalls44 reports
- MiniMed®630G MMT-1755KArtificial pancreas device system, threshold suspend4 recalls11 reports
- MiniMed® 630G MMT-1755KMKArtificial pancreas device system, threshold suspend4 recalls0 reports
- MiniMed® 630G MMT-1755KAutomated insulin dosing , threshold suspend4 recalls0 reports
- MiniMed™ 630G MMT-1755KRAutomated insulin dosing device system, single hormonal control4 recalls0 reports
- MiniMed® 630G MMT-1715KLAutomated insulin dosing , threshold suspend3 recalls30 reports
- MiniMed® 630G MMT-1755KAutomated insulin dosing , threshold suspend3 recalls0 reports
- MiniMed™ 630G MMT-1755KAutomated insulin dosing device system, single hormonal control3 recalls0 reports
- MiniMed™ 630G MMT-1755KRAutomated insulin dosing device system, single hormonal control3 recalls0 reports
- MiniMed® 630G MMT-1715KRAutomated insulin dosing device system, single hormonal control1 recall0 reports
- MiniMed® 630G MMT-1715KRAutomated insulin dosing , threshold suspend1 report
Recent MiniMed 630G adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER ALSO REPORTED THAT THE A POTENTIAL UNDER-DELIVERY. THE CUSTOMER WAS TREATED WITH AN INSULIN PUMP. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-397 AND MMT-1715K. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER STATED THAT THE AIR BUBBLES IN THE TUBING LIMIT INSULIN DELIVERY. THE CUSTOMER DID NOT ALLOW THE INSULIN TO WARM TO ROOM TEMPERATURE BEFORE FILLING THE RESERVOIR AND WAS ADVISED TO CHANGE THE INFUSION SET AND RESERVOIR. THE CUSTOMER ALSO REPORTED CONSISTENTLY HIGH BLOOD GLUCOSE LEVELS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-397. NO PRODUCT RETURN IS REQUIRED FOR MMT-1715K.
Patient problems: Hyperglycemia
MAUDE report 22394343 ↗MalfunctionJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 368 MG/DL WHILE TRAVELING BY PLANE, WHICH LATER REDUCED TO 138 MG/DL AFTER TREATMENT WITH THE INSULIN PUMP (SUBCUTANEOUS INSULIN INFUSION). BLOOD GLUCOSE VALUE AT THE TIME OF THE EVENT WAS 368 MG/DL. THE CUSTOMER WAS TREATED WITH INSULIN PUMP (SUBCUTANEOUS INSULIN INFUSION). THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, INCLUDING VERIFYING PUMP SETTINGS AND ADVISING THE CUSTOMER ON POTENTIAL CONTRIBUTING FACTORS SUCH AS SITE ISSUES, INSULIN QUALITY, AND TUBING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-1715K.
Patient problems: Hyperglycemia
MAUDE report 22394188 ↗MalfunctionJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE BLOOD GLUCOSE VALUE AT THE TIME OF THE EVENT WAS 300 MG/DL. THE CUSTOMER WAS TREATED WITH AN INSULIN PUMP (SUBCUTANEOUS INSULIN INFUSION). THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, AND IT WAS UNKNOWN WHETHER THE INSULIN PUMP WAS UTILIZED WITHIN 48 HOURS OF THE EVENT; ALSO, IT WAS UNKNOWN WHETHER THE AUTO MODE FEATURE WAS ACTIVE OR NOT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 22393101 ↗MalfunctionJul 3, 2025Power Problem; Power Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED REPLACE BATTERY NOW ALARMS WITHOUT PRIOR LOW BATTERY ALERTS, WITH THE BATTERY DEPLETING RAPIDLY AFTER BEING REPLACED. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED WHICH INCLUDED REVIEWING ALARM HISTORY, CONFIRMING THE USE OF NEW ALKALINE BATTERIES, AND VERIFYING THAT THE BATTERY CAP WAS NOT LOOSE, CRACKED, OR DAMAGED. THE ISSUE PERSISTED, AND THE PUMP WAS REPLACED. THE CUSTOMER WAS ADVISED TO MONITOR THE INSULIN PUMP AND CONTINUE USING IT UNTIL THE REPLACEMENT ARRIVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392982 ↗MalfunctionJul 3, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED BATTERY LEAKAGE AND RECURRING REPLACE BATTERY ALARMS WITH THEIR INSULIN PUMP. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER REPORTED A BATTERY FAILED ALARM, AND THE ISSUE WAS ATTRIBUTED TO BATTERY LEAKAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392976 ↗MalfunctionJul 3, 2025Moisture or Humidity Problem; Key or Button Unresponsive/not Working; Moisture or Humidity Problem; Key or Button Unresponsive/not Working
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED A BATTERY ISSUE AND A STUCK BUTTON ALARM WITH THEIR INSULIN PUMP. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER STATED THAT NO DAMAGE WAS FOUND ON THE BATTERY CAP AND BATTERY COMPARTMENT. ALSO, THE PUMP HAD BEEN EXPOSED TO MOISTURE AND CONTAINED FLUID IN THE BATTERY COMPARTMENT AND REQUIRED A REPLACEMENT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392502 ↗MalfunctionJul 3, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED CONDENSATION UNDER THE LCD SCREEN OF THE INSULIN PUMP AFTER OUTDOOR ACTIVITY. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, AND NO CRACKS OR EXTERNAL DAMAGE WERE FOUND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392214 ↗MalfunctionJul 3, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER OBSERVED A SMALL RED LIGHT UNDER THE BACK ARROW ON THE PUMP THAT REMAINED SOLID FOR APPROXIMATELY TWO MONTHS. THE LIGHT DID NOT FLASH DURING ALARMS, AND THE PUMP HAD NOT BEEN DROPPED OR BUMPED. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS NOT PERFORMED, AND THE PUMP WAS REPLACED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392202 ↗MalfunctionJul 3, 2025Battery Problem; Wireless Communication Problem; Battery Problem; Wireless Communication Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER ALSO REPORTED A REPLACE BATTERY NOW ALARM WITH A PRIOR LOW BATTERY ALERT, THE BATTERY LIFE WAS SHORTER THAN EXPECTED, THE SENSOR DID NOT CONNECT WITH THE INSULIN PUMP. BLOOD GLUCOSE VALUE AT THE TIME OF EVENT WAS 347 MG/DL. THE CUSTOMER WAS TREATED WITH INSULIN PUMP (SUBCUTANEOUS INSULIN INFUSION). THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED AND THE CUSTOMER WILL BE PROVIDE WITH AN INSULIN PUMP REPLACEMENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER DISCONTINUED USE OF THE INSULIN PUMP. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 22392007 ↗MalfunctionJul 3, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED A REPLACE BATTERY NOW ALARM WITHOUT A PRIOR LOW BATTERY WARNING. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT MMT-1715K. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22391994 ↗MalfunctionJul 3, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED RAPID BATTERY DEPLETION, WITH BATTERIES LASTING ONLY 36 HOURS DESPITE USING NEW ENERGIZER LITHIUM BATTERIES AND RECEIVING A REPLACEMENT BATTERY CAP. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, WHICH INCLUDED REVIEWING ALARM HISTORY, VERIFYING BATTERY USAGE, AND SUGGESTING ADJUSTMENTS TO SETTINGS SUCH AS BACKLIGHT TIMEOUT AND SCREEN MANAGEMENT, BUT THE ISSUE PERSISTED. THE CUSTOMER WAS ADVISED TO REVERT TO THEIR BACKUP PLAN PER HCP INSTRUCTIONS AND PLACE THE PUMP IN STORAGE MODE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22391930 ↗MalfunctionJun 27, 2025Break; Battery Problem; Break; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED TWO ISSUES WITH THE INSULIN PUMP: PHYSICAL DAMAGE WHERE THE CLIP ATTACHES AND RECURRING "REPLACE BATTERY NOW" ALARMS, REQUIRING BATTERY CHANGES EVERY OTHER DAY. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED AND THAT INCLUDED REVIEWING ALARM HISTORY, VERIFYING BATTERY USAGE, AND CONFIRMING THAT THE BATTERY CAP WAS NOT DAMAGED, BUT THE ISSUE PERSISTED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP, REVERT TO THEIR BACKUP PLAN PER HEALTHCARE PROFESSIONAL'S INSTRUCTIONS, AND INSERT NEW BATTERIES AS NEEDED UNTIL THE REPLACEMENT ARRIVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22346704 ↗MalfunctionJun 27, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED MULTIPLE "LOW BATTERY" AND "REPLACE BATTERY NOW" ALARMS WITHIN 20 HOURS, WITH THE BATTERY BECOMING HOT UPON REMOVAL. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED AND WHICH INCLUDED REVIEWING ALARM HISTORY, VERIFYING THE USE OF DURACELL BATTERIES, AND CONFIRMING THAT THE BATTERY CAP WAS NOT LOOSE, CRACKED, OR DAMAGED, BUT THE ISSUE PERSISTED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22346193 ↗MalfunctionJun 25, 2025Battery Problem; Material Integrity Problem; Battery Problem; Material Integrity Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED RECURRING "BATTERY FAILED" ALARM. THE CUSTOMER ALSO REPORTED A SMALL CRACK ALONG THE BACK OF THE PUMP EXTENDING FROM THE BATTERY CAP DOWN. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, WHICH INCLUDED REVIEWING ALARM HISTORY, CLEANING THE BATTERY COMPARTMENT, AND CONFIRMING NO DAMAGE OR CORROSION ON THE BATTERY CAP CONTACTS, BUT THE ISSUE PERSISTED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE, REVERT TO THE BACKUP PLAN PER THE HEALTHCARE PROFESSIONAL'S INSTRUCTIONS, AND PLACE THE PUMP IN STORAGE MODE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22326232 ↗MalfunctionJun 25, 2025Battery Problem; Battery Problem
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED REPEATED BATTERY FAILED ALARMS AND A LOW BATTERY INDICATOR DESPITE REPLACING THE BATTERY APPROXIMATELY 15 TIMES. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1715K. TROUBLESHOOTING WAS PERFORMED, AND THE CUSTOMER STATED THAT NO DAMAGE TO THE BATTERY CAP CONTACTS OR COMPARTMENT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22326123 ↗
Other Medtronic Minimed, Inc. products
- MiniMed 670G23 models · 141,169 reports · 4 recalls
- Paradigm REAL-Time Revel180 models · 49,838 reports · 2 recalls
- MiniMed 530G71 models · 48,745 reports · 1 recall
- MiniMed670G2 models · 28,871 reports · 3 recalls
- MiniMed630G2 models · 6,656 reports · 5 recalls
- MiniMed Reservoir6 models · 4,545 reports
- PARADIGM REAL-time95 models · 2,862 reports · 1 recall
- Guardian Sensor (3)11 models · 2,726 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered, combining the brand names used for this product line.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.