1.2mm Cross Cut Fissure - Intended for use in the cutting… recallZ-0063-2025 · Stryker Corporation · initiated Sep 5, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
FDA root cause category: Process control
What the company is doing
Stryker issued Urgent Medical Device Recall Letter on 9/5/24 via FedEx 2-day to Recall Department, Materials Manager, OR Director, Risk Manager. Letter states reason for recall, health risk and action to take: Actions needed 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments recall team directly by calling (269) 568-1481 or email at [email protected].
Recall details
- Recall number
- Z-0063-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 5, 2024
- Posted by the FDA
- Oct 11, 2024
- Recalling firm
- Stryker Corporation
- Quantity
- 307 units
- Product code
- HBE
- FDA submissions
- K143320
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Columbia.
Product description
1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012
Code information (models, lots, serial numbers)
GTIN: 04546540362346 Lot number: 22329017
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA 5820066012Stryker Corporation · Bur, Surgical, General & Plastic Surgery1 recall4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page