Brand Name: Atlan recallZ-0304-2025 · Draeger Medical, Inc. · initiated Oct 15, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

FDA root cause category: Process control

What the company is doing

On October 15, 2024 Urgent Medical Device Correction letters were sent to customers. Necessary Action: The corrective action is to replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly replacement. Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Monitor the patient s condition specifically their oxygenation status as a brief cessation of ventilation can potentially lead to e.g.; hypoxia, loss of lung recruitment, bradycardia and cardiac arrest. If preferred, the alarm priority of the "Ventilator error!!!" can be downgraded with "ALARM RESET" after switching to Man/Spont ventilation mode. Identification of affected medical devices: According to our records, you have received at least one Atlan anesthesia workstation manufactured by Draegerwerk AG & Co. KGaA that may be affected by this issue. Please refer to the attached list of affected serial numbers. Please ensure that all users of the Drger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. Please complete and return the attached Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Draeger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding

Recall details

Recall number
Z-0304-2025
Status
Open, Classified
Initiated by the company
Oct 15, 2024
Posted by the FDA
Nov 13, 2024
Recalling firm
Draeger Medical, Inc.
Quantity
1,492 units (139 units US, 1,353 OUS)
Product code
BSZ
FDA submissions
K230931
Distribution
US Nationwide distribution in the states of AL, AZ, CA, FL, GA, IA, IL, KY, MA, ME, ND, NE, NH, NJ, NY, OH, PA, TN, TX, WI, WY. International distribution: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Columbia, Croatia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Kenya, Latvia, Malaysia, Mauritius, Mexico, Morocco, Namibia, Nepal, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.

Product description

Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Component: Not applicable

Code information (models, lots, serial numbers)

Show the full code information (9k characters)

Model/Catalog Number: 8211500; UDI-DI Number: 04048675556176; Serial Numbers Distributed in the US: ASTB-0038; 0039; 0146; 0147; 0148; 0281; 0282; 0283; 0300; 0301; 0302; 0303; 0312; 0313; 0314; 0315; 0316 ASTC-0005; 0094; 0095; 0190; 0191 ASTD-0029; 0030; 0031; 0032; 0033; 0034; 0035; 0036; 0037; 0038; 0039; 0040; 0041; 0042; 0043; 0044; 0045; 0046 0057; 0136; 0137; 0138; 0139; 0140; 0141; 0142; 0143; 0148; 0149; 0150; 0151; 0152; 0153; 0155; 0156; 0157; 0158; 0170; 0171; 0186; 0235; 0236; 0237; 0257; 0258; 0259 ASTE-0002; 0003; 0004; 0005; 0006; 0007; 0008; 0029; 0030; 0031; 0032; 0080; 0134; 0135; 0136; 0152; 0153; 0154; 0155; 0180; 0181; 0209; 0210; 0211; 0212; 0213; 0214; 0215; 0216; 0217; 0218; 0254; 0255; 0318; 0319; 0320; 0321; 0322 ASTF-0007; 0008; 0009; 0010; 0145; 0146; 0147; 0148; 0149; 0150; 0177; 0178; 0379; 0380 ASTH-0083; 0084; 0107; 0160; 0161; 0184; 0185; 0207; 0208; 0233; 0241; 0242; 0254; 0370; 0392; 0432; 0464 Serial Numbers Distributed OUS ASTA-0279; 0293; 0302; 0303; 0304; 0306; 0334; 0335; 0336; 0344; 0347; 0348; 0360; 0369; 0373; 0375; 0376; 0377; 0379; 0380; 0381; 0386; 0389; 0390; 0392; 0393; 0399; 0412; 0422; 0433; 0435; 0437; 0438 ASTB-0009; 0011; 0012; 0015; 0016; 0017; 0019; 0024; 0025; 0026; 0027; 0035; 0036; 0042; 0043; 0044; 0045; 0047; 0050; 0056; 0058; 0059; 0060; 0062; 0067; 0069; 0070; 0072; 0073; 0085; 0086; 0087; 0088; 0089; 0090; 0091; 0092; 0096; 0097; 0098; 0099; 0100; 0101; 0102; 0104; 0105; 0106; 0107; 0108; 0109; 0110; 0111; 0112; 0113; 0114; 0115; 0116; 0117; 0118; 0119; 0120; 0121; 0123; 0124; 0125; 0126; 0127; 0128; 0129; 0130; 0131; 0132; 0133; 0135; 0136; 0137; 0138; 0139; 0140; 0141; 0142; 0143; 0144; 0145; 0147; 0149; 0152; 0153; 0154; 0155; 0156; 0157; 0158; 0159; 0163; 0164; 0166; 0167; 0168; 0169; 0170; 0171; 0172; 0173; 0174; 0183; 0184; 0185; 0186; 0187; 0188; 0189; 0190; 0191; 0192; 0193; 0194; 0196; 0197; 0198; 0199; 0200; 0201; 0202; 0203; 0204; 0205; 0207; 0208; 0209; 0210; 0212; 0213; 0221; 0222; 0223; 0224; 0225; 0226; 0227; 0228; 0229; 0230; 0231; 0238; 0239; 0240; 0241; 0242; 0243; 0244; 0245; 0246; 0247; 0251; 0252; 0253; 0254; 0255; 0256; 0257; 0258; 0259; 0260; 0261; 0267; 0268; 0269; 0270; 0271; 0272; 0273; 0274; 0275; 0284; 0285; 0292; 0293; 0294; 0295; 0296; 0297; 0298; 0299; 0301; 0310; 0311; ASTC-0009; 0010; 0011; 0012; 0013; 0014; 0015; 0016; 0017; 0018; 0019; 0020; 0021; 0022; 0023; 0024; 0027; 0028; 0029; 0030; 0031; 0032; 0033; 0034; 0035; 0043; 0044; 0045; 0051; 0052; 0053; 0054; 0055; 0056; 0057; 0058; 0059; 0060; 0061; 0065; 0066; 0067; 0071; 0072; 0073; 0074; 0075; 0076; 0084; 0085; 0086; 0087; 0088; 0096; 0097; 0098; 0110; 0111; 0112; 0113; 0114; 0115; 0116; 0117; 0120; 0121; 0122; 0128; 0129; 0130; 0131; 0132; 0134; 0135; 0141; 0142; 0143; 0144; 0145; 0146; 0147; 0148; 0149; 0155; 0156; 0157; 0158; 0159; 0160; 0161; 0162; 0164; 0165; 0166; 0167; 0168; 0170; 0171; 0172; 0188; 0189; 0193; 0197; 0198; 0199; 0204; 0205; 0206; 0207; 0208; 0209; 0210; 0211; 0212; 0213; 0214; 0215; 0216; 0217; 0231; 0232; 0233; 0234; 0235; 0236; 0237; 0238; 0239; 0240; 0241; 0246; 0247; 0248; 0249; 0250; 0251; 0252; 0253; 0254; 0255; 0256; 0257; 0258; 0259; 0260; 0261; 0262; 0263; 0264; 0265; 0266; 0267; 0268; 0269; 0270; 0271; 0272; 0273; 0274; 0275; 0276; 0277 ASTD-0001; 0002; 0003; 0004; 0005; 0006; 0007; 0017; 0018; 0019; 0020; 0021; 0022; 0023; 0024; 0025; 0047; 0048; 0049; 0053; 0054; 0055; 0056; 0058; 0059; 0060; 0067; 0068; 0069; 0070; 0071; 0072; 0073; 0074; 0075; 0076; 0077; 0078; 0079; 0104; 0105; 0106; 0107; 0108; 0109; 0110; 0111; 0112; 0113; 0114; 0115; 0116; 0117; 0118; 0119; 0125; 0126; 0127; 0128; 0177; 0178; 0179; 0180; 0184; 0185; 0196; 0197; 0198; 0199; 0200; 0201; 0202; 0206; 0207; 0208; 0209; 0210; 0211; 0212; 0213; 0214; 0215; 0216; 0217; 0218; 0219; 0220; 0221; 0222; 0223; 0224; 0234; 0247; 0248; 0249; 0250; 0251; 0252; 0253; 0254; 0255; 0256; 0260; 0261; 0262; 0263; 0264; 0269; 0270; 0271; 0272; 0273; 0274; 0275; 0276; 0277; 0278; 0279; 0280; 0281; 0282; 0283; 0284 ASTE-0009; 0010; 0011; 0018; 0019; 0020; 0021; 0022; 0023; 0024; 0025; 0026; 0027; 0028; 0033; 0036; 0037; 0038; 0039; 0040; 0045; 0046; 0047; 0048; 0049; 0050; 0051; 0052; 0053; 0054; 0055; 0061; 0066; 0067; 0068; 0069; 0070; 0071; 0072; 0073; 0074; 0075; 0076; 0081; 0082; 0088; 0089; 0090; 0091; 0092; 0093; 0094; 0095; 0096; 0097; 0098; 0099; 0100; 0101; 0102; 0103; 0104; 0105; 0106; 0107; 0109; 0110; 0111; 0112; 0113; 0114; 0115; 0116; 0117; 0118; 0119; 0120; 0121; 0132; 0133; 0137; 0138; 0139; 0140; 0141; 0142; 0143; 0144; 0145; 0146; 0147; 0148; 0149; 0150; 0151; 0156; 0157; 0158; 0159; 0160; 0161; 0162; 0167; 0168; 0169; 0170; 0182; 0183; 0184; 0185; 0186; 0188; 0189; 0190; 0191; 0192; 0193; 0194; 0204; 0205; 0206; 0207; 0208; 0227; 0228; 0229; 0230; 0231; 0232; 0233; 0234; 0235; 0236; 0237; 0238; 0239; 0240; 0241; 0242; 0251; 0252; 0253; 0256; 0257; 0258; 0259; 0260; 0261; 0262; 0263; 0264; 0265; 0266; 0267; 0268; 0269; 0270; 0271; 0272; 0273; 0274; 0275; 0276; 0277; 0283; 0284; 0285; 0286; 0287; 0288; 0289; 0290; 0291; 0292; 0293; 0294; 0295; 0296; 0297; 0298; 0307; 0308; 0309; 0310; 0311; 0312; 0313; 0314; 0315; 0316; 0317; 0323; 0324; 0330; 0331; 0332; 0333; 0334; 0335; 0336; 0337; 0338; 0339; 0340; 0341; 0342; 0343; 0344; 0345; 0346; 0347; 0363; 0364; 0365; 0366; 0367; 0368; 0369; 0376 ASTF-0011; 0012; 0013; 0020; 0021; 0022; 0023; 0024; 0025; 0026; 0038; 0039; 0040; 0041; 0042; 0043; 0044; 0045; 0046; 0047; 0056; 0057; 0058; 0059; 0060; 0064; 0065; 0066; 0067; 0068; 0069; 0083; 0084; 0085; 0087; 0088; 0089; 0090; 0091; 0092; 0094; 0095; 0096; 0097; 0098; 0099; 0100; 0101; 0102; 0103; 0104; 0105; 0106; 0107; 0108; 0109; 0114; 0115; 0116; 0117; 0120; 0126; 0127; 0128; 0129; 0130; 0131; 0132; 0133; 0134; 0142; 0143; 0144; 0152; 0153; 0154; 0156; 0157; 0158; 0159; 0160; 0161; 0163; 0164; 0165; 0171; 0172; 0173; 0174; 0175; 0176; 0179; 0180; 0181; 0182; 0183; 0184; 0185; 0186; 0187; 0188; 0189; 0190; 0191; 0192; 0193; 0199; 0200; 0201; 0202; 0203; 0204; 0205; 0206; 0207; 0208; 0209; 0210; 0211; 0212; 0213; 0214; 0215; 0216; 0217; 0218; 0219; 0223; 0224; 0225; 0226; 0227; 0228; 0229; 0230; 0231; 0232; 0233; 0234; 0235; 0236; 0237; 0238; 0239; 0240; 0241; 0242; 0243; 0244; 0245; 0246; 0247; 0248; 0249; 0250; 0251; 0252; 0253; 0254; 0255; 0256; 0257; 0258; 0259; 0260; 0261; 0262; 0263; 0264; 0265; 0266; 0267; 0268; 0269; 0270; 0271; 0272; 0273; 0274; 0275; 0276; 0277; 0278; 0279; 0280; 0281; 0282; 0283; 0284; 0285; 0286; 0287; 0288; 0289; 0290; 0291; 0292; 0293; 0294; 0295; 0296; 0297; 0298; 0299; 0300; 0302; 0303; 0304; 0305; 0306; 0307; 0308; 0309; 0310; 0311; 0312; 0313; 0314; 0315; 0316; 0317; 0318; 0319; 0320; 0321; 0322; 0323; 0324; 0325; 0326; 0327; 0328; 0329; 0330; 0331; 0332; 0333; 0334; 0335; 0336; 0337; 0338; 0339; 0340; 0341; 0342; 0343; 0344; 0345; 0346; 0347; 0348; 0349; 0350; 0351; 0352; 0353; 0354; 0355; 0356; 0357; 0358; 0359; 0360; 0361; 0362; 0363; 0364; 0365; 0366; 0367; 0368; 0369; 0370; 0371; 0372; 0373; 0374; 0375; 0376; 0377; 0378; 0381; 0382; 0383; 0384; 0385; 0386; 0387; 0388; 0389; 0390; 0391; 0392; 0393; 0394; 0395; 0396; 0397; 0398; 0399; 0400; 0401; 0402; 0403; 0404; 0405; 0407; 0408; 0412; 0413; 0414; 0415; 0416; 0417; 0418; 0419; 0420; 0421; 0422; 0423; 0424; 0425; 0426; 0427; 0428; 0429; 0430; 0431; 0432; 0433; 0439; 0440; 0441; 0442; 0443; 0444; 0445; 0446; 0447; 0448; 0449; 0450; 0451; 0452; 0453; 0454; 0455; 0456; 0457; 0458; 0459; 0460; 0461; 0462; 0463; 0464; 0465; 0466; 0467; 0468; 0469; 0470 ASTH-0004; 0013; 0014; 0015; 0016; 0017; 0018; 0019; 0020; 0021; 0022; 0028; 0029; 0030; 0031; 0032; 0033; 0034; 0035; 0036; 0037; 0038; 0048; 0049; 0050; 0051; 0052; 0053; 0054; 0055; 0056; 0057; 0058; 0059; 0060; 0061; 0062; 0063; 0064; 0065; 0066; 0067; 0068; 0069; 0070; 0071; 0072; 0073; 0074; 0077; 0078; 0079; 0080; 0085; 0086; 0087; 0089; 0090; 0091; 0100; 0101; 0102; 0103; 0104; 0105; 0106; 0108; 0109; 0110; 0111; 0112; 0113; 0114; 0127; 0128; 0129; 0130; 0131; 0132; 0133; 0134; 0135; 0136; 0137; 0138; 0139; 0140; 0141; 0154; 0155; 0156; 0157; 0158; 0159; 0162; 0163; 0164; 0165; 0166; 0167; 0168; 0177; 0178; 0179; 0180; 0181; 0182; 0183; 0199; 0200; 0201; 0202; 0203; 0205; 0206; 0210; 0211; 0212; 0213; 0214; 0215; 0216; 0231; 0232; 0234; 0235; 0236; 0237; 0238; 0239; 0240; 0247; 0248; 0249; 0250; 0251; 0252; 0253; 0262; 0263; 0264; 0265; 0268; 0283; 0284; 0285; 0286; 0287; 0288; 0289; 0290; 0291; 0292; 0300; 0301; 0321; 0322; 0323; 0340; 0341; 0342; 0343; 0344; 0345; 0346; 0347; 0348; 0349; 0350; 0351; 0352; 0353; 0354; 0366; 0367; 0368; 0369; 0371; 0372; 0373; 0374; 0375; 0376; 0377; 0378; 0379; 0380; 0381; 0391; 0397; 0398; 0399; 0400; 0402; 0411; 0433; 0438; 0439; 0454; 0455; 0456; 0457; 0458; 0459; 0460; 0461; 0462; 0463; 0471; 0472; 0473; 0474; 0475; 0482; 0483 ASTJ-0026; 0027; 0028; 0029; 0030; 0073; 0074; 0075; 0076; 0103; 0162; 0197; 0198; 0199; 0200; 0208; 0256 ASTK-0003; 0004; 0005; 0006; 0007; 0008; 0009

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K230931 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95561).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page