Ivenix, Inc. recallZ-3153-2024 · Fresenius Kabi USA, LLC · initiated Aug 30, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.
FDA root cause category: Software design
What the company is doing
An URGENT Medical Device Field Correction Important Safety Update notification letter dated 8/30/24 was sent to customers. Required Actions for Users: 1. Please fill out the Customer Reply Form attached to this letter to acknowledge receipt. 2. If your facility further distributes or transfers products to satellite sites or other locations, please ask these locations to contact the email address on the Customer Reply Form so we can send notice to those satellites directly. 3. Install new software version (5.9.2) for the Ivenix Large Volume Pump (LVP). See details below. How To Obtain and Install Software Version 5.9.2 To request installation, please reach out promptly to your Fresenius Kabi representative (1-855- 354-6387 or [email protected]). They will push the new software update to your pumps. When the LVP version 5.9.2 software update has been sent to the LVP, the following prompt appears before the LVP is shut down. Customer support - email Ivenix_su [email protected] Customer support - phone 1-855-354-6387 Normal business Hours are Monday through Friday 8:30 AM to 5 PM Eastern Time, After Hours Supported with Voicemail Messaging
Recall details
- Recall number
- Z-3153-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 30, 2024
- Posted by the FDA
- Sep 26, 2024
- Recalling firm
- Fresenius Kabi USA, LLC
- Quantity
- 17 units
- Product code
- FRN
- FDA submissions
- K183311
- Distribution
- US Nationwide.
Product description
Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.
Code information (models, lots, serial numbers)
UDI-DI: 00811505030122; Software Version 5.9.1 and prior
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ivenix, Inc. 2.0Fresenius Kabi AG · Pump, Infusion10 recalls0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Ivenix Infusion System LVP-0004Fresenius Kabi AG · Pump, Infusion7 recalls2 deaths4,328 reports
- Ivenix Infusion System 030-0028-08Fresenius Kabi AG · Pump, Infusion7 reports
- Ivenix Infusion System 030-0027-06Fresenius Kabi AG · Infusion Safety Management Software3 reports
- Ivenix Infusion System 030-0028-07Fresenius Kabi AG · Pump, Infusion2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.