Neptune E-SEP recallZ-0367-2025 · Stryker Corporation · initiated Oct 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
FDA root cause category: Under Investigation by firm
What the company is doing
Stryker issued Urgent Medical Device Recall Letter on 10/04/24 via Fed'X to Risk Manager, Materials Manager, OR Director. Letter states reason for recall, health risk and action to take: 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. On behalf of Stryker, we thank you sincerely for your help and support in completing this action and regret any inconvenience that may be caused. We would like to reassure you that Stryker is committed to ensuring that only conforming devices, meeting our high internal quality standards and your expectations, remain on the market. If you have any questions about this matter, contact Stryker Instruments recall team directly by calling (269) 290-0524 or email at [email protected]. A Medical Device Recall Expansion notification letter dated 9/2/25 was sent to customers. Additional Lots A total of 47 lot numbers across two catalog numbers are listed below. Distribution dates in scope range from 03/29/2023 05/17/2024. Actions Needed 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return a new copy of the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A re
Recall details
- Recall number
- Z-0367-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 4, 2024
- Posted by the FDA
- Nov 13, 2024
- Recalling firm
- Stryker Corporation
- Quantity
- 2,620 eaches
- Product code
- GEI
- FDA submissions
- K143145
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
Product description
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001
Code information (models, lots, serial numbers)
GTIN: 07613327117219 Lot Number: 64117291
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Neptune E-SEP 0703047001Stryker Corporation · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95502).
- Z-0365-2025 · Open, ClassifiedStryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing s…
- Z-0366-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch, Bulk Non-Sterile - Designed for general electrosurgical applications including cutting and coagulation and for…
- Z-0368-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and fo…
- Z-0369-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for rem…
- Z-0370-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for r…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page