Centricity recallZ-3160-2024 · GE Healthcare · initiated Jul 23, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.
FDA root cause category: Software Design Change
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECITON notice to its consignees on 07/23/2023 via letter. The notice explained the issue, potential risk to the patient, and provided the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions below. Follow one of the below recommendations to ensure the latest addendum is being reviewed: 1. Use Microsoft Edge or Google Chrome browsers when launching ZFP Viewer. OR 2. If you cannot use the browsers listed in #1, to find the addendum / latest report, review all the reports available for the study in the Series Selector. Navigation to the series selection can be found in the User Manual under Section 2.3.3 of The Series Selector, with subheading as Open a report/document using series selector"." GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-3160-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 23, 2024
- Posted by the FDA
- Sep 17, 2024
- Recalling firm
- GE Healthcare
- Quantity
- 803 units
- Product code
- LLZ
- FDA submissions
- K131977
- Distribution
- Worldwide distribution.
Product description
GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 2110344-010; d) 2110344-011; e) 2110344-013; f) 2110344-019; g) 2110344-025; h) 2110344-029; i) 2110344-030; j) 2110344-032 ; Radiological image processing system
Code information (models, lots, serial numbers)
Show the full code information (55k characters)
UDI/DI: 00840682102988: a) 2110344-008, Installed Product ID Numbers: B54834-Centricity Universal Viewer Zero Footprint / ZFP-02394744, KMCH-Centricity Universal Viewer Zero Footprint / ZFP-01819848, 355920-Centricity Universal Viewer Zero Footprint / ZFP-20996307, 225717-Centricity Universal Viewer Zero Footprint / ZFP-19666806, RF0827-Centricity Universal Viewer Zero Footprint / ZFP-19460712; b) 2110344-009, Installed Product ID Numbers: 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372918, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348000, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348009, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348010, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348035, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-19632983, SA1008-Centricity Universal Viewer Zero Footprint / ZFP-20620864; c) 2110344-010, Installed Product ID Numbers: 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-00098227, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-00098228, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-14194760, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-16023936, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-16023938, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-16023939, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-16023940, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-16023941, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-16023942, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-17192653, 1472-19-Centricity Universal Viewer Zero Footprint / ZFP-22125883, 946739-Centricity Universal Viewer Zero Footprint / ZFP-14767595, 753891-Centricity Universal Viewer Zero Footprint / ZFP-21739897, 753891-Centricity Universal Viewer Zero Footprint / ZFP-19737302, JP-340499-Centricity Universal Viewer Zero Footprint / ZFP-01275434, RU8902-Centricity Universal Viewer Zero Footprint / ZFP-17775055, RU9383-Centricity Universal Viewer Zero Footprint / ZFP-16868184; d) 2110344-011, Installed Product ID Numbers: 3805-14-Centricity Universal Viewer Zero Footprint / ZFP-01564901, 3805-14-Centricity Universal Viewer Zero Footprint / ZFP-02098755, 3805-14-Centricity Universal Viewer Zero Footprint / ZFP-02098757, 3805-14-Centricity Universal Viewer Zero Footprint / ZFP-02098758, 4084-1-Centricity Universal Viewer Zero Footprint / ZFP-02301039, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036722; e) 2110344-013, Installed Product ID Numbers: RU1473-Centricity Universal Viewer Zero Footprint / ZFP-02331227, RU3152-Centricity Universal Viewer Zero Footprint / ZFP-02314018, RF1194-Centricity Universal Viewer Zero Footprint / ZFP-19716329, RU6265-Centricity Universal Viewer Zero Footprint / ZFP-20080267, RU1107-Centricity Universal Viewer Zero Footprint / ZFP-01620728, RU9227-Centricity Universal Viewer Zero Footprint / ZFP-20080351, RU8412-Centricity Universal Viewer Zero Footprint / ZFP-19727241, RU7361-Centricity Universal Viewer Zero Footprint / ZFP-20084480; f) 2110344-019, Installed Product ID Numbers: 104586-2-Centricity Universal Viewer Zero Footprint / ZFP-01299076, 8121-6-Centricity Universal Viewer Zero Footprint / ZFP-01808438, 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01367875, 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372919, 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372920, 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372921, 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372922, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-00556036, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-00556037, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347655, 444402-Centricity Universal Viewer Zero Footprint / ZFP-20091853, JP-340206-Centricity Universal Viewer Zero Footprint / ZFP-01275380, JP-690277-Centricity Universal Viewer Zero Footprint / ZFP-19665249; g) 2110344-025, Installed Product ID Numbers: 104440-1-Centricity Universal Viewer Zero Footprint / ZFP-23264377, RU1152-Centricity Universal Viewer Zero Footprint / ZFP-20579574; h) 2110344-029, Installed Product ID Numbers: 105906-2-Centricity Universal Viewer Zero Footprint / ZFP-02312614, 4653-1-Centricity Universal Viewer Zero Footprint / ZFP-00161865, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-02371135, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-02371136, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-02371137, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-02371138, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-02371139, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-01983569, 10414-1-Centricity Universal Viewer Zero Footprint / ZFP-02257545, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609719, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609721, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609722, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609724, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609725, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609727, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609728, 4049-1-Centricity Universal Viewer Zero Footprint / ZFP-01805026, 4049-1-Centricity Universal Viewer Zero Footprint / ZFP-02302924, 103944-3-Centricity Universal Viewer Zero Footprint / ZFP-01477454, 103944-3-Centricity Universal Viewer Zero Footprint / ZFP-02286832, 3805-14-Centricity Universal Viewer Zero Footprint / ZFP-02098759, 1275-6-Centricity Universal Viewer Zero Footprint / ZFP-00161288, 1275-6-Centricity Universal Viewer Zero Footprint / ZFP-00161686, 1275-6-Centricity Universal Viewer Zero Footprint / ZFP-00161687, 1275-6-Centricity Universal Viewer Zero Footprint / ZFP-18469046, 1275-6-Centricity Universal Viewer Zero Footprint / ZFP-20653788, 106710-1-Centricity Universal Viewer Zero Footprint / ZFP-01810611, 106710-1-Centricity Universal Viewer Zero Footprint / ZFP-14768105, 106710-1-Centricity Universal Viewer Zero Footprint / ZFP-14769712, 102753-1-Centricity Universal Viewer Zero Footprint / ZFP-01810676, 102753-1-Centricity Universal Viewer Zero Footprint / ZFP-01978730, 102753-1-Centricity Universal Viewer Zero Footprint / ZFP-01978731, 102753-1-Centricity Universal Viewer Zero Footprint / ZFP-01978732, 105525-1-Centricity Universal Viewer Zero Footprint / ZFP-02315655, 105525-1-Centricity Universal Viewer Zero Footprint / ZFP-21306260, 104015-1-Centricity Universal Viewer Zero Footprint / ZFP-02157305, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-01740249, 3212-8-Centricity Universal Viewer Zero Footprint / ZFP-14761752, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-02157261, 10547-2-Centricity Universal Viewer Zero Footprint / ZFP-01294055, 10547-2-Centricity Universal Viewer Zero Footprint / ZFP-02337580, 9915-2-Centricity Universal Viewer Zero Footprint / ZFP-01459010, 9915-2-Centricity Universal Viewer Zero Footprint / ZFP-02257259, 9866-2-Centricity Universal Viewer Zero Footprint / ZFP-00101400, 9866-2-Centricity Universal Viewer Zero Footprint / ZFP-00101401, 10832-3-Centricity Universal Viewer Zero Footprint / ZFP-21207910, 105468-1-Centricity Universal Viewer Zero Footprint / ZFP-02254840, 4289-3-Centricity Universal Viewer Zero Footprint / ZFP-01976550, 102193-2-Centricity Universal Viewer Zero Footprint / ZFP-14765252, 1609-1-Centricity Universal Viewer Zero Footprint / ZFP-02344894, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-22188687, 3719-9-Centricity Universal Viewer Zero Footprint / ZFP-00161362, 3719-9-Centricity Universal Viewer Zero Footprint / ZFP-00161368, 3066-1-Centricity Universal Viewer Zero Footprint / ZFP-00101769, 3066-1-Centricity Universal Viewer Zero Footprint / ZFP-00101770, 3719-1-Centricity Universal Viewer Zero Footprint / ZFP-01987326, 1331-24-Centricity Universal Viewer Zero Footprint / ZFP-02041554, 1331-24-Centricity Universal Viewer Zero Footprint / ZFP-02190655, 11673-4-Centricity Universal Viewer Zero Footprint / ZFP-01814396, 101647-1-Centricity Universal Viewer Zero Footprint / ZFP-02205362, 106558-2-Centricity Universal Viewer Zero Footprint / ZFP-01987441, 4204-1-Centricity Universal Viewer Zero Footprint / ZFP-01459008, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-02341586, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-02354814, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-02354815, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-02354816, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-15291321, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-15291322, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-15291332, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-15291333, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-15291334, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-15291335, 6901-11-Centricity Universal Viewer Zero Footprint / ZFP-16355110, 105461-1-Centricity Universal Viewer Zero Footprint / ZFP-02035195, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-01974088, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-02354817, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-02354818, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-02354819, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-02354820, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-15291319, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-15291320, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-17594364, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-19020773, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-19020774, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-20621990, 6901-12-Centricity Universal Viewer Zero Footprint / ZFP-20621993, 1847-3-Centricity Universal Viewer Zero Footprint / ZFP-02097811, 1847-3-Centricity Universal Viewer Zero Footprint / ZFP-02255598, 1847-2-Centricity Universal Viewer Zero Footprint / ZFP-00169967, 1847-2-Centricity Universal Viewer Zero Footprint / ZFP-00169968, 1847-2-Centricity Universal Viewer Zero Footprint / ZFP-00169969, 1696-1-Centricity Universal Viewer Zero Footprint / ZFP-20694814, 1696-1-Centricity Universal Viewer Zero Footprint / ZFP-20699361, 1696-1-Centricity Universal Viewer Zero Footprint / ZFP-02186950, 1696-1-Centricity Universal Viewer Zero Footprint / ZFP-02269625, 100729-1-Centricity Universal Viewer Zero Footprint / ZFP-01721608, 6051-Centricity Universal Viewer Zero Footprint / ZFP-02277661, 2892781_Stvincentssyd-Centricity Universal Viewer Zero Footprint / ZFP-02186467, 1698348_Alfred-Centricity Universal Viewer Zero Footprint / ZFP-02032817, 1586578_Nedlands-Centricity Universal Viewer Zero Footprint / ZFP-20034444, 123456-Centricity Universal Viewer Zero Footprint / ZFP-14310910, 423834-Centricity Universal Viewer Zero Footprint / ZFP-24370371, 423834-Centricity Universal Viewer Zero Footprint / ZFP-24370372, 423834-Centricity Universal Viewer Zero Footprint / ZFP-24370373, 333266-Centricity Universal Viewer Zero Footprint / ZFP-16482049, 13628-Centricity Universal Viewer Zero Footprint / ZFP-02336211, 104295-2-Centricity Universal Viewer Zero Footprint / ZFP-01286791, 631789-Centricity Universal Viewer Zero Footprint / ZFP-01286792, 631789-Centricity Universal Viewer Zero Footprint / ZFP-21268829, B0201-12-Centricity Universal Viewer Zero Footprint / ZFP-02031960, 2227709-Centricity Universal Viewer Zero Footprint / ZFP-20080947, 1301843-Centricity Universal Viewer Zero Footprint / ZFP-14430019, S540080-Centricity Universal Viewer Zero Footprint / ZFP-00043934, M44850-Centricity Universal Viewer Zero Footprint / ZFP-01817813, M41884-Centricity Universal Viewer Zero Footprint / ZFP-02345947, PF1005-Centricity Universal Viewer Zero Footprint / ZFP-02301061, PF1005-Centricity Universal Viewer Zero Footprint / ZFP-02301134, U1982906 / NC1076-Centricity Universal Viewer Zero Footprint / ZFP-01987930, 4625913X - HC0565-Centricity Universal Viewer Zero Footprint / ZFP-00545594, X9000399 - HC0399-Centricity Universal Viewer Zero Footprint / ZFP-01282975, HC1451-Centricity Universal Viewer Zero Footprint / ZFP-01288243, 1657844X - 518086-Centricity Universal Viewer Zero Footprint / ZFP-00158637, X9000245 - HC0245-Centricity Universal Viewer Zero Footprint / ZFP-01462020, 362144-Centricity Universal Viewer Zero Footprint / ZFP-02263754, PC0070-Centricity Universal Viewer Zero Footprint / ZFP-02042361, UHAM02-Centricity Universal Viewer Zero Footprint / ZFP-00059329, X9003387 - HHEI01-Centricity Universal Viewer Zero Footprint / ZFP-21310713, 434487-Centricity Universal Viewer Zero Footprint / ZFP-22169581, 145774-Centricity Universal Viewer Zero Footprint / ZFP-01814045, 1257845X - 434487-Centricity Universal Viewer Zero Footprint / ZFP-22203508, 0828550068-Centricity Universal Viewer Zero Footprint / ZFP-02094418, 828550053-Centricity Universal Viewer Zero Footprint / ZFP-02253600, 434519-Centricity Universal Viewer Zero Footprint / ZFP-20677457, A1892038-Centricity Universal Viewer Zero Footprint / ZFP-02093306, A56306-Centricity Universal Viewer Zero Footprint / ZFP-02180921, A56600-Centricity Universal Viewer Zero Footprint / ZFP-01454929, A6700602-Centricity Universal Viewer Zero Footprint / ZFP-02178761, A60858-Centricity Universal Viewer Zero Footprint / ZFP-02286910, JP-380108-Centricity Universal Viewer Zero Footprint / ZFP-02255381, JP-340255-Centricity Universal Viewer Zero Footprint / ZFP-01974303, JP-280245-Centricity Universal Viewer Zero Footprint / ZFP-02255463, JP-280250-Centricity Universal Viewer Zero Footprint / ZFP-02360544, JP-140140-Centricity Universal Viewer Zero Footprint / ZFP-02154910, JP-140051-Centricity Universal Viewer Zero Footprint / ZFP-02099295, JP-290033-Centricity Universal Viewer Zero Footprint / ZFP-01275435, JP-290033-Centricity Universal Viewer Zero Footprint / ZFP-01275437, JP-290033-Centricity Universal Viewer Zero Footprint / ZFP-01275438, JP-250378-Centricity Universal Viewer Zero Footprint / ZFP-02186431, JP-090905-Centricity Universal Viewer Zero Footprint / ZFP-02041046, JP-130422-Centricity Universal Viewer Zero Footprint / ZFP-01980318, JP-130527-Centricity Universal Viewer Zero Footprint / ZFP-02093919, JP-130626-Centricity Universal Viewer Zero Footprint / ZFP-02043333, JP-130761-Centricity Universal Viewer Zero Footprint / ZFP-02248293, KW1102-Centricity Universal Viewer Zero Footprint / ZFP-02191178, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02156484, 60020_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02307677, 60020_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02291060, 709434-Centricity Universal Viewer Zero Footprint / ZFP-14765518, SA1075-Centricity Universal Viewer Zero Footprint / ZFP-01457247, UP285A-Centricity Universal Viewer Zero Footprint / ZFP-01815172, GPX458-Centricity Universal Viewer Zero Footprint / ZFP-01980394, GPX459-Centricity Universal Viewer Zero Footprint / ZFP-01980471, GPQ131-Centricity Universal Viewer Zero Footprint / ZFP-02254265, 407237-Centricity Universal Viewer Zero Footprint / ZFP-02290167, 529010-Centricity Universal Viewer Zero Footprint / ZFP-01824677, 34433-Centricity Universal Viewer Zero Footprint / ZFP-01294788, 16798599-Centricity Universal Viewer Zero Footprint / ZFP-01475999, 34435-Centricity Universal Viewer Zero Footprint / ZFP-00545336, 34449-Centricity Universal Viewer Zero Footprint / ZFP-22226438, 80674-Centricity Universal Viewer Zero Footprint / ZFP-01425411, 80674-Centricity Universal Viewer Zero Footprint / ZFP-21318253; i) 2110344-030, Installed Product ID Numbers: 4084-1-Centricity Universal Viewer Zero Footprint / ZFP-19600711, 9862-1-Centricity Universal Viewer Zero Footprint / ZFP-00100263, 9862-1-Centricity Universal Viewer Zero Footprint / ZFP-14765956, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096091, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348002, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090, 1503588-Centricity Universal Viewer Zero Footprint / ZFP-14549617, 6891796-Centricity Universal Viewer Zero Footprint / ZFP-19631000, 306184-Centricity Universal Viewer Zero Footprint / ZFP-19620282, 324227-Centricity Universal Viewer Zero Footprint / ZFP-02040656, 320092-Centricity Universal Viewer Zero Footprint / ZFP-19675475, 329997-Centricity Universal Viewer Zero Footprint / ZFP-19732482, 3440585-Centricity Universal Viewer Zero Footprint / ZFP-02286138, 3440585-Centricity Universal Viewer Zero Footprint / ZFP-20092101, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904827, JP-380123-Centricity Universal Viewer Zero Footprint / ZFP-20101283, JP-400021-Centricity Universal Viewer Zero Footprint / ZFP-19649615, JP-030135-Centricity Universal Viewer Zero Footprint / ZFP-01275425, JP-420052-Centricity Universal Viewer Zero Footprint / ZFP-14761038, JP-420052-Centricity Universal Viewer Zero Footprint / ZFP-15111121, JP-420052-Centricity Universal Viewer Zero Footprint / ZFP-15111126, JP-272900-Centricity Universal Viewer Zero Footprint / ZFP-02041843, JP-270263-Centricity Universal Viewer Zero Footprint / ZFP-01275411, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-01299106, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 328816-Centricity Universal Viewer Zero Footprint / ZFP-17805275, 4879575-Centricity Universal Viewer Zero Footprint / ZFP-15743070, U-4J5U54-Centricity Universal Viewer Zero Footprint / ZFP-01455930, RU7579-Centricity Universal Viewer Zero Footprint / ZFP-18957266, 41400-Centricity Universal Viewer Zero Footprint / ZFP-02291025; j) 2110344-032, Installed Product ID Numbers: 3915-10-Centricity Universal Viewer Zero Footprint / ZFP-02360681, 3915-10-Centricity Universal Viewer Zero Footprint / ZFP-18867313, 4653-1-Centricity Universal Viewer Zero Footprint / ZFP-00161866, 4653-1-Centricity Universal Viewer Zero Footprint / ZFP-19618935, 4653-1-Centricity Universal Viewer Zero Footprint / ZFP-21780800, 103710-1-Centricity Universal Viewer Zero Footprint / ZFP-02308031, 104586-2-Centricity Universal Viewer Zero Footprint / ZFP-23066609, 4415-6-Centricity Universal Viewer Zero Footprint / ZFP-01810693, 2195-15-Centricity Universal Viewer Zero Footprint / ZFP-01291411, 2195-15-Centricity Universal Viewer Zero Footprint / ZFP-00165673, 2195-15-Centricity Universal Viewer Zero Footprint / ZFP-00168142, 2195-15-Centricity Universal Viewer Zero Footprint / ZFP-00168143, 2195-15-Centricity Universal Viewer Zero Footprint / ZFP-01448429, 2195-10-Centricity Universal Viewer Zero Footprint / ZFP-20999747, 105281-1-Centricity Universal Viewer Zero Footprint / ZFP-01820425, 100256-3-Centricity Universal Viewer Zero Footprint / ZFP-01810795, 100256-3-Centricity Universal Viewer Zero Footprint / ZFP-19604803, 100256-3-Centricity Universal Viewer Zero Footprint / ZFP-19604804, 100256-3-Centricity Universal Viewer Zero Footprint / ZFP-19604805, 100256-3-Centricity Universal Viewer Zero Footprint / ZFP-19604806, 4415-8-Centricity Universal Viewer Zero Footprint / ZFP-20274847, 4415-2-Centricity Universal Viewer Zero Footprint / ZFP-19609729, 4415-7-Centricity Universal Viewer Zero Footprint / ZFP-20274843, 8142-4-Centricity Universal Viewer Zero Footprint / ZFP-02397428, 8142-4-Centricity Universal Viewer Zero Footprint / ZFP-19715369, 8142-4-Centricity Universal Viewer Zero Footprint / ZFP-19715370, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-00099694, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-01294051, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-02350740, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-16265092, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20019523, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20019524, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20019525, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20019526, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20019527, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20036734, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20036735, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20036736, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-20036737, 105953-2-Centricity Universal Viewer Zero Footprint / ZFP-02345969, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-22268979, 100620-23-Centricity Universal Viewer Zero Footprint / ZFP-22268125, 103696-8-Centricity Universal Viewer Zero Footprint / ZFP-16868149, 4420-6-Centricity Universal Viewer Zero Footprint / ZFP-01810727, 4420-6-Centricity Universal Viewer Zero Footprint / ZFP-20032680, 4420-6-Centricity Universal Viewer Zero Footprint / ZFP-20032681, 4420-6-Centricity Universal Viewer Zero Footprint / ZFP-20032690, 2204-1-Centricity Universal Viewer Zero Footprint / ZFP-01893572, 2317-3-Centricity Universal Viewer Zero Footprint / ZFP-00165278, 2317-3-Centricity Universal Viewer Zero Footprint / ZFP-00165279, 2317-3-Centricity Universal Viewer Zero Footprint / ZFP-02337066, 2317-3-Centricity Universal Viewer Zero Footprint / ZFP-02337067, 105441-1-Centricity Universal Viewer Zero Footprint / ZFP-01987335, 105950-1-Centricity Universal Viewer Zero Footprint / ZFP-02344194, 105950-1-Centricity Universal Viewer Zero Footprint / ZFP-02369612, 103835-2-Centricity Universal Viewer Zero Footprint / ZFP-02314205, 103835-2-Centricity Universal Viewer Zero Footprint / ZFP-02336078, 103835-2-Centricity Universal Viewer Zero Footprint / ZFP-15050663, 103835-2-Centricity Universal Viewer Zero Footprint / ZFP-15050664, 103835-2-Centricity Universal Viewer Zero Footprint / ZFP-15050665, 103944-3-Centricity Universal Viewer Zero Footprint / ZFP-02286833, 103944-3-Centricity Universal Viewer Zero Footprint / ZFP-02286834, 103944-3-Centricity Universal Viewer Zero Footprint / ZFP-02286835, 103944-3-Centricity Universal Viewer Zero Footprint / ZFP-02286836, 10345-1-Centricity Universal Viewer Zero Footprint / ZFP-01419025, 10345-1-Centricity Universal Viewer Zero Footprint / ZFP-02189674, 10345-9-Centricity Universal Viewer Zero Footprint / ZFP-01425634, 10363-2-Centricity Universal Viewer Zero Footprint / ZFP-00555476, 10363-2-Centricity Universal Viewer Zero Footprint / ZFP-21992896, 104473-1-Centricity Universal Viewer Zero Footprint / ZFP-01813754, 10388-1-Centricity Universal Viewer Zero Footprint / ZFP-14762271, 10388-1-Centricity Universal Viewer Zero Footprint / ZFP-19714873, 10388-1-Centricity Universal Viewer Zero Footprint / ZFP-19714877, 10388-1-Centricity Universal Viewer Zero Footprint / ZFP-19714878, 10536-1-Centricity Universal Viewer Zero Footprint / ZFP-00161373, 10536-1-Centricity Universal Viewer Zero Footprint / ZFP-00161372, 10536-1-Centricity Universal Viewer Zero Footprint / ZFP-21325653, 1635-8-Centricity Universal Viewer Zero Footprint / ZFP-18227238, 9880-6-Centricity Universal Viewer Zero Footprint / ZFP-21775960, 102719-1-Centricity Universal Viewer Zero Footprint / ZFP-00098294, 102719-1-Centricity Universal Viewer Zero Footprint / ZFP-00098295, 102719-6-Centricity Universal Viewer Zero Footprint / ZFP-21318832, 100171-15-Centricity Universal Viewer Zero Footprint / ZFP-20869504, 100171-15-Centricity Universal Viewer Zero Footprint / ZFP-00098296, B0201-64-Centricity Universal Viewer Zero Footprint / ZFP-22255230, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 102768-1-Centricity Universal Viewer Zero Footprint / ZFP-02302156, 10283-1-Centricity Universal Viewer Zero Footprint / ZFP-00097951, 3228-10-Centricity Universal Viewer Zero Footprint / ZFP-01442252, 3228-12-Centricity Universal Viewer Zero Footprint / ZFP-01451967, 3228-12-Centricity Universal Viewer Zero Footprint / ZFP-02003801, 3228-12-Centricity Universal Viewer Zero Footprint / ZFP-02256569, 10725-1-Centricity Universal Viewer Zero Footprint / ZFP-00099453, 10725-1-Centricity Universal Viewer Zero Footprint / ZFP-00099454, 10725-1-Centricity Universal Viewer Zero Footprint / ZFP-15291356, 10725-1-Centricity Universal Viewer Zero Footprint / ZFP-19609380, 10725-7-Centricity Universal Viewer Zero Footprint / ZFP-15291372, 4362-1-Centricity Universal Viewer Zero Footprint / ZFP-14640680, 4362-1-Centricity Universal Viewer Zero Footprint / ZFP-19732060, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-14640650, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 1542-5-Centricity Universal Viewer Zero Footprint / ZFP-00116852, 1542-5-Centricity Universal Viewer Zero Footprint / ZFP-01970562, 1542-5-Centricity Universal Viewer Zero Footprint / ZFP-02345660, 105482-1-Centricity Universal Viewer Zero Footprint / ZFP-02093928, 104073-1-Centricity Universal Viewer Zero Footprint / ZFP-19660132, 104073-1-Centricity Universal Viewer Zero Footprint / ZFP-21281793, 1609-1-Centricity Universal Viewer Zero Footprint / ZFP-00104758, 1609-1-Centricity Universal Viewer Zero Footprint / ZFP-00104772, 1609-1-Centricity Universal Viewer Zero Footprint / ZFP-00104773, 1609-1-Centricity Universal Viewer Zero Footprint / ZFP-02344892, 4418-3-Centricity Universal Viewer Zero Footprint / ZFP-15291378, 4418-3-Centricity Universal Viewer Zero Footprint / ZFP-19604646, 3922-25-Centricity Universal Viewer Zero Footprint / ZFP-01290685, 100254-7-Centricity Universal Viewer Zero Footprint / ZFP-00096557, 100254-7-Centricity Universal Viewer Zero Footprint / ZFP-14776418, 100254-7-Centricity Universal Viewer Zero Footprint / ZFP-14776426, 100254-1-Centricity Universal Viewer Zero Footprint / ZFP-21752638, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, B0201-64-Centricity Universal Viewer Zero Footprint / ZFP-22255230, 11673-4-Centricity Universal Viewer Zero Footprint / ZFP-02349623, 11673-4-Centricity Universal Viewer Zero Footprint / ZFP-02349624, 9894-1-Centricity Universal Viewer Zero Footprint / ZFP-02173723, 4157-3-Centricity Universal Viewer Zero Footprint / ZFP-00100067, 4157-3-Centricity Universal Viewer Zero Footprint / ZFP-00108112, 4157-3-Centricity Universal Viewer Zero Footprint / ZFP-02363496, 4157-3-Centricity Universal Viewer Zero Footprint / ZFP-02363498, 4157-6-Centricity Universal Viewer Zero Footprint / ZFP-02290179, 4157-3-Centricity Universal Viewer Zero Footprint / ZFP-20943326, 4157-11-Centricity Universal Viewer Zero Footprint / ZFP-21282125, 4541-1-Centricity Universal Viewer Zero Footprint / ZFP-00105159, 4541-1-Centricity Universal Viewer Zero Footprint / ZFP-00105160, 102359-1-Centricity Universal Viewer Zero Footprint / ZFP-02186080, 103918-1-Centricity Universal Viewer Zero Footprint / ZFP-00100288, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-20698641, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-00100649, 103696-33-Centricity Universal Viewer Zero Footprint / ZFP-20098431, 103696-30-Centricity Universal Viewer Zero Footprint / ZFP-21237479, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-20019515, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-22284069, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-20036716, 4325-52-Centricity Universal Viewer Zero Footprint / ZFP-01367195, 4325-52-Centricity Universal Viewer Zero Footprint / ZFP-02345640, 4325-52-Centricity Universal Viewer Zero Footprint / ZFP-02345641, 4325-51-Centricity Universal Viewer Zero Footprint / ZFP-01299096, 103696-3-Centricity Universal Viewer Zero Footprint / ZFP-16868150, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-00103256, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-02351584, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20019518, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20019519, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20019521, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20019522, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036724, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036725, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036726, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036727, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036728, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036731, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036732, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-20036733, 100085-5-Centricity Universal Viewer Zero Footprint / ZFP-22284088, 105348-1-Centricity Universal Viewer Zero Footprint / ZFP-01893555, 3726-1-Centricity Universal Viewer Zero Footprint / ZFP-00599126, 2440-6-Centricity Universal Viewer Zero Footprint / ZFP-00160891, 2440-6-Centricity Universal Viewer Zero Footprint / ZFP-00160911, 2440-6-Centricity Universal Viewer Zero Footprint / ZFP-20903230, 100505-1-Centricity Universal Viewer Zero Footprint / ZFP-02375853, 102667-4-Centricity Universal Viewer Zero Footprint / ZFP-02355027, 102667-4-Centricity Universal Viewer Zero Footprint / ZFP-02355028, 102667-4-Centricity Universal Viewer Zero Footprint / ZFP-02183317, 102667-4-Centricity Universal Viewer Zero Footprint / ZFP-02348702, 102667-4-Centricity Universal Viewer Zero Footprint / ZFP-02348704, 103246-1-Centricity Universal Viewer Zero Footprint / ZFP-01973593, DZ27??-Centricity Universal Viewer Zero Footprint / ZFP-02176131, 1503586-Centricity Universal Viewer Zero Footprint / ZFP-21016603, 000009335-Centricity Universal Viewer Zero Footprint / ZFP-19631916, AH5045-Centricity Universal Viewer Zero Footprint / ZFP-17594407, 313084-Centricity Universal Viewer Zero Footprint / ZFP-14776791, 268882-Centricity Universal Viewer Zero Footprint / ZFP-19714224, 293834-Centricity Universal Viewer Zero Footprint / ZFP-21244289, 293834-Centricity Universal Viewer Zero Footprint / ZFP-21244291, 293834-Centricity Universal Viewer Zero Footprint / ZFP-21244292, 293834-Centricity Universal Viewer Zero Footprint / ZFP-02004797, 283039-Centricity Universal Viewer Zero Footprint / ZFP-01819807, 283039-Centricity Universal Viewer Zero Footprint / ZFP-02170574, 283039-Centricity Universal Viewer Zero Footprint / ZFP-02283242, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605119, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605120, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605122, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605124, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605125, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605126, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605127, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605128, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605129, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605133, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605247, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605248, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605250, 105800-Centricity Universal Viewer Zero Footprint / ZFP-00605130, 282984-Centricity Universal Viewer Zero Footprint / ZFP-01373557, 282984-Centricity Universal Viewer Zero Footprint / ZFP-01373558, 282984-Centricity Universal Viewer Zero Footprint / ZFP-01373559, 282984-Centricity Universal Viewer Zero Footprint / ZFP-01373560, 282984-Centricity Universal Viewer Zero Footprint / ZFP-00802574, 589019-Centricity Universal Viewer Zero Footprint / ZFP-21738608, 589019-Centricity Universal Viewer Zero Footprint / ZFP-21738610, 98654-Centricity Universal Viewer Zero Footprint / ZFP-19719416, 533917-Centricity Universal Viewer Zero Footprint / ZFP-21740697, 533917-Centricity Universal Viewer Zero Footprint / ZFP-21740698, 533917-Centricity Universal Viewer Zero Footprint / ZFP-21740701, 533917-Centricity Universal Viewer Zero Footprint / ZFP-21740702, 104367-Centricity Universal Viewer Zero Footprint / ZFP-20832389, 75518-Centricity Universal Viewer Zero Footprint / ZFP-21258025, 75518-Centricity Universal Viewer Zero Footprint / ZFP-21738619, 75518-Centricity Universal Viewer Zero Footprint / ZFP-21738633, 75518-Centricity Universal Viewer Zero Footprint / ZFP-21738634, 241164-Centricity Universal Viewer Zero Footprint / ZFP-21740708, 2227709-Centricity Universal Viewer Zero Footprint / ZFP-22175336, 3627571-Centricity Universal Viewer Zero Footprint / ZFP-02171345, 1339046-Centricity Universal Viewer Zero Footprint / ZFP-19719421, 240578-Centricity Universal Viewer Zero Footprint / ZFP-22173127, 12743610-Centricity Universal Viewer Zero Footprint / ZFP-19600555, 7293623-Centricity Universal Viewer Zero Footprint / ZFP-20632623, 576427-Centricity Universal Viewer Zero Footprint / ZFP-19719437, 342130-Centricity Universal Viewer Zero Footprint / ZFP-19717012, WJPEOPLE-Centricity Universal Viewer Zero Footprint / ZFP-02171481, 333433-Centricity Universal Viewer Zero Footprint / ZFP-20080648, 483049-Centricity Universal Viewer Zero Footprint / ZFP-20042108, 445933-Centricity Universal Viewer Zero Footprint / ZFP-20039061, BEIJING302-Centricity Universal Viewer Zero Footprint / ZFP-02170935, 241220-Centricity Universal Viewer Zero Footprint / ZFP-20042482, 4477574-Centricity Universal Viewer Zero Footprint / ZFP-19716408, 4477574-Centricity Universal Viewer Zero Footprint / ZFP-20039208, 589019-Centricity Universal Viewer Zero Footprint / ZFP-19722710, FUDAN-TUMOUR-Centricity Universal Viewer Zero Footprint / ZFP-02171473, 331610-Centricity Universal Viewer Zero Footprint / ZFP-20632619, 331610-Centricity Universal Viewer Zero Footprint / ZFP-21740704, 331610-Centricity Universal Viewer Zero Footprint / ZFP-21740705, 331610-Centricity Universal Viewer Zero Footprint / ZFP-21740706, 104367-Centricity Universal Viewer Zero Footprint / ZFP-19719110, 240578-Centricity Universal Viewer Zero Footprint / ZFP-20080645, 75518-Centricity Universal Viewer Zero Footprint / ZFP-20611133, 241116-Centricity Universal Viewer Zero Footprint / ZFP-20915523, SH-CH-Centricity Universal Viewer Zero Footprint / ZFP-02171452, 2272753-Centricity Universal Viewer Zero Footprint / ZFP-20085689, 533917-Centricity Universal Viewer Zero Footprint / ZFP-20915520, 241164-Centricity Universal Viewer Zero Footprint / ZFP-20039201, 98654-Centricity Universal Viewer Zero Footprint / ZFP-19719417, 1377666-Centricity Universal Viewer Zero Footprint / ZFP-20038841, 507105-Centricity Universal Viewer Zero Footprint / ZFP-19719124, 2420831-Centricity Universal Viewer Zero Footprint / ZFP-20080656, YN-HAOBANG-Centricity Universal Viewer Zero Footprint / ZFP-02171461, M44770-Centricity Universal Viewer Zero Footprint / ZFP-00117027, A44986-Centricity Universal Viewer Zero Footprint / ZFP-22172875, 16371623-Centricity Universal Viewer Zero Footprint / ZFP-20081079, M3865783-Centricity Universal Viewer Zero Footprint / ZFP-00071606, 13984613-Centricity Universal Viewer Zero Footprint / ZFP-20031336, 61760-Centricity Universal Viewer Zero Footprint / ZFP-23290069, M2209952-Centricity Universal Viewer Zero Footprint / ZFP-00071585, A44957-Centricity Universal Viewer Zero Footprint / ZFP-01270574, M8388356-Centricity Universal Viewer Zero Footprint / ZFP-02394952, M614219-Centricity Universal Viewer Zero Footprint / ZFP-00116630, M3488569-Centricity Universal Viewer Zero Footprint / ZFP-00071697, M3488569-Centricity Universal Viewer Zero Footprint / ZFP-00071708, M3488569-Centricity Universal Viewer Zero Footprint / ZFP-23227681, MQ12345 / MQ1090-Centricity Universal Viewer Zero Footprint / ZFP-00162347, 13985614-Centricity Universal Viewer Zero Footprint / ZFP-20031340, M1171652-Centricity Universal Viewer Zero Footprint / ZFP-18347633, M1175651-Centricity Universal Viewer Zero Footprint / ZFP-18347636, A44986-Centricity Universal Viewer Zero Footprint / ZFP-01797479, M1003239-Centricity Universal Viewer Zero Footprint / ZFP-02360168, M1003239-Centricity Universal Viewer Zero Footprint / ZFP-14776398, L16502-Centricity Universal Viewer Zero Footprint / ZFP-13964016, C41917-Centricity Universal Viewer Zero Footprint / ZFP-00117236, D41917-Centricity Universal Viewer Zero Footprint / ZFP-00117239, M54807-Centricity Universal Viewer Zero Footprint / ZFP-19141534, M42176-Centricity Universal Viewer Zero Footprint / ZFP-00043764, M45052-Centricity Universal Viewer Zero Footprint / ZFP-01270575, 744459-Centricity Universal Viewer Zero Footprint / ZFP-02398817, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, 145196 - HC2028-Centricity Universal Viewer Zero Footprint / ZFP-02351755, X9002184 - PC2184-Centricity Universal Viewer Zero Footprint / ZFP-02362325, 530019-Centricity Universal Viewer Zero Footprint / ZFP-01282264, 1604200X - 509940-Centricity Universal Viewer Zero Footprint / ZFP-14194840, HC4017-Centricity Universal Viewer Zero Footprint / ZFP-02400997, 2530876X - PC0114-Centricity Universal Viewer Zero Footprint / ZFP-00261054, HC1627-Centricity Universal Viewer Zero Footprint / ZFP-00260658, 120111 - UESS01-Centricity Universal Viewer Zero Footprint / ZFP-02041830, UC1220-Centricity Universal Viewer Zero Footprint / ZFP-02299598, 600791-Centricity Universal Viewer Zero Footprint / ZFP-02099050, X9000727 - HC0727-Centricity Universal Viewer Zero Footprint / ZFP-00111408, 1887824X - 560828-Centricity Universal Viewer Zero Footprint / ZFP-14769682, 362520-Centricity Universal Viewer Zero Footprint / ZFP-00393290, 410014-Centricity Universal Viewer Zero Footprint / ZFP-16295260, HC1642-Centricity Universal Viewer Zero Footprint / ZFP-02395840, X9000094 - HC0094-Centricity Universal Viewer Zero Footprint / ZFP-00111203, X9007551 --Centricity Universal Viewer Zero Footprint / ZFP-14369166, HC0881-Centricity Universal Viewer Zero Footprint / ZFP-01423579, 142510 - HC2641-Centricity Universal Viewer Zero Footprint / ZFP-14430198, 5328294X - 424881-Centricity Universal Viewer Zero Footprint / ZFP-02395836, 278146-Centricity Universal Viewer Zero Footprint / ZFP-20631150, 263394-Centricity Universal Viewer Zero Footprint / ZFP-22285625, 614388-Centricity Universal Viewer Zero Footprint / ZFP-20995981, VEDANTAMRF-Centricity Universal Viewer Zero Footprint / ZFP-01973242, AIIMS-CN-Centricity Universal Viewer Zero Footprint / ZFP-01892392, NIMHANS-Centricity Universal Viewer Zero Footprint / ZFP-01890910, SCTIMST-Centricity Universal Viewer Zero Footprint / ZFP-02299430, JIPMER-Centricity Universal Viewer Zero Footprint / ZFP-01820556, CONTINENTAL-Centricity Universal Viewer Zero Footprint / ZFP-20944996, 540880-Centricity Universal Viewer Zero Footprint / ZFP-14767101, TMC-IN-Centricity Universal Viewer Zero Footprint / ZFP-02205262, 327750-Centricity Universal Viewer Zero Footprint / ZFP-22177091, 767262-Centricity Universal Viewer Zero Footprint / ZFP-21005868, 570450-Centricity Universal Viewer Zero Footprint / ZFP-21317343, 767262-Centricity Universal Viewer Zero Footprint / ZFP-21272858, 767262-Centricity Universal Viewer Zero Footprint / ZFP-22176296, 327750-Centricity Universal Viewer Zero Footprint / ZFP-22176480, 327750-Centricity Universal Viewer Zero Footprint / ZFP-22225271, 327750-Centricity Universal Viewer Zero Footprint / ZFP-24333848, 327750-Centricity Universal Viewer Zero Footprint / ZFP-24334469, 2347557-Centricity Universal Viewer Zero Footprint / ZFP-21314383, 327750-Centricity Universal Viewer Zero Footprint / ZFP-22137249, 428320-Centricity Universal Viewer Zero Footprint / ZFP-21005854, 428320-Centricity Universal Viewer Zero Footprint / ZFP-21005861, 327750-Centricity Universal Viewer Zero Footprint / ZFP-22226547, 570450-Centricity Universal Viewer Zero Footprint / ZFP-21272330.662878-Centricity Universal Viewer Zero Footprint / ZFP-20922238, 5379583-Centricity Universal Viewer Zero Footprint / ZFP-20984219, 4823580-Centricity Universal Viewer Zero Footprint / ZFP-22146167, 16745082-Centricity Universal Viewer Zero Footprint / ZFP-22223500, 5204593-Centricity Universal Viewer Zero Footprint / ZFP-20535380, 5383585-Centricity Universal Viewer Zero Footprint / ZFP-16174592, 5382589-Centricity Universal Viewer Zero Footprint / ZFP-02398975, 327810-Centricity Universal Viewer Zero Footprint / ZFP-20095699, 662878-Centricity Universal Viewer Zero Footprint / ZFP-20095687, 662878-Centricity Universal Viewer Zero Footprint / ZFP-20095693, 570450-Centricity Universal Viewer Zero Footprint / ZFP-02039952, 327750-Centricity Universal Viewer Zero Footprint / ZFP-21763022, 327750-Centricity Universal Viewer Zero Footprint / ZFP-22273495, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20984225, 16737663-Centricity Universal Viewer Zero Footprint / ZFP-24335015, 3076565-Centricity Universal Viewer Zero Footprint / ZFP-14224595, 2347557-Centricity Universal Viewer Zero Footprint / ZFP-20679129, 2347557-Centricity Universal Viewer Zero Footprint / ZFP-20679123, 767262-Centricity Universal Viewer Zero Footprint / ZFP-02402591, 767262-Centricity Universal Viewer Zero Footprint / ZFP-19417272, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20661270, 327750-Centricity Universal Viewer Zero Footprint / ZFP-01445613, 327750-Centricity Universal Viewer Zero Footprint / ZFP-01977821, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20034458, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20661156, 327750-Centricity Universal Viewer Zero Footprint / ZFP-19615100, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20021119, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20661247, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20661278, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20661288, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20661295, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20679117, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20585460, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20656005, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20656011, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20656018, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904858, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904852, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904845, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904838, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904832, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904585, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904579, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904572, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904565, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904559, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904550, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904543, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904536, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904307, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904291, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904283, 327750-Centricity Universal Viewer Zero Footprint / ZFP-20904273, 603806-Centricity Universal Viewer Zero Footprint / ZFP-14767092, 428320-Centricity Universal Viewer Zero Footprint / ZFP-19735893, 1416035-Centricity Universal Viewer Zero Footprint / ZFP-14195747, 1416035-Centricity Universal Viewer Zero Footprint / ZFP-20034225, 651342-Centricity Universal Viewer Zero Footprint / ZFP-02400072, 428320-Centricity Universal Viewer Zero Footprint / ZFP-21005846, 767259-Centricity Universal Viewer Zero Footprint / ZFP-20655989, 7821598-Centricity Universal Viewer Zero Footprint / ZFP-21272316, 16697676_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-21228217, 81100-Centricity Universal Viewer Zero Footprint / ZFP-02339355, 81077-Centricity Universal Viewer Zero Footprint / ZFP-21770142, 81077-Centricity Universal Viewer Zero Footprint / ZFP-20680055, IL1098-Centricity Universal Viewer Zero Footprint / ZFP-01451983, C53548-Centricity Universal Viewer Zero Footprint / ZFP-21224902, 575008-Centricity Universal Viewer Zero Footprint / ZFP-16672644, B53686-Centricity Universal Viewer Zero Footprint / ZFP-01271448, A51560-Centricity Universal Viewer Zero Footprint / ZFP-02336360, A3431341-Centricity Universal Viewer Zero Footprint / ZFP-01980190, A57198-Centricity Universal Viewer Zero Footprint / ZFP-14767566, A51043-Centricity Universal Viewer Zero Footprint / ZFP-15956283, A3784447-Centricity Universal Viewer Zero Footprint / ZFP-01270583, JP-070240-Centricity Universal Viewer Zero Footprint / ZFP-22298203, JP-283699-Centricity Universal Viewer Zero Footprint / ZFP-20978978, JP-340111-Centricity Universal Viewer Zero Footprint / ZFP-01275379, JP-122363-Centricity Universal Viewer Zero Footprint / ZFP-14374736, JP-120427-Centricity Universal Viewer Zero Footprint / ZFP-19984123, JP-180086-Centricity Universal Viewer Zero Footprint / ZFP-02300794, JP-070095-Centricity Universal Viewer Zero Footprint / ZFP-17594505, JP-070167-Centricity Universal Viewer Zero Footprint / ZFP-01275409, JP-280330-Centricity Universal Viewer Zero Footprint / ZFP-19630031, JP-280073-Centricity Universal Viewer Zero Footprint / ZFP-02394736, JP-280078-Centricity Universal Viewer Zero Footprint / ZFP-14194565, JP-280121-Centricity Universal Viewer Zero Footprint / ZFP-20608365, JP-283568-Centricity Universal Viewer Zero Footprint / ZFP-16450970, JP-280868-Centricity Universal Viewer Zero Footprint / ZFP-20922627, JP-080215-Centricity Universal Viewer Zero Footprint / ZFP-01825286, JP-080218-Centricity Universal Viewer Zero Footprint / ZFP-01367626, JP-080218-Centricity Universal Viewer Zero Footprint / ZFP-01367627, JP-040538-Centricity Universal Viewer Zero Footprint / ZFP-20022063, JP-040022-Centricity Universal Viewer Zero Footprint / ZFP-02157466, JP-040022-Centricity Universal Viewer Zero Footprint / ZFP-02289421, JP-040022-Centricity Universal Viewer Zero Footprint / ZFP-02289422, JP-040164-Centricity Universal Viewer Zero Footprint / ZFP-02099044, JP-150017-Centricity Universal Viewer Zero Footprint / ZFP-02088882, JP-273074-Centricity Universal Viewer Zero Footprint / ZFP-14761365, JP-110043-Centricity Universal Viewer Zero Footprint / ZFP-19611403, JP-110043-Centricity Universal Viewer Zero Footprint / ZFP-19715420, JP-110043-Centricity Universal Viewer Zero Footprint / ZFP-19715639, JP-250031-Centricity Universal Viewer Zero Footprint / ZFP-02399308, JP-220238-Centricity Universal Viewer Zero Footprint / ZFP-14764727, JP-130644-Centricity Universal Viewer Zero Footprint / ZFP-20084064, JP-130314-Centricity Universal Viewer Zero Footprint / ZFP-02274906, JP-130314-Centricity Universal Viewer Zero Footprint / ZFP-19616515, JP-130440-Centricity Universal Viewer Zero Footprint / ZFP-01979557, JP-130440-Centricity Universal Viewer Zero Footprint / ZFP-01979560, JP-130440-Centricity Universal Viewer Zero Footprint / ZFP-02040264, JP-130539-Centricity Universal Viewer Zero Footprint / ZFP-02304194, JP-130539-Centricity Universal Viewer Zero Footprint / ZFP-02332756, JP-130539-Centricity Universal Viewer Zero Footprint / ZFP-02332757, JP-130067-Centricity Universal Viewer Zero Footprint / ZFP-20483773, JP-131312-Centricity Universal Viewer Zero Footprint / ZFP-20617592, JP-300074-Centricity Universal Viewer Zero Footprint / ZFP-20509115, JP-300048-Centricity Universal Viewer Zero Footprint / ZFP-01275433, JP-300048-Centricity Universal Viewer Zero Footprint / ZFP-21210229, 851746-Centricity Universal Viewer Zero Footprint / ZFP-20923876, 851894-Centricity Universal Viewer Zero Footprint / ZFP-01973237, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-19638454, KW1007-Centricity Universal Viewer Zero Footprint / ZFP-14549479, KW1096-Centricity Universal Viewer Zero Footprint / ZFP-20541640, LB1160-Centricity Universal Viewer Zero Footprint / ZFP-01591506, SGB-MY-Centricity Universal Viewer Zero Footprint / ZFP-19630188, 2348649-Centricity Universal Viewer Zero Footprint / ZFP-14769747, HUSM-Centricity Universal Viewer Zero Footprint / ZFP-19614903, IPPT-Centricity Universal Viewer Zero Footprint / ZFP-19614914, 1322427-Centricity Universal Viewer Zero Footprint / ZFP-19660792, MSU-MY-Centricity Universal Viewer Zero Footprint / ZFP-01893497, 5499-Centricity Universal Viewer Zero Footprint / ZFP-14868737, 332068-Centricity Universal Viewer Zero Footprint / ZFP-20870045, PER503-Centricity Universal Viewer Zero Footprint / ZFP-21037599, SA2128-Centricity Universal Viewer Zero Footprint / ZFP-21330000, 2978539-Centricity Universal Viewer Zero Footprint / ZFP-22213020, SA1060-Centricity Universal Viewer Zero Footprint / ZFP-00066963, SA1060-Centricity Universal Viewer Zero Footprint / ZFP-20025907, SA1005-Centricity Universal Viewer Zero Footprint / ZFP-00065849, SA2329-Centricity Universal Viewer Zero Footprint / ZFP-19678190, SA 1175-Centricity Universal Viewer Zero Footprint / ZFP-19719174, 2978539-Centricity Universal Viewer Zero Footprint / ZFP-19600952, SA2410-Centricity Universal Viewer Zero Footprint / ZFP-20025015, SA1304-Centricity Universal Viewer Zero Footprint / ZFP-02396452, SA 2385-Centricity Universal Viewer Zero Footprint / ZFP-20501423, SA2315-Centricity Universal Viewer Zero Footprint / ZFP-19669732, SA1006-Centricity Universal Viewer Zero Footprint / ZFP-02328482, SA 1176-Centricity Universal Viewer Zero Footprint / ZFP-19609918, SA1209-Centricity Universal Viewer Zero Footprint / ZFP-00066545, SA2235-Centricity Universal Viewer Zero Footprint / ZFP-19660083, SA2191-Centricity Universal Viewer Zero Footprint / ZFP-20617518, SA1054-Centricity Universal Viewer Zero Footprint / ZFP-19722888, SA 2281-Centricity Universal Viewer Zero Footprint / ZFP-18895618, SA2110-Centricity Universal Viewer Zero Footprint / ZFP-01805103, SA1008-Centricity Universal Viewer Zero Footprint / ZFP-00602579, SA1008-Centricity Universal Viewer Zero Footprint / ZFP-20620862, SA2645-Centricity Universal Viewer Zero Footprint / ZFP-21174755, SA2110-Centricity Universal Viewer Zero Footprint / ZFP-20999430, SA1008-Centricity Universal Viewer Zero Footprint / ZFP-21201005, SA 2393-Centricity Universal Viewer Zero Footprint / ZFP-19620146, SA2006-Centricity Universal Viewer Zero Footprint / ZFP-20019291, KTPH-Centricity Universal Viewer Zero Footprint / ZFP-01452994, 243134-Centricity Universal Viewer Zero Footprint / ZFP-19629976, 1363272-Centricity Universal Viewer Zero Footprint / ZFP-01286173, 9287622 - 2415034-Centricity Universal Viewer Zero Footprint / ZFP-02041820, NUH-SG-Centricity Universal Viewer Zero Footprint / ZFP-01476439, TTSH-RAD-Centricity Universal Viewer Zero Footprint / ZFP-01977001, 16723955-Centricity Universal Viewer Zero Footprint / ZFP-20685541, 16723955-Centricity Universal Viewer Zero Footprint / ZFP-20685543, 17006190-Centricity Universal Viewer Zero Footprint / ZFP-22258838, MPX431-Centricity Universal Viewer Zero Footprint / ZFP-01813942, FPD108-Centricity Universal Viewer Zero Footprint / ZFP-20084307, FPD108-Centricity Universal Viewer Zero Footprint / ZFP-20084312, GED649-Centricity Universal Viewer Zero Footprint / ZFP-19666476, GPX021-Centricity Universal Viewer Zero Footprint / ZFP-02041108, DEN272-Centricity Universal Viewer Zero Footprint / ZFP-01370312, DEN272-Centricity Universal Viewer Zero Footprint / ZFP-21287696, DE492A-Centricity Universal Viewer Zero Footprint / ZFP-19633836, NPV793-Centricity Universal Viewer Zero Footprint / ZFP-19672387, 150080-Centricity Universal Viewer Zero Footprint / ZFP-01299636, 492861-Centricity Universal Viewer Zero Footprint / ZFP-02248288, 437918-Centricity Universal Viewer Zero Footprint / ZFP-00163452, 722050-Centricity Universal Viewer Zero Footprint / ZFP-00555881, 0853550429-Centricity Universal Viewer Zero Footprint / ZFP-02041927, 0853550209-Centricity Universal Viewer Zero Footprint / ZFP-02093686, 0853550675-Centricity Universal Viewer Zero Footprint / ZFP-01287925, 0853550199-Centricity Universal Viewer Zero Footprint / ZFP-02093677, 0853550804-Centricity Universal Viewer Zero Footprint / ZFP-01286189, 0853550198-Centricity Universal Viewer Zero Footprint / ZFP-02093675, 0853550198-Centricity Universal Viewer Zero Footprint / ZFP-20541728, 0853551073-Centricity Universal Viewer Zero Footprint / ZFP-19622953, 0853550185-Centricity Universal Viewer Zero Footprint / ZFP-02093535, 0853550993-Centricity Universal Viewer Zero Footprint / ZFP-02093449, 0853551064-Centricity Universal Viewer Zero Footprint / ZFP-17533740, 0853550200-Centricity Universal Viewer Zero Footprint / ZFP-02093882, 0853550200-Centricity Universal Viewer Zero Footprint / ZFP-20541715, 0853550312-Centricity Universal Viewer Zero Footprint / ZFP-01286591, 0853550201-Centricity Universal Viewer Zero Footprint / ZFP-02093879, 0853550201-Centricity Universal Viewer Zero Footprint / ZFP-20541705, 0853550616-Centricity Universal Viewer Zero Footprint / ZFP-20912105, 0853550218-Centricity Universal Viewer Zero Footprint / ZFP-02093540, 0853550218-Centricity Universal Viewer Zero Footprint / ZFP-17533749, 81188-Centricity Universal Viewer Zero Footprint / ZFP-18227246, 6002-Centricity Universal Viewer Zero Footprint / ZFP-00158571, 34365-Centricity Universal Viewer Zero Footprint / ZFP-14374646, 34225-Centricity Universal Viewer Zero Footprint / ZFP-00157020, 486961-Centricity Universal Viewer Zero Footprint / ZFP-20611232, RXQ50-Centricity Universal Viewer Zero Footprint / ZFP-00505094, RXQ50-Centricity Universal Viewer Zero Footprint / ZFP-01701456, 50158-Centricity Universal Viewer Zero Footprint / ZFP-01981985, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 81564-Centricity Universal Viewer Zero Footprint / ZFP-00436804, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231, 446860-Centricity Universal Viewer Zero Footprint / ZFP-22255231
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page