Skip to content

Alinity m recallZ-0234-2025 · Abbott Molecular, Inc. · initiated Sep 12, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Molecular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.

FDA root cause category: Device Design

What the company is doing

Abbott issued an "Urgent Field Safety Notice" dated 9/11/2024 to its consignees on 09/12/2024 via FedEx priority overnight express. The notice explained the issue, potential impact, and provided the following information: "Necessary Actions " If you experience an increased incidence of Error Code 9198 (Positive control is non-reactive) while testing the Alinity m HR HPVor Alinity m STI assay, please contact Abbott Technical Support at 1-800-553-7042 Option 2 for additional troubleshooting information. " Complete and return the associated Customer Reply Form. If you have forwarded this product to any other laboratories, please also forward this letter and customer reply form to that laboratory. Review this information with laboratory personnel and retain this communication for future reference." If you have any questions regarding this communication, please contact your local Abbott representative or call: 1-224-361-7000.

Recall details

Recall number
Z-0234-2025
Status
Open, Classified
Initiated by the company
Sep 12, 2024
Posted by the FDA
Oct 29, 2024
Recalling firm
Abbott Molecular, Inc.
Quantity
15068 units
Product code
QEP
FDA submissions
K222379
Distribution
Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA, IN, IL, LA, MA, MI, MN, MO, NC, NE, NJ, NY, OR, PA, SC, TN, TX, VA, WA, and WI; and foreign (OUS) countries of: AUSTRIA, BELGUIM, BULGARIA, CAMBODIA, CANADA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REP., EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, KAZAKHSTAN, KENYA, LUXEMBOURG, MALAWI, MALAYSIA, MONTENEGRO, MOROCCO, NETHERLANDS, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDIA ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TANZANIA, THAILAND, UNITED KINGDOM, AND VIETNAM.

Product description

Abbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;

Code information (models, lots, serial numbers)

a) REF 09N17-095, UDI/DI 00884999048591; b) REF 09N17-090, UDI/DI 00884999047945; c) REF 09N17-091, UDI/DI 00884999049277 All Lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95442).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page