Skip to content

MICRO-X Rover Mobile X-ray System, # MXU-RV35 recallZ-0665-2025 · Micro-X Ltd. · initiated Nov 28, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Micro-X Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.

FDA root cause category: Device Design

What the company is doing

On 11/28/23 Modification 17 was sent to customers telling them that Control Boards need to be replaced with Control Board PCBA, 12425-04 that will eliminate the issue. For technical assistance contact: [email protected], 1-206-249-8764 Firm asked distributors to do the following: On next visit please confirm that the bulletins were completed and return the documentation.

Recall details

Recall number
Z-0665-2025
Status
Open, Classified
Initiated by the company
Nov 28, 2023
Posted by the FDA
Dec 12, 2024
Recalling firm
Micro-X Ltd.
Quantity
18
Product code
IZL
FDA submissions
K211423
Distribution
US Nationwide Distribution: NJ, CA, PR

Product description

MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71

Code information (models, lots, serial numbers)

MXU-RV35 UDI-DI: 9357123000037 and MXU-RV71 UDI-DI: 935712300005 using Control Board PCBA, 12425-03 (or earlier). Serial Numbers: MSN00334, MSN00348, MSN00349, MSN00351, MSN00352, MSN00361, MSN00364, MSN00365,MSN00382, MSN00383, MSN00384, MSN00385, MSN00386, MSN00387, MSN00388, MSN00390, MSN00391, MSN00395

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K211423 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page