MICRO-X Rover Mobile X-ray System, # MXU-RV35 recallZ-0580-2025 · Micro-X Ltd. · initiated Oct 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Micro-X Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.

FDA root cause category: Environmental control

What the company is doing

On 10/22/24, Service Bulletin SB007-23-1.0 was emailed to customers restating receiving, storage, and operation conditions and informing customers that extreme humidity environments can adversely affect the equipment. For further information contact your local service representative: phone: 1-206-249-8764 Email: [email protected]

Recall details

Recall number
Z-0580-2025
Status
Open, Classified
Initiated by the company
Oct 22, 2024
Posted by the FDA
Nov 27, 2024
Recalling firm
Micro-X Ltd.
Quantity
6
Product code
IZL
FDA submissions
K211423
Distribution
US: TX, PR, MD, CA

Product description

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Code information (models, lots, serial numbers)

UDI-DI: 9357123000037, Serial Numbers: 348, 349, 361, 351, 365, 391

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K211423 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.