Skip to content

SOLTIVE recallZ-0135-2025 · Olympus Corporation of the Americas · initiated Aug 13, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

FDA root cause category: Under Investigation by firm

What the company is doing

Olympus issued Urgent Medical Device Correction Letter on 9/12/24 to Operating Room Director, Risk Management. Letter states reason foe recall, health risk and action to take: 1. Carefully read the content of this notification. 2. Please check all areas of your facility to determine if you have the devices specified above. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the Soltive Laser System Instructions for Use. 4. If you have further distributed this product, identify your customers, and forward them this notification. 5. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0452 c. Complete the form as instructed. If you require additional information, please do not hesitate to contact me at [email protected] or by phone at (647) 999-3203

Recall details

Recall number
Z-0135-2025
Status
Open, Classified
Initiated by the company
Aug 13, 2024
Posted by the FDA
Oct 18, 2024
Recalling firm
Olympus Corporation of the Americas
Quantity
159 units
Product code
GEX
FDA submissions
K221306
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Product description

Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS

Code information (models, lots, serial numbers)

UDI 00821925044135 All Serial Numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95380).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page