Skip to content

Homocysteine Enzymatic Assay recallZ-0489-2025 · Roche Diagnostics Operations, Inc. · initiated Oct 17, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)

FDA root cause category: Under Investigation by firm

What the company is doing

Roche issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/17/2024 via UPS Ground. The notice explained the issue, clinical significance, and requested the following actions be taken: "" HCYS kits lot number 747029 expired on September 30, 2024. Ensure any inventory you have of this lot has been discarded per your local guidelines. " Follow your laboratory s standard operating procedures to determine specific clinical implications for your patients and whether retrospective review of patient results is required. " Consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site." For questions: contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336

Recall details

Recall number
Z-0489-2025
Status
Open, Classified
Initiated by the company
Oct 17, 2024
Posted by the FDA
Nov 19, 2024
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
458 units
Product code
LPS
FDA submissions
K113793
Distribution
US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,

Product description

cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine

Code information (models, lots, serial numbers)

UDI/DI 04015630929979, Lot Number 747029, exp. 30-SEPT-2024

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K113793 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.