We found 3 models of MiniCap by Baxter Healthcare Corporation in the FDA's device registry (GUDID). Together they are named in 3,423 adverse event reports in the FDA's MAUDE database, received between Jan 2, 2018 and Sep 22, 2026 — classified as 5 deaths, 86 injuries, 3,332 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1,234 reports).
7 FDA recall records name at least one of these models by its device identifier, 6 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
MiniCap recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 23, 2024Open, Classified
There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
Z-0563-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 23, 2024Open, Classified
There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
Z-0562-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 23, 2024Open, Classified
There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
Z-0559-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 21, 2024Open, Classified
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
Z-0377-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Oct 21, 2024Open, Classified
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
Z-0376-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Oct 21, 2024Open, Classified
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
Z-0373-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Nov 10, 2020Terminated
Potential for no-flow and leaks under the twist clamp.
Z-0768-2021 · Root cause: Process change control · Recall details · FDA record ↗
Most reported MiniCap problems
Device problems
- Detachment of Device or Device Component1,234
- Fluid/Blood Leak719
- Disconnection475
- Crack288
- Difficult to Open or Close223
- Failure to Disconnect222
- Break138
- Connection Problem128
Patient problems
- No Clinical Signs, Symptoms or Conditions2,861
- No Consequences Or Impact To Patient364
- No Patient Involvement105
- Peritonitis83
- Abdominal Pain14
- Unspecified Infection10
- Sepsis3
- Bacterial Infection2
MiniCap models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- MiniCap 5C4482SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE3 recalls5 deaths2,961 reports
- MiniCap 5C4483SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE4 recalls433 reports
- MiniCap EZPC4466SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE29 reports
Recent MiniCap adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Break
It was reported that the twist clamp of a Minicap Transfer Set broke; described as the patient ¿had a transfer set change recently where the white twist part broke¿. This was identified during multiple steps of peritoneal dialysis therapy. The transfer set was replaced. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50854526 ↗MalfunctionJun 25, 2026Free or Unrestricted Flow
IT WAS REPORTED THAT THE MINICAP TRANSFER SET LEAKED AFTER THE ¿SWITCH¿ WAS TIGHTENED AND IN PLACE. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THE TRANSFER SET HAD BEEN IN USE FOR THREE MONTHS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25620288 ↗MalfunctionJun 10, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A MINICAP TRANSFER SET LEAKED AT THE TWIST CLAMP. THIS OCCURRED DURING THE FIRST EXCHANGE OF PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25460910 ↗MalfunctionMay 22, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS A SEPARATION BETWEEN THE FEMALE CONNECTOR AND MAIN BODY OF THE MINICAP TRANSFER SET. THE BLUE PORTION OF THE TRANSFER SET WAS LOOSE AND ROTATING. THIS WAS IDENTIFIED DURING A TRANSFER SET CHANGE. THE TRANSFER SET WAS NOT USED AND WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25260369 ↗MalfunctionMay 20, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS A SEPARATION BETWEEN THE FEMALE CONNECTOR AND THE MAIN BODY OF A MINICAP TRANSFER SET; DESCRIBED AS ¿THE DARK BLUE PART GOT DISCONNECTED FROM THE BODY OF THE TRANSFER SET¿. THIS WAS IDENTIFIED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25231191 ↗MalfunctionMay 19, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS A SEPARATION BETWEEN THE FEMALE CONNECTOR AND THE MAIN BODY OF A MINICAP TRANSFER SET. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25217974 ↗MalfunctionMay 18, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS A SEPARATION BETWEEN THE FEMALE CONNECTOR AND MAIN BODY OF THE MINICAP TRANSFER SET. THIS OCCURRED WHEN DISCONNECTING FROM THE CYCLER AFTER PERITONEAL DIALYSIS THERAPY. THE PATIENT LINE WOULD NOT DISCONNECT INITIALLY. THE PATIENT NEEDED SOMETHING TO HELP THEM GRIP ONTO THE CONNECTIONS AND WAS EVENTUALLY ABLE TO DISCONNECT. THE TRANSFER SET WAS REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25203354 ↗MalfunctionMay 14, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS A SEPARATION BETWEEN THE FEMALE CONNECTOR AND THE MAIN BODY OF A MINICAP TRANSFER SET; DESCRIBED AS ¿THE NAVY BLUE THREADS ARE SEPARATING FROM THE LIGHT BLUE COMPONENT OF THE TRANSFER SET¿. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THE TRANSFER SET HAD BEEN IN USE FOR ONE WEEK AT THE TIME OF THE EVENT. AS A RESULT, THE TRANSFER SET WAS REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176287 ↗MalfunctionMay 13, 2026Free or Unrestricted Flow
IT WAS REPORTED THAT A MINICAP TRANSFER SET LEAKED WITH THE TWIST CLAMP CLOSED. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THE TRANSFER SET WAS REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25165083 ↗MalfunctionMay 13, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A MINICAP TRANSFER SET HAD IODINE LEAKAGE. THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25164775 ↗MalfunctionMay 12, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT A MINICAP TRANSFER HAD A SEPARATION ISSUE; DESCRIBED AS ¿THE DARK BLUE END PART GOT DISCONNECTED FROM THE BODY OF THE TRANSFER SET¿. THIS OCCURRED WHEN THE PATIENT WAS ABOUT TO DISCONNECT FROM THE PD CYCLER AFTER AN AUTOMATED PERITONEAL DIALYSIS (PD) THERAPY SESSION. AS A RESULT, THE TRANSFER SET WAS REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25153406 ↗MalfunctionMay 5, 2026Difficult to Open or Close
IT WAS REPORTED THAT THE TWIST CLAMP OF THE MINICAP TRANSFER SET COULD NOT BE CLOSED. THIS WAS IDENTIFIED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084122 ↗MalfunctionApr 28, 2026Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THERE WAS A SLIT IN THE PAPER SIDE OF THE PACKAGING OF THE MINICAP TRANSFER SET. THIS WAS IDENTIFIED BEFORE USE OF THE DEVICE FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019818 ↗MalfunctionApr 24, 2026Difficult to Open or Close
IT WAS REPORTED THAT THE TWIST CLAMP OF THE MINICAP TRANSFER SET COULD NOT BE CLOSED. THIS WAS IDENTIFIED DURING A TRANSFER SET REPLACEMENT. A NEW TRANSFER SET WAS USED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24985049 ↗MalfunctionApr 20, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A LEAK BETWEEN THE TITANIUM ADAPTER AND MINICAP TRANSFER SET. THIS OCCURRED DURING USE OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24925688 ↗
Other Baxter Healthcare Corporation products
- EXACTAMIX28 models · 6,739 reports · 13 recalls
- Novum IQ3 models · 1,540 reports · 7 recalls
- CLEARLINK/CONTINU-FLO/DUO-VENT3 models · 873 reports · 3 recalls
- CLEARLINK19 models · 710 reports · 4 recalls
- CLEARLINK/CONTINU-FLO8 models · 416 reports · 4 recalls
- CLEARLINK/DUO-VENT5 models · 282 reports · 1 recall
- REPEATER6 models · 189 reports
- DUO-VENT5 models · 31 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page