Skip to content

Cortera recallZ-0125-2025 · XTANT Medical Holdings, Inc · initiated Aug 21, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from XTANT Medical Holdings, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Spinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with released manufacturing specifications, and if implanted the screws may not have adequate strength or performance.

FDA root cause category: Process change control

What the company is doing

On 8/21/24, recall notices were mailed and emailed to customers who were asked to do the following: 1) Examine affected devices and quarantine any screws with DNI markings. 2) Notify your staff about the recall. If affected devices were further distributed, provide the recall notice to these customers. 3) Complete and return the response forms via email to [email protected]

Recall details

Recall number
Z-0125-2025
Status
Open, Classified
Initiated by the company
Aug 21, 2024
Posted by the FDA
Oct 17, 2024
Recalling firm
XTANT Medical Holdings, Inc
Quantity
330
Product code
NKB
FDA submissions
K221403
Distribution
US Nationwide distribution in the states of IL, WI, FL, PA, CA, TX, MI, MO, VA, OH, MA.

Product description

Cortera Spinal Fixation System Screw, Poly, Solid, 6.5mm x 45mm, REF: 150100-65045

Code information (models, lots, serial numbers)

UDI-DI: 00810076563190, Lot: DNI

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page