ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound… recallZ-3215-2024 · Siemens Medical Solutions USA, Inc. · initiated Aug 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
FDA root cause category: Software design
What the company is doing
On 8/15/2024, correction notices were distributed to customers who were asked to do the following: To avoid potential misdiagnosis by an under-represented volume flow measurement value, use the factory default display configuration of l/min (liters per minute). Review of previous examination results ONLY if a clinical assessment reviewed a volume flow measurement(s) with the unit of measurement configured to display a value in milliliters. Firm's Customer Service Engineer will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available in the summer of 2024. Ensure that all users of ACUSON Juniper and ACUSON Maple systems within your organization, and others who may need to be informed, receive the safety relevant information provided with this notice and take the actions specified herein. If you have further questions, please contact the firm's Ultrasound Service Customer Care Center at 1-800-888-7436.
Recall details
- Recall number
- Z-3215-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 15, 2024
- Posted by the FDA
- Sep 24, 2024
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Quantity
- 7022
- Product code
- IYN
- FDA submissions
- K201130
- Distribution
- US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe
Product description
ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.
Code information (models, lots, serial numbers)
11335791, UDI-DI: 04056869152059: 1.0/VA10G and prior, 1.5/ VB10H and prior , 2.0/ VB11C and prior, 2.5/ VB30D and prior. 11653093, UDI-DI: 04056869957357: 2.0/ VB11C and prior, 2.5/ VB30D and prior.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACUSON Juniper Diagnostic Ultrasound System 11335791Siemens Medical Solutions USA, Inc. · Transducer, ultrasonic, diagnostic3 recalls2 reports
- ACUSON Juniper Select Diagnostic Ultrasound System 11653093Siemens Medical Solutions USA, Inc. · Transducer, ultrasonic, diagnostic2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95254).
- Z-3216-2024 · Open, ClassifiedACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software
- Z-3217-2024 · Open, ClassifiedACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page