Quantum Perfusion Systems for CPB, Roller Pumps use… recallZ-0224-2025 · Spectrum Medical Ltd. · initiated Mar 13, 2023
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spectrum Medical Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
FDA root cause category: Component design/selection
What the company is doing
All potentially affected product was identified. The firm corrected all products between 03/13/2024 and 08/07/2024 during service visits.
Recall details
- Recall number
- Z-0224-2025
- Status
- Completed
- Initiated by the company
- Mar 13, 2023
- Posted by the FDA
- Oct 24, 2024
- Recalling firm
- Spectrum Medical Ltd.
- Quantity
- 172 units
- Product code
- DTQ
- FDA submissions
- K173834
- Distribution
- US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.
Product description
Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI/DI 05060434420770: Serial Numbers: a) AU6004040, AU6004041, AU6004042, AU6004043, AU6004044, AU6004046, AU6004050, AU6004051, AU6004052, AU6004053, AU6004054, AU6004056, AU6004057, AU6004058, AU6004060, AU6004061, AU6004062, AU6004063, AU6004064, AU6004065, AU6004066, AU6004069, AU6004070, AU6004071, AU6004072, AU6004073, AU6004074, AU6004075, AU6004136, AU6004137, AU6004138, AU6004139, AU6004140, AU6004141, AU6004142, AU6004143, AU6004144, AU6004145, AU6004146, AU6004147, AU6004148, AU6004159, AU6004160, AU6004161, AU6004162, AU6004167, AU6004168, AU6004169, AU6004170, AU6004171, AU6004172, AU6004173, AU6004174, AU6004175, AU6004176, AU6004177, AU6004178, AU6004179, AU6004180, AU6004181, AU6004182, AU6004183, AU6004184, AU6004185, AU6004186, AU6004187, AU6004188, AU6004195, AU6004197, AU6004198, AU6004200, AU6004201, AU6004202, AU6004203, AU6004204, AU6004205, AU6004207, AU6004332, AU6004333, AU6004348, AU6004349, AU6004350, AU6004352, AU6004353, AU6004354, AU6004532, AU6004577, AU6004578, AU6004579, AU6004580, AU6004587, AU6004588, AU6004590, AU6004591, AU6004603, AU6004604, AU6004605, AU6004606, AU6004607, AU6004609, AU6004610, AU6004613, AU6004731, AU6004771, AU6004773, AU6004776, AU6004777, AU6004778, AU6004779, AU6004781, AU6004782, AU6004919, AU6004920, AU6004921, AU6004922, AU6004923, AU6004924, AU6004925, AU6004926, AU6004927, AU6004929, AU6004930, AU6004955, AU6004957, AU6004958, AU6004960, AU6004963, AU6004966, AU6004968, AU6004970, AU6004971, AU6004972, AU6005010, AU6005021, AU6005022, AU6005023, AU6005024, AU6005025, AU6005026, AU6005027, AU6005028, AU6005029, AU6005030, AU6005037, AU6005038, AU6005039, AU6005040, AU6005041, AU6005042, AU6005057, AU6005065, AU6005109, AU6005110, AU6005118, AU6005124, AU6005125, AU6005355, AU6006351; Serial Numbers: b) AU8001273, AU8001275, AU8001346, AU8001347, AU8001348, AU8001359, AU8001360, AU8001366, AU8001403, AU8001406, AU8001407, AU8001446, AU8001447, AU8001449
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Quantum QRP8Spectrum Medical Limited · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall24 reports
Other devices under the same FDA submission
4 devices cleared under K173834 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Quantum QRP6Spectrum Medical Limited · Console, Heart-Lung Machine, Cardiopulmonary Bypass66 reports
- Quantum QRP4Spectrum Medical Limited · Console, Heart-Lung Machine, Cardiopulmonary Bypass29 reports
- Quantum ® QPF2Spectrum Medical Limited · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 report
- Quantum QPSSpectrum Medical Limited · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page