Enterprise Imaging XERO Viewer 8Recalls & reported problems

Agfa HealthCare NV · Enterprise Imaging XERO Viewer 8.x · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05400874000710), received between Aug 27, 2024 and Aug 28, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 6, 2024Open, Classified

    Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.

    Z-0055-2025 · Root cause: Software design · Recall details · FDA record ↗

  2. Dec 5, 2022Open, Classified

    There is a software defect that can cause issues with images.

    Z-1004-2023 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Problem: Parameter Calculation Error2

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error

    ON (B)(6) 2024, A CUSTOMER REPORTED TO AGFA HEALTHCARE THAT WHILE USING THE XERO VIEWER, REFERENCE/LOCALIZER LINE ON THE CORONAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER VIEWPORTS, WAS OFFSET BY 20MM FROM THE ACTUAL POSITION. AS A RESULT OF THIS ISSUE, A SPINAL SURGERY WAS PERFORMED ON THE WRONG VERTEBRA. EVENT DETAILS: ON (B)(6) 2024 AT 12:37, A PATIENT UNDERWENT A CT OF THE LUMBAR SPINE BY REQUEST OF THE NEUROSURGEON TO DETERMINE THE SOURCE OF LOWER BACK PAIN THAT WAS RADIATING INTO THE RIGHT LEG. THE STUDY WAS REPORTED ON (B)(6) 2024 AT 13:00 BY THE RADIOLOGIST WHO IDENTIFIED A RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL AND A POSSIBLE CONFLICT WITH THE L5 NERVE ROOT. THAT SAME DAY THE PATIENT UNDERWENT A POSTERIOR FORAMINOTOMY AT L4/L5 WHICH IS IN CONTRAST TO WHAT THE RADIOLOGIST REPORT IDENTIFIED IN RELATION TO THE RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL. THIS WAS DUE TO REFERENCE LINE ON THE LATERAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER IMAGE VIEWPORTS, WAS OFFSET 20MM (AT L4/L5) FROM THE ACTUAL CORRECT POSITION (AT L5/S1). AS THE PATIENT'S LOWER BACK PAIN DID NOT SUBSIDE, AN ADDITION

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20084885 ↗
  • InjuryAug 27, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error

    ON AUGUST 7 2024, A CUSTOMER REPORTED TO AGFA HEALTHCARE THAT WHILE USING THE XERO VIEWER, REFERENCE/LOCALIZER LINE ON THE CORONAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER VIEWPORTS, WAS OFFSET BY 20MM FROM THE ACTUAL POSITION. AS A RESULT OF THIS ISSUE, A SPINAL SURGERY WAS PERFORMED ON THE WRONG VERTEBRA. EVENT DETAILS: ON (B)(6) 2024 AT 12:37, A PATIENT UNDERWENT A CT OF THE LUMBAR SPINE BY REQUEST OF THE NEUROSURGEON TO DETERMINE THE SOURCE OF LOWER BACK PAIN THAT WAS RADIATING INTO THE RIGHT LEG. THE STUDY WAS REPORTED ON JULY 24, 2024 AT 13:00 BY THE RADIOLOGIST WHO IDENTIFIED A RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL AND A POSSIBLE CONFLICT WITH THE L5 NERVE ROOT. THAT SAME DAY THE PATIENT UNDERWENT A POSTERIOR FORAMINOTOMY AT L4/L5 WHICH IS IN CONTRAST TO WHAT THE RADIOLOGIST REPORT IDENTIFIED IN RELATION TO THE RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL. THIS WAS DUE TO REFERENCE LINE ON THE LATERAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER IMAGE VIEWPORTS, WAS OFFSET 20MM (AT L4/L5) FROM THE ACTUAL CORRECT POSITION (AT L5/S1). AS THE PATIENT'S LOWER BACK PAIN DID NOT SUBSIDE, AN ADDI

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20080238 ↗

Other System, Image Processing, Radiological devices