Enterprise Imaging XERO Viewer- software application used… recallZ-0055-2025 · Agfa Healthcare NV · initiated Sep 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Agfa Healthcare NV or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
FDA root cause category: Software design
What the company is doing
AGFA issued Urgent: Mandatory Correction Important Information on 9/10/24. Letter states reason for recall, health risk and action to take: AGFA HealthCare will provide the software corrections for the affected versions. A patch is available for each of the following XERO Viewer version streams: - 8.4.x - 8.3.x - 8.2.x - 8.1.4.x - 8.1.2.x For versions of XERO Viewer prior to 8.1.2, the issue can be prevented from happening at your site by: - Using the XERO Viewer Extended Tools component (Xtend), if enabled. - Using the Enterprise Imaging Desktops viewer. - There is an option to disable the features affected by this issue. This can be configured by your administrator or contact AGFA Support. Our Support organization representative will contact all customers on any of these versions and coordinate the correction activity. The patch installation will not require any downtime. We request that you complete and return the attached mandatory feedback form. Please direct questions or concerns to your AGFA HealthCare Customer Support, and please reference PRB0762354. AGFA HealthCare Customer Support: 1.866.599.3868
Recall details
- Recall number
- Z-0055-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 6, 2024
- Posted by the FDA
- Oct 10, 2024
- Recalling firm
- Agfa Healthcare NV
- Quantity
- 801 units
- Product code
- LLZ
- FDA submissions
- K170434
- Distribution
- US Nationwide distribution. Foreign: Argentina, Australia, Belgium, Chile, Canada, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Georgia, Greece, Hong Kong, Iceland, Ireland, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Nicaragua, Norway, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Tanzania, Tunisia, United Arab Emirates, United Kingdom
Product description
Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000, 8.3.1.000, 8.3.2.000, 8.4.0.000, 8.4.1.000 (including all SU s and HF s)
Code information (models, lots, serial numbers)
UDI-DI: 05400874000710 Software Versions affected: All versions of Enterprise Imaging XERO Viewer are affected 8.1.2 or higher 8.1.4 or higher 8.2.0.100 or higher 8.2.2.000 or higher 8.3.2.000 or higher 8.4.0.000 or higher Enterprise Imaging XERO Viewer main version streams 8.4.0, 8.4.1, 8.3.2, 8.3.1, 8.3.0, 8.2.2, 8.2.1, 8.2.0, 8.1.5, 8.1.4, 8.1.3, 8.1.2, 8.1.1, 8.1, 8.0.1, 8.0.0, ICIS View 2014.1, ICIS View 2014.2
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Enterprise Imaging XERO Viewer 8Agfa HealthCare NV · System, Image Processing, Radiological2 recalls2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page