Skip to content

Cortera 150100-65045Recalls & reported problems

Xtant Medical Holdings, Inc. · Screw, Poly, Solid, 6.5x45 · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810076563190), received between Sep 12, 2024 and Sep 18, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 21, 2024Open, Classified

    Spinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with released manufacturing specifications, and if implanted the screws may not have adequate strength or performance.

    Z-0125-2025 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Labelling, Instructions for Use or Training Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL FIXATION SURGERY INVOLVING THE PEDICLE SCREW SYSTEM. DURING THE SURGERY IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTY WITH SITTING THE ROD INTO THE SCREW. NO PATIENT COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20246892 ↗
  • MalfunctionSep 12, 2024Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    IT WAS REPORTED THAT A PEDICLE SCREW WITH DNI IMPRINTED ON THE SCREW TULIP WAS IMPLANTED. THE SCREW DID NOT LOCK DOWN AND WAS REMOVED AND REPLACED WITH A CLINICAL CORTERA SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20206255 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.