NovoHip Total Hip System recallZ-0483-2025 · Novosource, LLC · initiated Sep 19, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Novosource, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mislabeled: Packaged in a box labeled as 32 long but containing a 28 long head by physical marking, may result in joint instability if implanted
FDA root cause category: Mixed-up of materials/components
What the company is doing
NovoSource LLC issued Urgent Medical Device Recall letter on 9/13/24 via email. Letter states reason for recall, health risk and action to take: "Immediate Action Discontinue Use: All customers and healthcare providers should immediately discontinue the use of the affected product batches. Surgical teams should not use any Biolox Delta Ceramic Head implants from the mislabeling-affected batches in any hip replacement surgeries. "Return or Segregation of Affected Product: Affected implants should be removed from active inventory and returned to NovoSource. Providers should follow instructions for proper packaging and shipping of the returned products. Any products not yet returned should be clearly labeled as "Do Not Use" and quarantined from inventory until shipment back to NovoSource. Instructions on how to return products will be provided, including the appropriate return address and contact person for support during this process. " For Unused Products: Customers should ensure that all unused products from the affected batches are returned to NovoSource. This will prevent any risk of improper head-liner fitting due to the mislabeling. "For Implanted Products: If the mislabeled product has already been implanted, surgeons should be prepared to monitor the patient for symptoms consistent with the mismatch. A mismatch of a 32 long head implanted into a 28 liner would likely have instability issues. A 28 long head implanted into a 32 liner will likely act very similarly in stability and alignment, while wear should not be catastrophic. The surgeon should monitor the patient to determine if symptoms appear that would result in the need to revise. Actions to Be Taken Pending Corrective or Removal Action: "Stop Usage: Healthcare facilities should stop any further use of the affected product batches pending completion of the recall process. NovoSource will provide updates on the investigation and resolution. "Notification to Surgical Teams: IF IMPLANTED, it is essential
Recall details
- Recall number
- Z-0483-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 19, 2024
- Posted by the FDA
- Nov 18, 2024
- Recalling firm
- Novosource, LLC
- Quantity
- 1 unit
- Product code
- LPH
- FDA submissions
- K132158
- Distribution
- Florida, West Virginia
Product description
NovoHip Biolox Ball Head 32mmL Ref: 1345-2803 NovoHip Total Hip System - NovoHip Total Hip System is a non-cemented hip prosthesis that consists of a 4-part total hip replacement system including femoral stem, femoral head, acetabular poly liner, and acetabular metal (or shell) components.
Code information (models, lots, serial numbers)
UDI-DI 00811382037849; Lot Number CER0023
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NovoHip Total Hip System 1345-3203Novosource, LLC · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented1 recall1 report
Product lines
Other devices under the same FDA submission
1 device cleared under K132158 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NovoHip Total Hip System 1345-2803Novosource, LLC · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page