Spectrum Medical Limited · Quantum Roller Pump - 4 inch · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

The Quantum QRP4 is a 4-inch roller pump console used to drive blood flow during cardiopulmonary bypass surgery. Its 34 reports mostly describe the pump stopping or failing to pump mid-procedure, sometimes with hardware errors, dropped connections, or a pump that would not roll. None of the reports mention death or injury, and no recalls name this device. Reports over the past 12 months totaled 10, down from 21 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 05060434420794), received between Jul 26, 2024 and Feb 25, 2026 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (12 reports).

2 other recalls cover devices cleared under the same FDA submission (K173834); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 3 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Pumping Stopped12
  • Pumping Problem6
  • Mechanical Problem3
  • Appropriate Device Problem Term/Code Not Available2
  • Insufficient Device Problem Information2
  • Mechanical Jam2
  • Circuit Failure1
  • Damaged Thread1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2026Pumping Stopped

    USER REPORTED THAT PUMP WAS NOT ROLLING DURING PROCEDURE. THIS INTERRUPTION IN BLOOD FLOW IS CONSIDERED REPORTABLE PER SPECTRUM MEDICAL'S CRITERIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24443280 ↗
  • MalfunctionFeb 3, 2026Mechanical Problem

    USER CONTACTED SPECTRUM TO REPORT A FAULTY VENT PUMPT. WHEN CONNECTED TO THE FRAME, THE PUMP TILE DISPLAYED "DÉFAILLANCE MATÉRIEL" (HARDWARE FAILURE) . THE USER FIRST TESTED WITH ANOTHER CABLE, THEN TOOK THE CEREBRO PUMP AND CONNECTED IT IN PLACE OF, WHICH ALLOWED THEM TO CONTINUE THE CASE. THERE WAS NO PATIENT HARM OR DELAY TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246342 ↗
  • MalfunctionDec 18, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    USER REPORTED THAT INTERMITTENT CONNECTIVITY ISSUES CAUSING THE PUMP TO STOP DURING THE PROCEDURE. THERE WAS NO PATIENT HARM; HOWEVER, UNINTENDED PUMP STOPS ARE CONSIDERED REPORTABLE BY SPECTRUM MEDICAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838246 ↗
  • MalfunctionDec 15, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ON SITE VISIT CONFIRMED FAULTY LED ON A ROLLER PUMP, FOLLOWING A COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23802882 ↗
  • MalfunctionDec 11, 2025Pumping Problem; Pumping Problem

    USER REPORTED THAT THE CARDIOPLEGIA PUMP DID NOT OPERATE CORRECTLY. THERE WAS NO PATIENT HARM AND THE USER RESOLVED THE ISSUE BY UNPLUGGING AND PLUGGING THE PUMP SAP CABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23780414 ↗
  • MalfunctionOct 22, 2025Failure to Pump; Failure to Pump

    USER REPORTED THAT A PUMP SHOWED A SPEED ERROR MESSAGE AND FAILED TO ROTATE. THE USER WAS UNABLE TO TROUBLESHOOT, BEFORE SWAPPING TO A BACKUP DEVICE TO SUCCESSFULLY COMPLETE THE CASE. WE CONSIDER INABILITY TO PUMP TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23355639 ↗
  • MalfunctionSep 4, 2025Pumping Problem; Pumping Problem

    USER REPORTED THAT THE BOBBIN ON THE PUMP IS STRIPPED AND WOULD NOT MAINTAIN A LOCKED POSITION. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22964537 ↗
  • MalfunctionAug 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING PREVENTATIVE MAINTENANCE IT NOTED THAT THE MODULE HAD AN INTERNAL CABLE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22893443 ↗
  • MalfunctionAug 26, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING PREVENTATIVE MAINTENANCE IT NOTED THAT THE MODULE HAD AN INTERNAL CABLE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22893306 ↗
  • MalfunctionAug 14, 2025Pumping Stopped; Pumping Stopped

    USER REPORTED THAT THE BOBBIN OF THE ROLLER PUMP WAS VERY STIFF BEFORE BYPASS BUT CONTINUED TO BE USED BY TEAM. USER REPORTED THAT THEY HAD DIFFICULTY SETTING THE TUBING IN CORRECT RACEWAY DUE TO STIFF BOBBIN DURING THE CASE THE TUBING "RUCKED UP " AND SPILT CAUSING AN INTERRUPTION TO THE BLOOD FLOW. THERE WAS NO PATIENT HARM AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22802521 ↗
  • MalfunctionJul 30, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    USER REPORTED THAT LOOSE BOBBINS, WHICH PREVENTED THE PUMP FROM MAINTAINING A LOCKED POSITION. THERE WAS NO PATIENT HARM AS THE ERROR WAS IDENTIFIED DURING SET-UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22669015 ↗
  • MalfunctionJul 17, 2025Mechanical Problem; Mechanical Problem

    DURING PREVENTATIVE MAINTENANCE, IT WAS DISCOVERED THAT THE LOCKED POSITION OF THE ROLLER PUMP DID NOT HOLD THE POSITION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22534677 ↗
  • MalfunctionJul 16, 2025Mechanical Problem; Mechanical Problem

    USER REPORTED THAT THE BOBBIN LOCKING MECHANISM ON THE ROLLER PUMP NOT WORKING CORRECTLY. THERE WAS NO PATIENT INVOLVEMENT AND THEREFORE NO RISK OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22521068 ↗
  • MalfunctionJun 18, 2025Pumping Stopped; Pumping Stopped

    DURING A CASE IT WAS REPORTED THAT THERE WAS A HARDWARE FAULT MESSAGE AND A FLASHING RED/YELLOW LIGHT ON THE PUMP. A BACK UP WAS USED TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER BLOOD FLOW INTERRUPTIONS TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22282325 ↗
  • MalfunctionJun 12, 2025Pumping Stopped; Pumping Stopped

    USER REPORTED THAT DURING A PROCEDURE A QUANTUM ROLLER PUMP STOPPED WORKING. THE USER CHOSE TO REPLACE THE QUANTUM ROLLER PUMP. THE REST OF THE WAS COMPLETED WITHOUT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22193933 ↗
  • MalfunctionMay 8, 2025Pumping Problem; Pumping Problem

    USER REPORTED PUMP FAILURE WHILE DELIVERING CARDIOPLEGIA TO PATIENT. THERE WAS NO PATIENT HARM; HOWEVER, THIS EVENT IS CONSIDERED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21989371 ↗
  • MalfunctionApr 28, 2025Pumping Stopped; Pumping Stopped

    USER REPORTED ROLLER FAILURE THAT CAUSED TUBING TO BUNCH. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919228 ↗
  • MalfunctionApr 23, 2025Mechanical Jam; Mechanical Jam

    USER REPORTED ROLLER PUMP JAM WHEN TRYING TO LOCK TUBING INTO PLACE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21890845 ↗
  • MalfunctionFeb 10, 2025Pumping Stopped; Pumping Stopped

    DURING PREVENTATIVE MAINTENANCE, IT WAS FOUND THAT THE QUANTUM ROLLER PUMP HAD AN INTERNAL CABLE ISSUE, CAUSING IT TO STOP DURING USE. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21344244 ↗
  • MalfunctionJan 28, 2025Pumping Stopped; Pumping Stopped

    DURING A CASE THE RP (B)(6) THAT IS A CARDIOPLEGIA PUMP STOPPED TO RUN AND SHOWED THE FOLLOWING ALARM:"SPEED ERROR". THE CLINICIAN SWAPPED WITH ANOTHER 4" PUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21251061 ↗
  • MalfunctionJan 28, 2025Pumping Stopped; Pumping Stopped

    DURING PROCEDURE, PERFUSIONIST RECEIVED A "SPEED ERROR" ALARM AND THE PUMP STOPPED. USER REPLACED THE PUMP AND THE CASE CONTINUED. THERE WAS NO PATIENT HARM. (B)(6) CONSIDERS ANY EVENT IN WHICH THE PUMP STOPS TO BE REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21250216 ↗
  • MalfunctionJan 15, 2025Pumping Stopped; Pumping Stopped

    USER REPORTED THAT A ROLLER PUMP ABRUPTLY STOPPED DURING A CASE AND THE USER WAS UNABLE TO RESTART. USER REPORTED A "STRANGE NOISE" BEFORE IT STOPPED. PUMP STOP IS CONSIDERED REPORTABLE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21157968 ↗
  • MalfunctionDec 6, 2024Pumping Stopped; Pumping Stopped

    PERFUSIONIST REPORTED THAT A ROLLER PUMP STOPPAGE DURING A CASE. THE PERFUSIONIST USED A HAND CRANK TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER PUMP STOPPAGES TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20862453 ↗
  • MalfunctionDec 4, 2024Pumping Stopped; Pumping Problem; Pumping Stopped; Pumping Problem

    PERFUSIONIST REPORTED, THAT A ROLLER PUMP STOPPAGE, DURING A CASE. A BACK UP WAS USED TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER, PUMP STOPPAGES TO BE REPORTABLE. DESPITE, NO HARM TO THE PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20843696 ↗
  • MalfunctionDec 2, 2024Pumping Problem; Pumping Problem

    PERFUSIONIST REPORTED THAT THE PUMP WOULD NOT STOP WHEN INSTRUCTED VIA THE KNOB OR TOUCHSCREEN. A BACK UP WAS USED TO COMPLETE THE CASE SUCCESSFULLY. WE CONSIDER INABILITY TO CONTROL A PUMP TO BE REPORTABLE, DESPITE NO HARM TO THE PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20816771 ↗

FDA clearance and classification

Cleared through 510(k) K173834 on May 25, 2018 (decision: Substantially Equivalent). Submitted by Spectrum Medical , Ltd. as “Quantum Pump Console”. The FDA review took 158 days. FDA record ↗

The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K173834. One submission often covers many models, so check the product and lot details in each record.

  • Z-0224-2025 · Mar 13, 2023 · Completed
    Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-0000
  • Z-2238-2019 · May 29, 2019 · Terminated
    Quantum Pump Console

Common questions about the Quantum QRP4

Has the Quantum QRP4 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05060434420794), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Quantum QRP4?

The FDA's MAUDE database lists 29 adverse event reports for this device, received between Jul 26, 2024 and Feb 25, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (12), “Pumping Problem” (6) and “Mechanical Problem” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Quantum QRP4?

None of the reports are classified as deaths, 0 as injuries and 29 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Quantum QRP4 FDA approved?

It was cleared through 510(k) K173834 on May 25, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Quantum QRP4, and is it still sold?

The FDA's device database lists Spectrum Medical Limited as the company responsible for it (brand Quantum, model QRP4). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices

All Console, Heart-Lung Machine, Cardiopulmonary Bypass devices →