Monarch recallZ-0070-2025 · Merit Medical Systems, Inc. · initiated Aug 19, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.
FDA root cause category: Packaging process control
What the company is doing
On 8/19/24, recall notices were emailed to customers asking them to do the following: 1) If affected devices are discovered, quarantine them, discontinue use and distribution, and return to the recalling firm. 2) Ensure that applicable personnel within your organization are made aware of this field action. 3) If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4) Complete and return the customer response form via email to [email protected] If you have any questions concerning this communication, please contact the firm's Sales Representatives or Customer Service via email at [email protected] or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri.
Recall details
- Recall number
- Z-0070-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 19, 2024
- Posted by the FDA
- Oct 11, 2024
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 7,099
- Product code
- OEQ
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of France, Italy, Jamaica, Ireland, Paraguay, Australia, India, Nepal, Japan.
Product description
Custom Inflation Kit, REF: K05-00052M, K05-00053H, K05-00109G, K05-00137K, K05-00296L, K05-00399D, K05-00404C, K05-00410G, K05-00438G, K05-00502F, K05-00564L, K05-00587D, K05-00612Q, K05-00704F, K05-00724F, K05-00793B, K05-00798, K05-00890D, K05-00890D, K05-00944D, K05-00956D, K05-01018C, K05-01051, K05-01104G, K05-01140D, K05-01194G, K05-01274D, K05-01304A, K05-01756B, K05-01820B, K05-01853A, K05-01855L, K05-01895A, K05-01933A, K05-02161, K05-02431A, K05-02487A, K05-02489C, K05-02495B, K05-02554, K05-02573, K05-02575, K05-02595, K05-02595, K05-02657A, K05-02810, K05-02887, K05-03069A, K05-03095B, K05-03112, K05-03124, K05-03150, K05-50037G
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
REF/UDI-DI/Lot(Expiration): K05-00052M/00884450138120/H2934735(5/5/2027), H2943603(5/19/2027). K05-00053H/00884450052921/H2918538(4/22/2027), H2934736(5/5/2027). K05-00109G/00884450464533/H2918443(4/15/2027), H2925236(4/22/2027), H2943604(5/19/2027). K05-00137K/00884450290316/H2925237(12/31/2026), H2934737(5/12/2027). K05-00296L/00884450467824/H2918499(4/29/2027), H2925271(4/29/2027). K05-00399D/00884450138298/H2929510(4/29/2027), H2934739(5/12/2027). K05-00404C/00884450138304/H2918500(4/29/2027). K05-00410G/00884450142295/H2918501(4/15/2027), H2934833(5/12/2027). K05-00438G/00884450142301/H2918502(4/15/2027), H2934834(5/5/2027), H2978145(6/3/2027). K05-00502F/00884450138427/H2918505(4/15/2027), H2934837(5/5/2027). K05-00564L/00884450257166/H2934839(5/5/2027). K05-00587D/00884450052952/H2918446(4/22/2027), H2934744(5/5/2027), H2938620(5/12/2027). K05-00612Q/00884450138519/H2925240(4/22/2027), H2934746(5/5/2027), H2943608(5/26/2027). K05-00704F/00884450138588/H2918508(4/15/2027), H2938622(5/12/2027). K05-00724F/00884450146538/H2918509(4/15/2027). K05-00793B/00884450236987/H2918448(4/15/2027), H2929514(4/29/2027), H2938623(5/12/2027). K05-00798/00884450046951/H2925242(4/22/2027), H2943609(5/19/2027). K05-00890D/00884450453629/H2918449(4/15/2027), H2929515(4/29/2027). K05-00890D/00884450453629/H2934749(5/5/2027), H2943610(5/19/2027). K05-00944D/00884450151631/H2918450(4/15/2027), H2929516(4/29/2027). K05-00956D/00884450142448/H2918511(4/15/2027), H2925278(4/22/2027), H2938625(5/12/2027). K05-01018C/00884450152324/H2925244(4/22/2027), H2938626(5/12/2027). K05-01051/00884450047057/H2925280(4/22/2027). K05-01104G/00884450145401/H2925285(4/22/2027), H2929540(4/29/2027), H2934846(5/5/2027). K05-01140D/00884450147931/H2925286(4/22/2027), H2934847(5/5/2027), H2972616(5/26/2027). K05-01194G/00884450409091/H2925288(4/22/2027). K05-01274D/00884450160596/H2918453(4/22/2027), H2934750(5/12/2027), H2938628(5/12/2027). K05-01304A/00884450279014/H2972617(5/26/2027). K05-01756B/00884450145449/H2929520(4/29/2027). K05-01820B/00884450639115/H2929521(4/29/2027), H2943614(5/19/2027). K05-01853A/00884450153864/H2918460(4/15/2027), H2934761(5/12/2027), H2943615(5/19/2027). K05-01855L/00884450453667/H2918518(4/15/2027), H2978189(6/3/2027). K05-01895A/00884450141458/H2929522(5/26/2027). K05-01933A/00884450462768/H2918519(4/15/2027), H2934855(5/5/2027). K05-02161/00884450139738/H2918522(4/15/2027), H2925298(4/29/2027), H2943617(5/19/2027), H2978192(6/3/2027). K05-02431A/00884450150191/H2929525(4/29/2027). K05-02487A/00884450140468/H2918472(4/15/2027). K05-02489C/00884450639177/H2938638(5/12/2027). K05-02495B/00884450257326/H2925259(4/22/2027), H2934775(5/5/2027). K05-02554/00884450148372/H2934780(5/5/2027). K05-02573/00884450148884/H2918476(4/15/2027). K05-02575/00884450158500/H2938639(5/12/2027). K05-02595/00884450165881/H2918525(4/15/2027), H2929547(4/29/2027), H2938640(5/12/2027). K05-02595/00884450165881/H2972630(5/26/2027), H2982566(6/10/2027). K05-02657A/00884450639214/H2918480(4/15/2027), H2925265(4/29/2027), H2934784(5/5/2027), H2938643(5/12/2027), H2982568(6/10/2027). K05-02810/00884450253083/H2929528(4/29/2027), H2938646(5/12/2027). K05-02887/00884450285923/H2918527(4/29/2027), H2925313(4/29/2027), H2934862(5/12/2027). K05-03069A/00884450461440/H2938301(3/31/2027). K05-03095B/00884450639429/H2925319(4/22/2027), H2938649(5/12/2027). K05-03112/00884450404492/H2938650(5/12/2027). K05-03124/00884450432631/H2934820(5/5/2027), H2938651(5/12/2027), H2978212(6/3/2027). K05-03150/00884450480045/H2925320(4/22/2027). K05-50037G/0088445045609/H2925266(4/29/2027), H2934824(5/5/2027), H2938652(5/12/2027).
Devices named in this recall
49 devices in the FDA device registry whose identifier appears in the recall's code information.
- Merit Medical® 00884450138519Merit Medical Systems, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Merit Medical® 00884450160596Merit Medical Systems, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Merit Medical® 00884450046951Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450047057Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450052921Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450052952Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450138120Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450138298Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450138304Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450138427Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450138588Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450139738Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450140468Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450141458Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450142295Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450142301Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450142448Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450145401Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450145449Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450146538Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450147931Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450148372Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450148884Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450150191Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450151631Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450152324Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450153864Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450158500Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450165881Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Monarch® 00884450236987Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450253083Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450257166Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450257326Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450279014Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450285923Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450290316Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450404492Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450409091Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450432631Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450453629Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450453667Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450462768Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450464533Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450467824Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450480045Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450639115Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450639177Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450639214Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
- Merit Medical® 00884450639429Merit Medical Systems, Inc. · Angiography/angioplasty kit1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95343).
- Z-0064-2025 · Open, ClassifiedBlue Diamond Digital Inflation Device, REF: IN7152/C, IN7352/C, IN7403/C, IN7802/C, IN7852/C, IN7130/C
- Z-0065-2025 · Open, ClassifiedbasixTOUCH Inflation Device, REF: IN8140/A
- Z-0066-2025 · Open, ClassifiedDiamondTOUCH Digital Inflation Device, REF: IN9135/A, IN9152/A
- Z-0067-2025 · Open, ClassifiedIntelliSystem Inflation Device, REF: IN1525/D
- Z-0068-2025 · Open, ClassifiedBlueFIRE Inflation Device, REF: IN10140
- Z-0069-2025 · Open, ClassifiedMONARCH Inflation Device, REF: IN2430/B, IN2530/B, IN2130/B
- Z-0071-2025 · Open, ClassifiedbasixSKY Inflation Device, REF: SKY1802
- Z-0072-2025 · Open, ClassifiedbasixTOUCH Inflation Device, REF: IN8130/JPC, IN8152/C, IN8302/C, IN8352/C, IN8802/C, IN8130/C, IN8130/C
- Z-0073-2025 · Open, ClassifiedStabiliT TOUCH Syringe, REF: IN8VCF/B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page