Skip to content

MICRO-X Rover Mobile X-ray System, # MXU-RV35 recallZ-0664-2025 · Micro-X Ltd. · initiated May 31, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Micro-X Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

FDA root cause category: Software design

What the company is doing

The firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.

Recall details

Recall number
Z-0664-2025
Status
Open, Classified
Initiated by the company
May 31, 2023
Posted by the FDA
Dec 12, 2024
Recalling firm
Micro-X Ltd.
Quantity
5
Product code
IZL
FDA submissions
K211423
Distribution
Worldwide and US Nationwide Distribution: NJ, PR, CA

Product description

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Code information (models, lots, serial numbers)

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K211423 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page