Neptune E-SEP recallZ-0365-2025 · Stryker Corporation · initiated Oct 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
FDA root cause category: Under Investigation by firm
What the company is doing
Stryker issued Urgent Medical Device Recall Letter on 10/04/24 via Fed'X to Risk Manager, Materials Manager, OR Director. Letter states reason for recall, health risk and action to take: 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. On behalf of Stryker, we thank you sincerely for your help and support in completing this action and regret any inconvenience that may be caused. We would like to reassure you that Stryker is committed to ensuring that only conforming devices, meeting our high internal quality standards and your expectations, remain on the market. If you have any questions about this matter, contact Stryker Instruments recall team directly by calling (269) 290-0524 or email at [email protected]. A Medical Device Recall Expansion notification letter dated 9/2/25 was sent to customers. Additional Lots A total of 47 lot numbers across two catalog numbers are listed below. Distribution dates in scope range from 03/29/2023 05/17/2024. Actions Needed 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return a new copy of the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A re
Recall details
- Recall number
- Z-0365-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 4, 2024
- Posted by the FDA
- Nov 13, 2024
- Recalling firm
- Stryker Corporation
- Quantity
- 1,120,860 eaches
- Product code
- GEI
- FDA submissions
- K143145
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
Product description
Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-000
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI-DI: 07613327117202 Lot Numbers: 53863666 54409859 56319954 56945653 57699375 59053479 65178945 53863667 54409860 56319955 56945654 57699376 59053480 65178946 53863668 54409862 56319956 57699353 57699377 59053481 65178947 53863669 54409919 56319957 57699354 57699378 59053482 65178949 53863670 54409920 56319958 57699355 57699379 59053483 65347997 53863671 54409921 56319959 57699359 57699380 61288937 65784636 54403959 54409922 56319960 57699360 57699381 61288943 65784637 54403960 54409923 56319961 57699361 57699382 62825781 65784638 54403961 54409924 56892676 57699362 57805609 64117282 65784639 54403962 54409925 56892677 57699363 57805610 64477143 65876797 54403963 54409926 56892689 57699364 57805611 64477144 65961918 54403964 54409927 56932527 57699365 58320366 64477145 66133414 54403965 54533880 56932528 57699366 58514707 64855504 66275138 54409851 54533881 56932533 57699367 58766534 64855505 V53051139 54409852 54533882 56932534 57699368 58766535 64855515 V53051140 54409853 54533883 56945645 57699369 58766536 64855516 V53267506 54409854 54533884 56945646 57699370 58766539 64973261 V53526100 54409855 54654973 56945647 57699371 58766540 64973262 V53708277 54409856 54665296 56945648 57699372 58791419 64973263 V53708278 54409857 56319952 56945651 57699373 59053477 64973266 V53934891 54409858 56319953 56945652 57699374 59053478 64973267 V53934892 V53934893 Additional Lot Numbers: 64262566 66387346 66599251 66758654 67377504 67652611 67794514 64973268 66387347 66705549 66758655 67377505 67652612 68735135 65178952 66387348 66705550 67017675 67377506 67652613 68735136 65178953 66387349 66758652 67017676 67481713 67794512 68735137 66387345 66599250 66758653 67207470 67652610 67794513 68735138
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Neptune E-SEP 0703047000Stryker Corporation · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 recalls131 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95502).
- Z-0366-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch, Bulk Non-Sterile - Designed for general electrosurgical applications including cutting and coagulation and for…
- Z-0367-2025 · Open, ClassifiedStryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removi…
- Z-0368-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and fo…
- Z-0369-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for rem…
- Z-0370-2025 · Open, ClassifiedStryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for r…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page