Skip to content

TORNIER PERFORM HUMERAL SYSTEM recallZ-3309-2024 · Tornier, Inc · initiated Aug 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tornier, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.

FDA root cause category: Under Investigation by firm

What the company is doing

Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via email. The notice explained the issue, potential risk, and requested the following actions be taken: Identify and isolate the affected units, and respond to the notice via email. Upon receipt of the response, Stryker will arrange for the return of the device. Those consignees who further distributed the product (to other sites or customers) were directed to notify those parties. For questions or concerns, contact: [email protected].

Recall details

Recall number
Z-3309-2024
Status
Open, Classified
Initiated by the company
Aug 29, 2024
Posted by the FDA
Sep 27, 2024
Recalling firm
Tornier, Inc
Quantity
38 units
Product code
PAO
FDA submissions
K201315
Distribution
US Nationwide distribution.

Product description

Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty

Code information (models, lots, serial numbers)

GTIN 00846832084494, Batch Lot # 1495AY

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page