MiniCap recallZ-0373-2025 · Baxter Healthcare Corporation · initiated Oct 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Baxter issued an "Important Medical Device Correction" notice to consignees beginning on 10/21/2024 via letter. In February 2025, Baxter Kidney Care segment launched a standalone company, Vantive. Vantive issued an updated notice to consignees beginning on 2/19/25 via letter. The notice explained the issue, hazard involved, and requested the following actions be taken: For patients older than 6 months of age, healthcare providers are advised to continue using both the peroxide-cured silicone tubing version and/or the platinum-cured silicone tubing version of the MiniCap Extended Life PD transfer sets, as neither is anticipated to present safety risks related to PCB/PCBAs. 2. For patients younger than 6 months: a. Healthcare providers should prioritize the use of platinum-cured silicone tubing sets or other alternatives that may be available. b. If no alternatives are available, healthcare providers should continue to connect the peroxide-cured silicone tubing sets and prioritize the use of shorter transfer sets. c. If the patient currently has the peroxide-cured silicone tubing, they do not need to have it replaced early, as data demonstrates that PCBA levels decrease over treatment time. 3. If you received this communication directly from Vantive, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. 4. If you purchased this product from a distributor, please note that responding via the Vantive customer portal is not applicable. 5. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers.
Recall details
- Recall number
- Z-0373-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 21, 2024
- Posted by the FDA
- Nov 20, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 2,706,558 units
- Product code
- KDJ
- FDA submissions
- K152675
- Distribution
- Worldwide Distribution (US Nationwide distribution)
Product description
Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482; use in Peritoneal Dialysis
Code information (models, lots, serial numbers)
UDI/DI 00085412007731, Lot/Serial Numbers: All lots including and manufactured after H19I26088
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MiniCap 5C4482Baxter Healthcare Corporation · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE3 recalls5 deaths2,961 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K152675 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LOCKING TITANIUM ADAPTER FOR PERITONEAL DIALYSIS CATHETER 5C4129Vantive US Healthcare LLC · SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95525).
- Z-0376-2025 · Open, ClassifiedBaxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4483; use in Peritoneal Dialysis
- Z-0377-2025 · Open, ClassifiedBaxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number R5C4482; use in Peritoneal Dialysis
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.