basixTOUCH recallZ-0072-2025 · Merit Medical Systems, Inc. · initiated Aug 19, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens which could lead to infection.
FDA root cause category: Packaging process control
What the company is doing
On 8/19/24, recall notices were emailed to customers asking them to do the following: 1) If affected devices are discovered, quarantine them, discontinue use and distribution, and return to the recalling firm. 2) Ensure that applicable personnel within your organization are made aware of this field action. 3) If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4) Complete and return the customer response form via email to [email protected] If you have any questions concerning this communication, please contact the firm's Sales Representatives or Customer Service via email at [email protected] or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri.
Recall details
- Recall number
- Z-0072-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 19, 2024
- Posted by the FDA
- Oct 11, 2024
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 16,439
- Product code
- PTM
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of France, Italy, Jamaica, Ireland, Paraguay, Australia, India, Nepal, Japan.
Product description
basixTOUCH Inflation Device, REF: IN8130/JPC, IN8152/C, IN8302/C, IN8352/C, IN8802/C, IN8130/C, IN8130/C
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
REF/UDI-DI/Lot(Expiration): IN8130/JPC/00884450621073/H2929530(4/29/2027). IN8152/C/00884450621301/H2925219(4/22/2027), H2929412(4/29/2027), H2938005(5/12/2027), H2938006(5/12/2027), H2943498(5/19/2027), H2943499(5/19/2027), H2972593(5/26/2027), H2972594(5/26/2027), H2978092(6/3/2027). IN8302/C/00884450621325/H2925221(4/22/2027). IN8352/C/00884450621349/H2934719(5/5/2027). IN8802/C/00884450621417/H2925222(4/22/2027), H2934720(5/12/2027), H2943517(5/19/2027), H2972598(5/26/2027). IN8130/C/00884450621059/H2925215(4/22/2027), H2929410(4/29/2027), H2934714(5/5/2027), H2937998(5/12/2027), H2938000(5/12/2027), H2943487(5/19/2027). IN8130/C/00884450621059/H2972590(5/26/2027), H2978089(6/3/2027), H2978090(6/3/2027), H2978091(6/3/2027), H2989571(6/17/2027).
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- basixTOUCH™ 00884450621417Merit Medical Systems, Inc. · Syringe, Balloon Inflation, Exempt2 recalls0 reports
- basixTOUCH™ 00884450621059Merit Medical Systems, Inc. · Syringe, Balloon Inflation, Exempt1 recall1 report
- basixTOUCH™ 00884450621301Merit Medical Systems, Inc. · Syringe, Balloon Inflation, Exempt1 recall0 reports
- basixTOUCH™ 00884450621325Merit Medical Systems, Inc. · Syringe, Balloon Inflation, Exempt1 recall0 reports
- basixTOUCH™Merit Medical Systems, Inc. · Syringe, Balloon Inflation, Exempt1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95343).
- Z-0064-2025 · Open, ClassifiedBlue Diamond Digital Inflation Device, REF: IN7152/C, IN7352/C, IN7403/C, IN7802/C, IN7852/C, IN7130/C
- Z-0065-2025 · Open, ClassifiedbasixTOUCH Inflation Device, REF: IN8140/A
- Z-0066-2025 · Open, ClassifiedDiamondTOUCH Digital Inflation Device, REF: IN9135/A, IN9152/A
- Z-0067-2025 · Open, ClassifiedIntelliSystem Inflation Device, REF: IN1525/D
- Z-0068-2025 · Open, ClassifiedBlueFIRE Inflation Device, REF: IN10140
- Z-0069-2025 · Open, ClassifiedMONARCH Inflation Device, REF: IN2430/B, IN2530/B, IN2130/B
- Z-0070-2025 · Open, ClassifiedCustom Inflation Kit, REF: K05-00052M, K05-00053H, K05-00109G, K05-00137K, K05-00296L, K05-00399D, K05-00404C, K05-00410G, K05-00438G, K05-00502F, K05-00564L, K05-00587D, K05-00612…
- Z-0071-2025 · Open, ClassifiedbasixSKY Inflation Device, REF: SKY1802
- Z-0073-2025 · Open, ClassifiedStabiliT TOUCH Syringe, REF: IN8VCF/B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page