Skip to content

SMARTFREEZE recallZ-0303-2025 · Boston Scientific Corporation · initiated Oct 10, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.

FDA root cause category: Device Design

What the company is doing

Boston Scientific disseminated a letter to Physicians/Healthcare Professionals on 10/10/2024 via courier. The letter notified its consignees of updates to the instructions for use (IFU) for the POLARx and POLARx FIT Cryoablation Catheter related to the risk of atrioesophageal (AE) fistula. The firm requested that the information be shared with clinicians who use the use the Boston Scientific POLARx Cryoablation System, including the POLARx Catheter, the POLARx FIT Catheter and the SMARTFREEZE Console, and to share the communication with any other organization to which these devices may have been transferred. Users were directed to immediately post the information on or near the product to ensure this information is easily accessible to all users of the device.

Recall details

Recall number
Z-0303-2025
Status
Open, Classified
Initiated by the company
Oct 10, 2024
Posted by the FDA
Nov 12, 2024
Recalling firm
Boston Scientific Corporation
Quantity
1035 units
Product code
OAE
FDA submissions
P220032
Distribution
Worldwide

Product description

Boston Scientific SMARTFREEZE CRYO CONSOLE: Console: Material Number M004CRBS4000; CONSOLE REFURB: Material Number M004CRBS400R0; CONSOLE HOSPITAL, Material Number M004CRBS400H0, CONSOLE ZERO, Material Number M004CRBS400Z0

Code information (models, lots, serial numbers)

GTIN 08714729992691

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under P220032 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95514).

Sources and updates