Enterprise Imaging XERO Viewer 8Recalls & reported problems
Agfa HealthCare NV · Enterprise Imaging XERO Viewer 8.x · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05400874000710), received between Aug 27, 2024 and Aug 28, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Sep 6, 2024Open, Classified
Software defect, when an image has non-square pixel spacing, the reference/localizer line is incorrectly placed on the scout image in the XERO Viewet, may display the incorrect crosshair/line position on the non-coplanar/scout image with respect to the axial slice location. Inaccurate placement of reference lines or crosshairs have potential risk of misdiagnosis or mistreatment due to unintended shifts in the display of anatomical locations.
Z-0055-2025 · Root cause: Software design · Recall details · FDA record ↗
- Dec 5, 2022Open, Classified
There is a software defect that can cause issues with images.
Z-1004-2023 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Problem: Parameter Calculation Error2
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error
ON (B)(6) 2024, A CUSTOMER REPORTED TO AGFA HEALTHCARE THAT WHILE USING THE XERO VIEWER, REFERENCE/LOCALIZER LINE ON THE CORONAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER VIEWPORTS, WAS OFFSET BY 20MM FROM THE ACTUAL POSITION. AS A RESULT OF THIS ISSUE, A SPINAL SURGERY WAS PERFORMED ON THE WRONG VERTEBRA. EVENT DETAILS: ON (B)(6) 2024 AT 12:37, A PATIENT UNDERWENT A CT OF THE LUMBAR SPINE BY REQUEST OF THE NEUROSURGEON TO DETERMINE THE SOURCE OF LOWER BACK PAIN THAT WAS RADIATING INTO THE RIGHT LEG. THE STUDY WAS REPORTED ON (B)(6) 2024 AT 13:00 BY THE RADIOLOGIST WHO IDENTIFIED A RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL AND A POSSIBLE CONFLICT WITH THE L5 NERVE ROOT. THAT SAME DAY THE PATIENT UNDERWENT A POSTERIOR FORAMINOTOMY AT L4/L5 WHICH IS IN CONTRAST TO WHAT THE RADIOLOGIST REPORT IDENTIFIED IN RELATION TO THE RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL. THIS WAS DUE TO REFERENCE LINE ON THE LATERAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER IMAGE VIEWPORTS, WAS OFFSET 20MM (AT L4/L5) FROM THE ACTUAL CORRECT POSITION (AT L5/S1). AS THE PATIENT'S LOWER BACK PAIN DID NOT SUBSIDE, AN ADDITION
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20084885 ↗InjuryAug 27, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error
ON AUGUST 7 2024, A CUSTOMER REPORTED TO AGFA HEALTHCARE THAT WHILE USING THE XERO VIEWER, REFERENCE/LOCALIZER LINE ON THE CORONAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER VIEWPORTS, WAS OFFSET BY 20MM FROM THE ACTUAL POSITION. AS A RESULT OF THIS ISSUE, A SPINAL SURGERY WAS PERFORMED ON THE WRONG VERTEBRA. EVENT DETAILS: ON (B)(6) 2024 AT 12:37, A PATIENT UNDERWENT A CT OF THE LUMBAR SPINE BY REQUEST OF THE NEUROSURGEON TO DETERMINE THE SOURCE OF LOWER BACK PAIN THAT WAS RADIATING INTO THE RIGHT LEG. THE STUDY WAS REPORTED ON JULY 24, 2024 AT 13:00 BY THE RADIOLOGIST WHO IDENTIFIED A RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL AND A POSSIBLE CONFLICT WITH THE L5 NERVE ROOT. THAT SAME DAY THE PATIENT UNDERWENT A POSTERIOR FORAMINOTOMY AT L4/L5 WHICH IS IN CONTRAST TO WHAT THE RADIOLOGIST REPORT IDENTIFIED IN RELATION TO THE RIGHT SIDED FORAMINAL EXTRUSION AT THE L5/S1 LEVEL. THIS WAS DUE TO REFERENCE LINE ON THE LATERAL SCOUT IMAGE, REPRESENTING THE AXIAL CT SLICE POSITION DISPLAYED ON THE OTHER IMAGE VIEWPORTS, WAS OFFSET 20MM (AT L4/L5) FROM THE ACTUAL CORRECT POSITION (AT L5/S1). AS THE PATIENT'S LOWER BACK PAIN DID NOT SUBSIDE, AN ADDI
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20080238 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page