Medtronic Percept RC Implantable Neurostimulators (INS)… recallZ-0206-2025 · Medtronic Neuromodulation · initiated Sep 24, 2024

Class II Open, Classified All recalls posted in October 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A limited number of Percept RC Implantable Neurostimulators (INS), Model B35300 will not be able to communicate with the A610 DBS Clinician Programmer Application upon use and an "Invalid Device" message will appear.

FDA root cause category: Under Investigation by firm

What the company is doing

Medtronic neuromodulation issued an URGENT MEDICAL DEVICE RECALL notice to its consignees of 09/24/2024 via letter. The notice explained the issue, potential risk, and requested the affected devices in stock be returned. The notice is to be shared with those to whom the device was distributed. If the device has been implanted, the consignee was directed to contact their Medtronic Representative to determine if there are additional steps that may establish communication with the INS. On 07-Oct-24 and 08-Oct-24, a Medtronic representative successfully performed an update using a manufacturing software tool to reestablish communication to the affected devices. Consignees were provided a consignee notification regarding this action.

Recall details

Recall number
Z-0206-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 24, 2024
Posted by the FDA
Oct 23, 2024
Recalling firm
Medtronic Neuromodulation
Quantity
39 units
Product code
MHY
FDA submissions
P960009
Distribution
International distribution to the countries of Germany, UK, and Israel.

Product description

Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.

Code information (models, lots, serial numbers)

GTIN 00763000813468, Serial Numbers: NND504236H, NND504252H; GTIN 00763000813475, Serial Numbers: NND504165H, NND504169H, NND504174H, NND504181H, NND504208H, NND504212H, NND504215H, NND504216H, NND504217H, NND504218H, NND504221H, NND504222H, NND504223H, NND504224H, NND504229H, NND504230H, NND504232H, NND504234H, NND504235H, NND504243H, NND504255H, NND504156H, NND504157H, NND504159H, NND504160H; GTIN 00763000813529, Serial Numbers: NND504167H, NND504178H, NND504179H, NND504180H, NND504182H, NND504227H, NND504228H; GTIN A7630008135201, Serial Numbers: NND504167H, NND504178H, NND504180H; GTIN A7630008134701, Serial Numbers: NND504181H, NND504243H

Other devices under the same FDA submission

90 devices cleared under P960009 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.