Skip to content

Ivenix, Inc. recallZ-3153-2024 · Fresenius Kabi USA, LLC · initiated Aug 30, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.

FDA root cause category: Software design

What the company is doing

An URGENT Medical Device Field Correction Important Safety Update notification letter dated 8/30/24 was sent to customers. Required Actions for Users: 1. Please fill out the Customer Reply Form attached to this letter to acknowledge receipt. 2. If your facility further distributes or transfers products to satellite sites or other locations, please ask these locations to contact the email address on the Customer Reply Form so we can send notice to those satellites directly. 3. Install new software version (5.9.2) for the Ivenix Large Volume Pump (LVP). See details below. How To Obtain and Install Software Version 5.9.2 To request installation, please reach out promptly to your Fresenius Kabi representative (1-855- 354-6387 or [email protected]). They will push the new software update to your pumps. When the LVP version 5.9.2 software update has been sent to the LVP, the following prompt appears before the LVP is shut down. Customer support - email Ivenix_su [email protected] Customer support - phone 1-855-354-6387 Normal business Hours are Monday through Friday 8:30 AM to 5 PM Eastern Time, After Hours Supported with Voicemail Messaging

Recall details

Recall number
Z-3153-2024
Status
Open, Classified
Initiated by the company
Aug 30, 2024
Posted by the FDA
Sep 26, 2024
Recalling firm
Fresenius Kabi USA, LLC
Quantity
17 units
Product code
FRN
FDA submissions
K183311
Distribution
US Nationwide.

Product description

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.

Code information (models, lots, serial numbers)

UDI-DI: 00811505030122; Software Version 5.9.1 and prior

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page