Atlan A350 XL 8621600Recalls & reported problems
Drägerwerk AG & Co. KGaA · Available with either electronically or mechanically controlled gas mixer variant · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04048675556183), received between Sep 20, 2023 and Mar 30, 2026 — classified as 1 death, 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (11 reports).
4 FDA recall records name this device's identifier, 4 still open.
5 other recalls cover devices cleared under the same FDA submission (K230931); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 7, 2026Class IOpen, Classified
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Z-2208-2026 · Root cause: Process change control · Recall details · FDA record ↗
- Sep 9, 2025Class IOpen, Classified
The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Z-2617-2025 · Root cause: Process control · Recall details · FDA record ↗
- Oct 15, 2024Class IOpen, Classified
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Z-0305-2025 · Root cause: Process control · Recall details · FDA record ↗
- Nov 17, 2023Class IIOpen, Classified
Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.
Z-0971-2024 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Deliver11
- Gas Output Problem7
- Protective Measures Problem3
- Use of Device Problem3
- Decrease in Pressure2
- Failure to Analyze Signal2
- Gas/Air Leak2
- Inaccurate Delivery2
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Bradycardia1
- Hypoxia1
- Respiratory Failure1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
DeathMar 30, 2026Decrease in Pressure; Inaccurate Delivery; Gas/Air Leak
IT WAS REPORTED THAT THE PATIENT EXPIRED DURING SURGERY. THERE IS NO ALLEGED DEVICE MALFUNCTION OR ISSUES WITH THE MACHINE'S PERFORMANCE REPORTED.
Patient problems: Respiratory Failure
MAUDE report 24726713 ↗MalfunctionMar 24, 2026Unexpected Shutdown
IT WAS REPORTED THAT THE DEVICE SHUT DOWN DURING A CASE. THERE WAS NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24679622 ↗MalfunctionMar 17, 2026Gas Output Problem; Failure to Deliver; Intermittent Communication Failure
IT WAS REPORTED THAT THE UNIT FAILED DURING CASE AND THAT THE SCREEN WENT BLANK. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616176 ↗MalfunctionNov 26, 2025Gas Output Problem; Failure to Deliver; Protective Measures Problem; Gas Output Problem; Failure to Deliver; Protective Measures Problem
THERE WAS A VENT FAIL DURING USE REPORTED; IT WAS CONFIRMED THAT THE EVENT DID NOT LEAD TO HEALTH CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23655512 ↗InjuryAug 25, 2025Decrease in Pressure; Inaccurate Delivery; Gas/Air Leak; Decrease in Pressure; Inaccurate Delivery; Gas/Air Leak
IT WAS REPORTED THAT THE PATIENT WAS NOT RECEIVING ANY FLOW. THE CASE IN QUESTION WAS REPORTEDLY STARTED WITH MASK VENTILATION AND DURING THIS TIME, THERE WAS NO FLOW WAVEFORM SEEN ON THE DEVICE. THE PATIENT¿S SATS WERE DROPPING AND THEY STARTED TO INTUBATE. ONCE INTUBATED, THEY ATTEMPTED TO VENTILATE BUT AGAIN NO FLOW WAS DETECTED. DURING THIS TIME, THEY STARTED COMPRESSIONS BECAUSE THE PATIENT WAS BECOMING HYPOXIC WHICH CAUSED THE HEART RATE TO DROP. THE DEVICE WAS REPORTEDLY CHECKED AND NO INDICATIONS FOR DEVICE MALFUNCTION WERE FOUND WHICH WOULD EXPLAIN WHY NO FLOW WAS DELIVERED. AS FURTHER REPORTED, THE ONLY ALARM WAS ¿LOW FIO2¿. THEN, THE PHYSICIAN NOTICED THE MISSING CONNECTION OF THE PATIENT AT THE Y-PIECE. ONCE THEY CONNECTED THE CIRCUIT, THE PROBLEM WAS SOLVED. THERE WERE NO FURTHER CONSEQUENCES FOR THE PATIENT REPORTED. BASED ON AVAILABLE INFORMATION, MOST LIKELY AN EXTERNAL LEAKAGE WAS THE ROOT CAUSE. IT CANNOT BE EXCLUDED THAT A USE ERROR CAUSED OR CONTRIBUTED TO THE REPORTED SYMPTOM.
Patient problems: Bradycardia; Hypoxia
MAUDE report 22882840 ↗MalfunctionJul 31, 2025Use of Device Problem; Failure to Deliver; No Apparent Adverse Event; Use of Device Problem; Failure to Deliver; No Apparent Adverse Event
IT WAS REPORTED THAT THE DEVICE HAD DISCREPANCIES ON O2 READINGS. LOG ANALYSIS REVEALED THAT ALSO A VENTILATOR FAILURE OCCURRED ON THE REPORTED DATE/TIME OF EVENT. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22673712 ↗MalfunctionJul 30, 2025Gas Output Problem; Failure to Analyze Signal; Failure to Deliver; Gas Output Problem; Failure to Analyze Signal; Failure to Deliver
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE A VENTILATOR FAILURE OCCURRED. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22661598 ↗MalfunctionJul 7, 2025Gas Output Problem; Failure to Deliver; Protective Measures Problem; Gas Output Problem; Failure to Deliver; Protective Measures Problem
IT WAS REPORTED THAT THE DEVICE DISPLAYED THE ERROR ¿VENTILATOR FAILURE, PLEASE VENTILATE MANUALLY¿ DURING PC VENTILATION. MANUAL VENTILATION WAS POSSIBLE WITHOUT ANY PROBLEMS. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22413069 ↗MalfunctionMay 28, 2025Gas Output Problem; Failure to Deliver; Protective Measures Problem; Gas Output Problem; Failure to Deliver; Protective Measures Problem
IT WAS REPORTED THAT THE DEVICE POSTED AN ALARM FOR A VENTILATION FAILURE. MANUAL VENTILATION WAS POSSIBLE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22094533 ↗MalfunctionMar 27, 2025Gas Output Problem; Incorrect Interpretation of Signal; Failure to Deliver; Gas Output Problem; Incorrect Interpretation of Signal; Failure to Deliver
IT WAS REPORTED THAT, ON THE (B)(6) 2025, AROUND 16.40 PM, WHILE THE PATIENT WAS ON THE VENTILATOR, THE DEVICE ALARMED FOR A VENTILATOR FAILURE. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21706770 ↗MalfunctionMar 7, 2025Gas Output Problem; Failure to Analyze Signal; Failure to Deliver; Gas Output Problem; Failure to Analyze Signal; Failure to Deliver
IT WAS REPORTED THAT THE DEVICE STOPPED VENTILATING THE PATIENT DURING A SURGICAL INTERVENTION IN TWO OCCASIONS. ONE ON (B)(6) 2024 AND ONE ON (B)(6) 2024. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21545065 ↗MalfunctionFeb 19, 2025Incorrect Interpretation of Signal; Use of Device Problem; Failure to Deliver; Incorrect Interpretation of Signal; Use of Device Problem; Failure to Deliver
IT WAS REPORTED THAT A VENTILATOR FAILURE OCCURRED DURING USE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21415089 ↗MalfunctionFeb 5, 2025Use of Device Problem; Failure to Deliver; Use of Device Problem; Failure to Deliver
VENT FAIL MESSAGE DURING USE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21310013 ↗MalfunctionSep 20, 2023Failure to Deliver; Device Handling Problem; Failure to Deliver; Device Handling Problem
IT WAS REPORTED THAT THE MACHINE FAILED IN THE MIDDLE OF A CASE WITH A "VENTILATOR FAILURE" MESSAGE. THE PROCEDURE WAS FINISHED WITH THE PATIENT BREATHING SPONTANEOUSLY. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17779817 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K230931. One submission often covers many models, so check the product and lot details in each record.
- Z-2207-2026 · May 7, 2026 · Open, ClassifiedAtlan A350. Model Number: 8211500. anesthesia workstation
- Z-2616-2025 · Sep 9, 2025 · Open, ClassifiedBrand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Product Description: The Atlan is an anesthesia workstation intended for u
- Z-0667-2025 · Oct 15, 2024 · Open, ClassifiedBrand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonat
- Z-0668-2025 · Oct 15, 2024 · Open, ClassifiedBrand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neon
- Z-0304-2025 · Oct 15, 2024 · Open, ClassifiedBrand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Component: Not applicable
Other Gas-Machine, Anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,109 reports
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,204 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,015 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,540 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page