Panda recallZ-3081-2024 · Wipro GE Healthcare Private Ltd. · initiated Aug 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Wipro GE Healthcare Private Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
FDA root cause category: Process control
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/09/2024 by letter disseminated through a traceable means. The notice explained the safety issue and the potential for adverse impact to user or patient. The notice also provided information for the safe operation of the device until GE Healthcare is able to inspect and test the units. A GE Healthcare representative will contact the consignee to arrange for the correction. Questions: Contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-3081-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 9, 2024
- Posted by the FDA
- Sep 20, 2024
- Recalling firm
- Wipro GE Healthcare Private Ltd.
- Quantity
- 137 units
- Product code
- FMT
- FDA submissions
- K201628
- Distribution
- US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.
Product description
GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
1) M1112198-01045518, UDI/DI 00840682152709, Serial Numbers: PBW23420278SA, PBW23390175SA, PBW24241071SA, PBW24241072SA, PBW24241075SA, PBW23340090SA; 2) M1112198-01045579, UDI/DI 00840682152709, Serial Numbers: PBW24241078SA, PBW24241079SA, PBW24241063SA, PBW24241076SA; 3) M1112198-01047148, UDI/DI 00840682152709, Serial Numbers: PBW24241068SA, PBW24241070SA, PBW24241065SA, PBW24241064SA, PBW24241066SA, PBW24241067SA, PBW24241069SA; 4) M1112198-01047271, UDI/DI 00840682152716, Serial Numbers: PBW23492144SA, PBW23482075SA, PBW23492149SA, PBW24062087SA, PBW24062078SA, PBW24062081SA, PBW23492145SA, PBW24152231SA, PBW24152225SA, PBW24152229SA, PBW23482085SA, PBW24202294SA, PBW24202292SA, PBW24202293SA, PBW24212323SA, PBW24212324SA, PBW24152226SA, PBW24152227SA, PBW24052060SA, PBW23492142SA; 5) M1112198-01048731, UDI/DI 00840682152709, Serial Numbers: PBW24241087SA, PBW24241086SA, PBW24241085SA, PBW23320042SA, PBW23320041SA, PBW24241084SA, PBW23320035SA; 6) M1112198-01048740, UDI/DI 00840682152716, Serial Numbers: PBW23482093SA; 7) M1112198-01052386, UDI/DI 00840682152716, Serial Numbers: PBW23482100SA, PBW24202309SA, PBW24202313SA, PBW24202310SA, PBW24202315SA, PBW24202312SA, PBW24202314SA, PBW24202311SA; 8) M1112198-01054697, UDI/DI 00840682152716, Serial Numbers: PBW24132200SA; 9) M1112198-01068942, UDI/DI 00840682152716, Serial Numbers: PBW23472056SA, PBW24062069SA, PBW24112144SA, PBW24062072SA, PBW24112157SA, PBW24112149SA, PBW24112145SA, PBW24112158SA, PBW24112148SA, PBW24112147SA, PBW24112155SA, PBW24112150SA, PBW24112153SA, PBW24112156SA, PBW24112152SA, PBW24112154SA, PBW24112146SA, PBW24112151SA, PBW24142209SA; 10) M1112198-01084972, UDI/DI 00840682152716, Serial Numbers: PBW23462025SA, PBW23462038SA, PBW23462039SA, PBW23472051SA, PBW23472049SA, PBW23472050SA, PBW24132204SA, PBW24132203SA, PBW24132205SA, PBW24122189SA, PBW24122190SA; 11) M1112198-01090848, UDI/DI 00840682152716, Serial Numbers: PBW24122198SA, PBW24122195SA, PBW24122197SA; 12) M1112198-01093457, UDI/DI 00840682152716, Serial Numbers: PBW24202308SA, PBW24132201SA; 13) M1112198-01117766, UDI/DI 00840682152709, Serial Numbers: PBW24110390SA; 14) M1112198-01139893, UDI/DI 00840682152716, Serial Numbers: PBW23242002SA, PBW23242003SA, PBW23242000SA, PBW23242001SA, PBW23512155SA, PBW24062070SA, PBW24172255SA, PBW24172253SA, PBW24172254SA, PBW24172251SA, PBW24172250SA, PBW24212327SA, PBW24212328SA; 15) M1112198-01156745, UDI/DI 00840682152716, Serial Numbers: PBW24192258SA; 16) M1112198-01167230, UDI/DI 00840682152716, Serial Numbers: PBW23462035SA, PBW23462036SA, PBW23462040SA, PBW24112159SA; 17) M1112198-01177090, UDI/DI 00840682152716, Serial Numbers: PBW23482099SA; 18) M1112198-01183599, UDI/DI 00840682152716, Serial Numbers: PBW23492140SA, PBW23482090SA, PBW23482092SA, PBW23482089SA, PBW23482086SA, PBW23492139SA, PBW23482087SA, PBW23482088SA, PBW23482091SA, PBW23482103SA, PBW23482104SA, PBW23482105SA; 19) M1112198-01185372, UDI/DI 00840682152716, Serial Numbers: PBW23482079SA, PBW23482081SA, PBW23482082SA, PBW23482083SA, PBW23482078SA; 20) M1112198-01193801, UDI/DI 00840682152716, Serial Numbers: PBW23502154SA, PBW24172252SA; 21) M1112198-01200764, UDI/DI 00840682152716, Serial Numbers: PBW23472043SA; 22) M1112198-01245548, UDI/DI 00840682152716, Serial Numbers: PBW24062101SA; 23) M1112198-01253742, UDI/DI 00840682152716, Serial Numbers: PBW24052062SA, PBW24052063SA, PBW24052066SA, PBW24052065SA, PBW24052064SA; 24) M1112198-01274059, UDI/DI 00840682152716, Serial Numbers: PBW24192278SA, PBW24192279SA.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Panda i-Res Warmer 1.0Datex-Ohmeda, Inc. · WARMER, INFANT RADIANT1 recall15 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K201628 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Panda iRes WarmerDatex-Ohmeda, Inc. · WARMER, INFANT RADIANT34 reports
- Giraffe WarmerDatex-Ohmeda, Inc. · WARMER, INFANT RADIANT27 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95216).
- Z-3080-2024 · Open, ClassifiedGE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based o…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page