BD Phoenix M50 Automated Microbiology System recallZ-0592-2025 · Becton Dickinson & Co. · initiated Oct 31, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
FDA root cause category: Software design
What the company is doing
On October 31, 2024 URGENT" MEDICAL DEVICE PRODUCT CORRECTION letters were sent to consignees. IT, Safety and Security Actions to be Taken: BD encourages customers to follow best practices for maintaining strong security measures to protect hospital networks and medical devices including: "Ensure access to potentially vulnerable devices is limited to authorized personnel "Inform authorized users of issue, and ensure all relevant passwords are tightly controlled "Monitor and log network traffic attempting to reach medical device managementenvironments for suspicious activity "Where possible, isolate affected devices in a secure VLAN or behind firewalls with restrictedaccess that only permits communication with trusted hosts in other networks when needed "Impacted devices do not require use of RDP ports and these should be disabled or blockedif enabled "Ensure permissions on file shares are appropriately established and enforced, and monitorand log access for evidence of suspicious activity "Disconnect devices from the network if connectivity is not necessary Next Steps to be Taken by Customer: 1.Ensure the contents of this notification are read and understood. 2.Share and post this customer letter within your facility network including IT and safety andsecurity department. Forward to any customers to whom you may have distributed theproduct(s) to ensure awareness. 3.Complete the attached Customer Response Form and return it to the BD contact noted on theForm so that BD may acknowledge your receipt of this notification per FDA requirements. 4.Report any adverse health consequences experienced with the use of these products to BD.Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 If you require further assistance, please contact: Technical Service +1 888 451 5514 Technical Questi
Recall details
- Recall number
- Z-0592-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 31, 2024
- Posted by the FDA
- Dec 3, 2024
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 4,411 units (567 US, 3,844 OUS)
- Product code
- LON
- FDA submissions
- K131331, K040099, K020321
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.
Product description
BD Phoenix M50 Automated Microbiology System Version or Model: 443624 Catalog Number: 443624
Code information (models, lots, serial numbers)
Show the full code information (35k characters)
Catalog No: 443624; UDI-DI:00382904436247; Serial Numbers: PF4176, PF4182, PF2302, PF4198, PF4149, PF4189, PF4202, PF2699, PF4200, PF4207, PF4209, PF2704, PF2640, PF2769, PF3383, PF2742, PF5292, PF4215, PF4204, PF0571, PF4188, PF4216, PF1642, PF4231, PF0929, PF4240, PF3443, PF0934, PF4232, PF4212, PF0566, PF3126, PF5408, PF5436, PF1716, PF5384, PF5396, PF5398, PF5404, PF3345, PF5443, PF3203, PF3128, PF0598, PF3447, PF0601, PF5489, PF5495, PF5022, PF5063, PF3362, PF3340, PF5459, PF5492, PF5462, PF5499, PF0703, PF5466, PF0880, PF0705, PF5525, PF5469, PF5471, PF3397, PF4309, PF4397, PF5547, PF0632, PF0638, PF4443, PF0637, PF5200, PF5549, PF5202, PF5550, PF0643, PF5049, PF0639, PF3204, PF5207, PF5094, PF5209, PF5117, PF1352, PF5217, PF5148, PF5231, PF5195, PF5097, PF5162, PF5109, PF5169, PF5179, PF5183, PF2063, PF5197, PF5203, PF5168, PF5323, PF5329, PF0687, PF5352, PF5353, PF5418, PF5350, PF2796, PF5359, PF5227, PF3169, PF5351, PF5365, PF5218, PF5361, PF5389, PF5265, PF5266, PF5328, PF5337, PF2274, PF0707, PF6126, PF0718, PF5356, PF5403, PF6167, PF5027, PF5349, PF5371, PF6174, PF4049, PF4071, PF4444, PF4064, PF4292, PF4305, PF4306, PF4315, PF4317, PF5042, PF4285, PF4287, PF4304, PF4373, PF4376, PF4407, PF4297, PF5043, PF4294, PF5072, PF5082, PF4251, PF4254, PF4393, PF4394, PF1032, PF5140, PF4252, PF5143, PF4253, PF5144, PF4335, PF4379, PF4390, PF4419, PF4308, PF5414, PF4295, PF5431, PF4337, PF4100, PF4415, PF4420, PF1054, PF2590, PF4346, PF5633, PF4278, PF4439, PF5874, PF5877, PF4338, PF4436, PF4430, PF5887, PF5011, PF5024, PF5048, PF5790, PF1478, PF1483, PF5047, PF4406, PF5134, PF1932, PF5021, PF5947, PF5029, PF5050, PF5130, PF5142, PF5145, PF5156, PF4412, PF1481, PF1091, PF5125, PF1096, PF5129, PF1099, PF5135, PF5088, PF1154, PF5108, PF1305, PF5164, PF6098, PF5111, PF1298, PF1373, PF1387, PF5153, PF6304, PF1477, PF6305, PF1513, PF6307, PF1743, PF5058, PF5064, PF5087, PF5078, PF0575, PF0534, PF0828, PF0928, PF5013, PF0672, PF0673, PF0675, PF1969, PF0429, PF1497, PF0533, PF2269, PF0633, PF5051, PF0841, PF1011, PF0719, PF0721, PF0737, PF0758, PF5107, PF5275, PF5493, PF5518, PF5163, PF0306, PF5472, PF0409, PF5494, PF5508, PF5509, PF5517, PF5521, PF0463, PF5522, PF2316, PF5527, PF2299, PF1124, PF0468, PF2407, PF0515, PF0628, PF0630, PF5582, PF0626, PF0879, PF0677, PF1112, PF5257, PF0541, PF0794, PF1122, PF1119, PF5548, PF1001, PF5555, PF0984, PF5561, PF5571, PF5572, PF5575, PF5589, PF5591, PF2200, PF5594, PF2202, PF5598, PF2096, PF5599, PF0004, PF1569, PF1045, PF1662, PF1405, PF1641, PF1412, PF1414, PF2003, PF2132, PF2223, PF2324, PF1528, PF1147, PF2369, PF2653, PF2561, PF5608, PF5611, PF2697, PF2268, PF1153, PF2546, PF0197, PF0199, PF0268, PF5733, PF5270, PF4099, PF4172, PF5613, PF0493, PF0494, PF0496, PF5957, PF1196, PF5966, PF1189, PF1195, PF1188, PF1192, PF0082, PF1202, PF0085, PF1201, PF1207, PF1206, PF1214, PF1191, PF1229, PF1230, PF1231, PF0614, PF0275, PF0190, PF0273, PF1257, PF1262, PF1063, PF1062, PF2379, PF0792, PF1208, PF0938, PF0942, PF1376, PF1384, PF1386, PF1390, PF0097, PF0094, PF0096, PF1408, PF0165, PF1285, PF0214, PF2360, PF2068, PF0307, PF2393, PF4056, PF2406, PF4017, PF3368, PF3197, PF3059, PF1354, PF1410, PF2564, PF1425, PF5173, PF5254, PF3412, PF4238, PF1430, PF1432, PF4384, PF1434, PF4223, PF4222, PF1417, PF0612, PF1422, PF1396, PF5832, PF0065, PF5833, PF0071, PF5862, PF5869, PF5822, PF0264, PF5825, PF3147, PF1507, PF5829, PF1423, PF5821, PF1427, PF5818, PF5830, PF5860, PF5861, PF1452, PF1437, PF0232, PF5693, PF1436, PF5682, PF2480, PF1508, PF1517, PF0751, PF1566, PF1720, PF6189, PF6171, PF6243, PF6254, PF6175, PF5240, PF2426, PF1582, PF2306, PF0319, PF0322, PF4325, PF1626, PF5480, PF1627, PF1752, PF1748, PF0917, PF1527, PF1531, PF0318, PF1816, PF0892, PF1537, PF1309, PF1539, PF0205, PF1371, PF1450, PF1451, PF1791, PF2236, PF2130, PF0013, PF2374, PF0023, PF0025, PF0028, PF0038, PF1591, PF1951, PF0034, PF1607, PF2752, PF1603, PF2760, PF0049, PF0054, PF4195, PF1630, PF1631, PF2975, PF3253, PF4068, PF5302, PF5320, PF0105, PF0102, PF1634, PF1637, PF1644, PF1648, PF1654, PF0600, PF1651, PF2330, PF2355, PF1669, PF2357, PF1673, PF1676, PF4367, PF2521, PF2836, PF2840, PF0144, PF5800, PF5791, PF5855, PF1677, PF5560, PF0150, PF0156, PF1686, PF1692, PF6086, PF6102, PF2839, PF1703, PF6112, PF1700, PF6124, PF6375, PF6125, PF6377, PF1116, PF1727, PF3172, PF1744, PF0290, PF0250, PF0181, PF4076, PF0233, PF0298, PF0296, PF0297, PF0291, PF1756, PF1758, PF0325, PF4326, PF1759, PF4371, PF0342, PF1772, PF0327, PF5005, PF0332, PF0234, PF0358, PF0349, PF0352, PF0086, PF0513, PF0615, PF0015, PF1820, PF1823, PF0570, PF0108, PF0572, PF0716, PF0712, PF0739, PF0717, PF0262, PF0823, PF4063, PF0263, PF0762, PF1830, PF1140, PF1832, PF0328, PF1824, PF0982, PF1827, PF0338, PF1169, PF1006, PF0532, PF1536, PF1554, PF0497, PF1277, PF1870, PF1965, PF1967, PF3050, PF2116, PF2083, PF0564, PF2503, PF0500, PF2502, PF2501, PF2213, PF2588, PF2087, PF2662, PF2782, PF2732, PF2992, PF2953, PF3010, PF1876, PF2206, PF1890, PF1247, PF1891, PF0839, PF4171, PF4184, PF3365, PF5243, PF2084, PF5189, PF5605, PF6333, PF0586, PF2477, PF4058, PF4055, PF2580, PF4083, PF1962, PF0884, PF1002, PF1003, PF0427, PF3225, PF3235, PF3341, PF3389, PF3411, PF1524, PF1995, PF2004, PF1583, PF2201, PF1624, PF1576, PF4429, PF1593, PF1279, PF1818, PF1834, PF2023, PF2025, PF5286, PF5069, PF2358, PF2422, PF2423, PF2036, PF5276, PF5290, PF5343, PF2219, PF2530, PF2651, PF2055, PF4400, PF2513, PF2054, PF2056, PF2965, PF2960, PF3112, PF3111, PF2067, PF5445, PF5648, PF4225, PF5958, PF6008, PF6054, PF6327, PF6329, PF1149, PF2088, PF2090, PF0041, PF2135, PF1990, PF2126, PF0053, PF4226, PF0017, PF0201, PF0415, PF0253, PF0426, PF5256, PF5261, PF0526, PF0972, PF0989, PF0217, PF0213, PF0593, PF0676, PF0971, PF2301, PF1084, PF1218, PF1616, PF1086, PF1113, PF1222, PF1379, PF1482, PF1485, PF0072, PF3145, PF1615, PF0073, PF2701, PF1707, PF1710, PF1740, PF2541, PF1859, PF1927, PF2108, PF3317, PF0110, PF3322, PF0109, PF4075, PF4134, PF0837, PF0908, PF1415, PF5229, PF0069, PF5255, PF0635, PF5333, PF5330, PF2419, PF5773, PF5798, PF0106, PF4284, PF2347, PF2346, PF2819, PF3217, PF3231, PF2566, PF0188, PF5003, PF5687, PF0414, PF0416, PF4051, PF1211, PF1180, PF1210, PF1263, PF3326, PF1133, PF1178, PF3339, PF4095, PF4266, PF4279, PF2844, PF2978, PF1266, PF3008, PF2519, PF3063, PF2523, PF2499, PF2242, PF2239, PF3200, PF3271, PF3269, PF3066, PF3085, PF3353, PF3355, PF3178, PF3400, PF4002, PF3404, PF3212, PF3379, PF3440, PF3376, PF3406, PF5215, PF5192, PF4291, PF4094, PF3301, PF5380, PF3331, PF5260, PF5464, PF0050, PF0435, PF0436, PF2643, PF1121, PF0499, PF0581, PF2676, PF5538, PF0588, PF4343, PF0649, PF0650, PF0870, PF0005, PF0103, PF0967, PF1223, PF2221, PF2217, PF1976, PF1982, PF1991, PF2712, PF1938, PF1999, PF3133, PF5190, PF3138, PF3137, PF2218, PF2456, PF2483, PF2485, PF2490, PF2557, PF2190, PF2761, PF2138, PF2142, PF2531, PF2549, PF2559, PF2558, PF2654, PF3020, PF2981, PF2993, PF2790, PF3021, PF4319, PF2831, PF5334, PF0743, PF5310, PF1600, PF2854, PF5744, PF5315, PF2930, PF2932, PF3388, PF5379, PF5568, PF3205, PF0787, PF2894, PF2925, PF2964, PF5732, PF1454, PF3006, PF3009, PF5916, PF5689, PF5949, PF5965, PF5978, PF3030, PF6313, PF6310, PF3034, PF5979, PF5039, PF0080, PF3067, PF0089, PF3079, PF0182, PF3099, PF6042, PF3118, PF3119, PF3121, PF5461, PF6285, PF3262, PF3264, PF6287, PF6308, PF5845, PF3405, PF5680, PF5788, PF6136, PF6312, PF6314, PF4005, PF4065, PF6148, PF4061, PF4164, PF6317, PF4169, PF6081, PF4130, PF4138, PF5222, PF3174, PF5223, PF6160, PF4136, PF3176, PF4137, PF4048, PF5988, PF5990, PF4424, PF5991, PF5992, PF5635, PF5554, PF1368, PF5781, PF6351, PF5763, PF6355, PF5782, PF6358, PF6396, PF1584, PF0722, PF0754, PF1131, PF2044, PF1959, PF0461, PF5900, PF5902, PF5905, PF5908, PF5896, PF1135, PF1136, PF6082, PF6083, PF6077, PF6115, PF6116, PF1535, PF1017, PF1601, PF0584, PF0453, PF0518, PF0523, PF0756, PF1221, PF2433, PF2452, PF2416, PF1855, PF2165, PF2707, PF3096, PF2555, PF2744, PF2794, PF3414, PF4022, PF0101, PF0257, PF6139, PF3309, PF3316, PF6079, PF0218, PF4237, PF0237, PF0379, PF4259, PF4112, PF4177, PF4183, PF3346, PF0951, PF0958, PF1098, PF0953, PF4243, PF0994, PF0896, PF1171, PF5628, PF5592, PF6217, PF6235, PF6220, PF2100, PF5623, PF1894, PF1895, PF6258, PF1893, PF1097, PF1355, PF1360, PF0714, PF5852, PF5884, PF1488, PF1493, PF5704, PF5692, PF5722, PF5724, PF5597, PF1663, PF5729, PF1658, PF1650, PF0035, PF5752, PF5726, PF5759, PF0747, PF0753, PF1653, PF1655, PF5967, PF1804, PF1879, PF1961, PF1118, PF1950, PF1948, PF6147, PF0450, PF0587, PF6242, PF1892, PF1960, PF1761, PF6173, PF1469, PF1132, PF1973, PF4128, PF1964, PF2338, PF2339, PF2337, PF2365, PF0948, PF2434, PF2192, PF2349, PF2332, PF2686, PF2127, PF2125, PF0411, PF0704, PF0044, PF2612, PF2571, PF0465, PF2678, PF4409, PF2788, PF0193, PF2874, PF2871, PF2867, PF0194, PF0195, PF0725, PF0278, PF0281, PF4217, PF0819, PF0820, PF0821, PF3393, PF0752, PF0965, PF0913, PF0916, PF0956, PF4057, PF0061, PF0961, PF0063, PF5539, PF1532, PF1561, PF1595, PF5883, PF5669, PF5939, PF0083, PF0088, PF5981, PF3051, PF5984, PF6015, PF6058, PF2154, PF6052, PF6151, PF3017, PF3036, PF3047, PF3041, PF3044, PF3080, PF6131, PF3082, PF3088, PF3089, PF3091, PF6341, PF6342, PF6360, PF6362, PF0037, PF0293, PF2071, PF1562, PF1599, PF0135, PF1904, PF0198, PF0200, PF0372, PF3065, PF2472, PF3144, PF3153, PF0390, PF2066, PF2661, PF2656, PF2719, PF2734, PF2745, PF0698, PF2747, PF2748, PF2750, PF3154, PF3155, PF3157, PF0256, PF3236, PF0128, PF0187, PF2998, PF3416, PF0191, PF0174, PF0930, PF0227, PF5154, PF0230, PF0822, PF0231, PF0825, PF0238, PF0240, PF0239, PF0946, PF0392, PF0926, PF2989, PF1038, PF1040, PF1069, PF4151, PF4160, PF1070, PF1078, PF1237, PF1000, PF0597, PF5668, PF5663, PF0599, PF4006, PF0610, PF5716, PF3395, PF4380, PF4328, PF4044, PF1013, PF4039, PF4021, PF1255, PF4111, PF4106, PF0366, PF5878, PF5390, PF0619, PF1148, PF4323, PF0172, PF5933, PF5347, PF0167, PF6093, PF6120, PF6158, PF0621, PF1143, PF1251, PF5348, PF5714, PF5808, PF0921, PF0177, PF5804, PF0179, PF5812, PF5851, PF5848, PF5449, PF5490, PF0178, PF5746, PF5926, PF0421, PF5952, PF5971, PF5931, PF1248, PF6200, PF1246, PF6067, PF5976, PF6237, PF1573, PF1367, PF6331, PF0196, PF1505, PF0368, PF0370, PF1287, PF6068, PF1139, PF0886, PF0849, PF2174, PF1224, PF6198, PF1253, PF1471, PF0803, PF1335, PF1233, PF0299, PF6201, PF1668, PF1467, PF1945, PF0812, PF1812, PF1519, PF0817, PF1321, PF2980, PF0689, PF2600, PF1344, PF1526, PF1777, PF0247, PF1781, PF0357, PF1708, PF0160, PF1810, PF0170, PF1019, PF0265, PF1782, PF0447, PF0445, PF1709, PF0361, PF1861, PF0446, PF1041, PF0729, PF0180, PF0858, PF4052, PF1864, PF0668, PF3303, PF4024, PF0749, PF2395, PF1714, PF2079, PF4185, PF2030, PF5793, PF0100, PF0944, PF0726, PF0693, PF0301, PF0099, PF2114, PF2051, PF2078, PF2412, PF2111, PF0833, PF0398, PF2131, PF0404, PF0403, PF2110, PF0462, PF2112, PF1786, PF0478, PF0483, PF0475, PF1887, PF0346, PF1954, PF0348, PF2298, PF0228, PF2210, PF2031, PF0551, PF0553, PF0356, PF0544, PF0546, PF0547, PF2304, PF2070, PF2294, PF0224, PF0552, PF0556, PF0624, PF2335, PF0376, PF0987, PF2404, PF2195, PF1993, PF2046, PF2058, PF2540, PF2297, PF2869, PF4173, PF0397, PF0400, PF2879, PF2043, PF2918, PF3256, PF3258, PF2027, PF3252, PF2408, PF3104, PF3148, PF3246, PF3275, PF2553, PF0640, PF0659, PF3284, PF0730, PF3116, PF0731, PF2535, PF3115, PF3242, PF3290, PF3328, PF2920, PF3306, PF0831, PF2148, PF3114, PF4113, PF3278, PF0750, PF3324, PF3367, PF4398, PF4418, PF0408, PF3239, PF0442, PF0460, PF3319, PF3310, PF3369, PF4318, PF4413, PF5038, PF5123, PF3366, PF3370, PF3372, PF1134, PF0070, PF3386, PF3387, PF0051, PF0087, PF0448, PF5238, PF5285, PF0505, PF0134, PF0137, PF5289, PF3216, PF0594, PF0670, PF4387, PF0340, PF4351, PF3429, PF5297, PF5287, PF4178, PF4093, PF5679, PF0888, PF5805, PF0850, PF0871, PF0922, PF1056, PF0846, PF0851, PF1034, PF1026, PF2950, PF3109, PF4042, PF5847, PF0890, PF0891, PF4131, PF4132, PF0893, PF1010, PF4158, PF0206, PF4087, PF4062, PF3413, PF3330, PF5713, PF0991, PF4353, PF1007, PF1008, PF5945, PF0979, PF4144, PF1005, PF6059, PF1031, PF1037, PF1039, PF1048, PF0815, PF0852, PF1029, PF1030, PF1036, PF0999, PF4191, PF0915, PF1022, PF1057, PF3427, PF1083, PF0152, PF0980, PF0981, PF1183, PF5422, PF1024, PF6033, PF6038, PF0878, PF1079, PF4218, PF1088, PF1094, PF0874, PF2873, PF0877, PF4449, PF2872, PF0602, PF4255, PF2875, PF5026, PF1109, PF0876, PF1162, PF1161, PF1158, PF4265, PF5075, PF0282, PF5146, PF4282, PF1179, PF4299, PF4276, PF1182, PF1106, PF2061, PF1274, PF1243, PF1280, PF1484, PF1340, PF5475, PF1739, PF1783, PF1736, PF1212, PF1930, PF1738, PF1256, PF1460, PF2284, PF2251, PF2320, PF2253, PF2345, PF2281, PF2515, PF2278, PF2498, PF5879, PF5930, PF2518, PF2282, PF5836, PF5857, PF2273, PF2816, PF2812, PF2813, PF2818, PF2397, PF2814, PF2398, PF5929, PF2817, PF2340, PF1337, PF3229, PF6325, PF3233, PF1314, PF1540, PF1546, PF3343, PF6272, PF5339, PF5031, PF5233, PF1292, PF5242, PF1313, PF1543, PF1853, PF5305, PF1293, PF5237, PF1552, PF4316, PF1336, PF5313, PF0767, PF1303, PF1316, PF1317, PF5817, PF1290, PF5451, PF1295, PF1474, PF5844, PF5870, PF1363, PF5937, PF6210, PF1275, PF0580, PF1310, PF1391, PF1456, PF0454, PF0434, PF0441, PF0488, PF0848, PF1357, PFVV0013, PF0489, PF1399, PF1458, PF1261, PF0663, PF1381, PF1445, PF0910, PF1639, PF1438, PF1444, PF1851, PF1850, PF1442, PF0968, PF0835, PF1472, PF5332, PF1538, PF1835, PF5344, PF5346, PF5377, PF0857, PF1621, PF2363, PF1620, PF0832, PF1284, PF1306, PF1674, PF0853, PF1397, PF1453, PF1636, PF1899, PF1633, PF1551, PF1558, PF1289, PF1632, PF1638, PF1567, PF1671, PF1300, PF1635, PF5456, PF1487, PF1680, PF2041, PF1937, PF1665, PF1670, PF2059, PF1296, PF5448, PF5454, PF1698, PF1694, PF0927, PF3130, PF1695, PF2437, PF2438, PF3102, PF2197, PF2241, PF2597, PF2598, PF1726, PF1725, PF1733, PF2532, PF1742, PF5476, PF5481, PF1746, PF2660, PF1741, PF1732, PF1787, PF5016, PF1729, PF2263, PF2685, PF0959, PF1737, PF2258, PF0960, PF5059, PF2257, PF2256, PF2693, PF1790, PF1764, PF2266, PF1829, PF4431, PF4434, PF2705, PF4437, PF2710, PF0983, PF1770, PF1800, PF4322, PF1749, PF1797, PF2168, PF1844, PF1768, PF2376, PF2949, PF0420, PF0529, PF0413, PF1815, PF5044, PF5012, PF3018, PF1881, PF2497, PF2708, PF2442, PF2443, PF2444, PF1798, PF2446, PF5004, PF1885, PF5008, PF3110, PF5092, PF3052, PF3195, PF1883, PF1841, PF1299, PF1882, PF5074, PF1846, PF2163, PF1842, PF1845, PF1913, PF1914, PF1320, PF1323, PF1942, PF1837, PF2149, PF2150, PF1943, PF1922, PF1946, PF1364, PF1923, PF1921, PF1699, PF5093, PF2581, PF5095, PF5091, PF1988, PF1920, PF5105, PF1916, PF5116, PF1978, PF1910, PF5127, PF1994, PF1060, PF1061, PF2248, PF1977, PF1066, PF1970, PF1980, PF1981, PF2020, PF1068, PF2008, PF1724, PF1728, PF5513, PF2028, PF2029, PF2636, PF2343, PF2220, PF2641, PF1987, PF5639, PF5681, PF2032, PF5686, PF2037, PF5637, PF2033, PF5641, PF5644, PF2060, PF2038, PF2034, PF2075, PF5701, PF5715, PF1087, PF5736, PF3081, PF5541, PF0021, PF2711, PF2717, PF0056, PF2342, PF0147, PF0149, PF2292, PF2318, PF2528, PF5751, PF5551, PF5766, PF2279, PF2315, PF2928, PF2380, PF2929, PF2313, PF2936, PF0235, PF2375, PF2940, PF2334, PF5558, PF2967, PF5584, PF2947, PF2312, PF2285, PFVV0004, PF2368, PF3266, PF2459, PF3254, PF2401, PF2955, PF2290, PF2457, PF2468, PF1282, PF1286, PF2471, PF5757, PF5769, PF5772, PF2999, PF5775, PF2862, PF2277, PF2458, PF2461, PF2463, PF2469, PF2921, PF2951, PF2445, PF2399, PF1325, PF5838, PF5779, PF2615, PF3014, PF2544, PF1324, PF1326, PF2547, PF5810, PF2341, PF1329, PF2146, PF2171, PF2069, PF2177, PF5856, PF2086, PF5977, PF2093, PF2091, PF2169, PF2180, PF0248, PF6001, PF2230, PF6016, PF2235, PF6027, PF6049, PF2234, PF5888, PF2226, PF2250, PF2227, PF2246, PF2252, PF2247, PF0074, PF2589, PF0258, PF2613, PF2388, PF2510, PF2516, PF2517, PF2512, PF0945, PF0947, PF2543, PF2522, PF2617, PF5609, PF2618, PF5615, PF2603, PF1215, PF5282, PF1014, PF1236, PF1241, PF1216, PF2596, PF1217, PF2614, PF6149, PF6156, PF6161, PF2608, PF2567, PF2552, PF1071, PF2465, PF2624, PF2621, PF2620, PF1239, PF2683, PF2631, PF6095, PF6135, PF2675, PF2674, PF2703, PF2729, PF2727, PF2715, PF6185, PF2716, PF2721, PF2134, PF2109, PF2698, PF6255, PF0943, PF0521, PF2688, PF6230, PF2756, PF2759, PF0932, PF6260, PF2755, PF1072, PF2766, PF2767, PF2763, PF2778, PF1244, PF2780, PF2781, PF2776, PF1503, PF2775, PF2789, PF2770, PF2753, PF2798, PF1322, PF2822, PF1234, PF2808, PF2830, PF2802, PF2804, PF2838, PF6282, PF2848, PF2827, PF2821, PF2823, PF1350, PF2851, PF2853, PF1778, PF2845, PF2849, PF2857, PF1058, PF1020, PF0977, PF0674, PF1025, PF0816, PF0183, PF0861, PF1506, PF2880, PF1706, PF2882, PF2891, PF1490, PF2912, PF1811, PF2884, PF2886, PF1807, PF2897, PF2889, PF2911, PF2881, PF1705, PF1808, PF1780, PF1863, PF2933, PF1929, PF2047, PF6363, PF6376, PF6413, PF2115, PF2914, PF1931, PF1050, PF1055, PF2629, PF1021, PF1033, PF2539, PF1027, PF2931, PF1043, PF1051, PF1046, PF1052, PF1023, PF2974, PF0419, PF1044, PF2215, PF2944, PF2211, PF2926, PF2977, PF2910, PF1064, PF2969, PF2976, PF6406, PF6416, PF2204, PF1042, PF1110, PF6419, PF2792, PF2303, PF0885, PF0882, PF2987, PF0875, PF0583, PF0691, PF0873, PF3005, PF3000, PF0692, PF0688, PF2994, PF2113, PF1862, PF1107, PF2991, PF0538, PF3003, PF1080, PF1157, PF1349, PF1409, PF3039, PF1093, PF1416, PF1516, PF1717, PF2348, PF1160, PF1515, PF1159, PF1164, PF2238, PF1167, PF3004, PF2300, PF3056, PF0098, PF2400, PF2072, PF2959, PF2074, PF0558, PF3037, PF1186, PF1181, PF3071, PF1165, PF1205, PF1269, PF1278, PF1273, PF1194, PF2305, PF3042, PF2089, PF3060, PF1213, PF2352, PF3077, PF2293, PF2410, PF1268, PF3173, PF3170, PF2283, PF2275, PF2296, PF2538, PF3188, PF2331, PF3183, PF2449, PF1315, PF1260, PF3213, PF1338, PF3093, PF4249, PF4230, PF1312, PF3075, PF3167, PF3187, PF3162, PF3163, PF1288, PF3164, PF3196, PF1479, PF3064, PF3279, PF3192, PF1311, PF3287, PF0684, PF5421, PF3286, PF3277, PF3108, PF3318, PF3208, PF5360, PF1375, PF3321, PF3098, PF3323, PF3289, PF5429, PF3285, PF1374, PF4192, PF3206, PF1455, PF3283, PF3101, PF3190, PF3141, PF0700, PF3273, PF1475, PF5372, PF5503, PF5511, PF3234, PF5437, PF0777, PF1585, PF3073, PF1439, PF1443, PF0735, PF3304, PF3232, PF1541, PF3307, PF3332, PF0487, PF3237, PF1281, PF3227, PF1345, PF3396, PF1617, PF1385, PF1614, PF3409, PF3240, PF1589, PF1587, PF1643, PF1400, PF3394, PF1605, PF1691, PF3425, PF5262, PF1672, PF3375, PF1581, PF2635, PF2642, PF1675, PF4014, PF5201, PF2785, PF4016, PF1682, PF3029, PF4029, PF2537, PF1681, PF4023, PF4345, PF1683, PF4030, PF4034, PF1604, PF1294, PF1697, PF4097, PF4041, PF1721, PF5288, PF0159, PF4066, PF4074, PF1722, PF4119, PF4115, PF1731, PF5306, PF5535, PF4116, PF5296, PF4121, PF1734, PF1792, PF4104, PF4118, PF4140, PF1766, PF1747, PF4122, PF1779, PF1789, PF1750, PF1788, PF0768, PF4145, PF0770, PF4150, PF4165, PF4161, PF0279, PF4167, PF0439, PF0773, PF0695, PF1767, PF1801, PF0774, PF4355, PF4314, PF4350, PF0484, PF1771, PF1794, PF1793, PF0872, PF0784, PF1795, PF1840, PF0778, PF4404, PF1796, PF5345, PF1884, PF2267, PF5291, PF1888, PF2193, PF0807, PF1871, PF0808, PF1849, PF1406, PF1394, PF1889, PF5032, PF1402, PF1839, PF1843, PF1912, PF5709, PF1919, PF5708, PF1917, PF1918, PF1685, PF1814, PF1915, PF5473, PF1925, PF1952, PF0406, PF1928, PF1559, PF1688, PF5293, PF0360, PF1869, PF1926, PF1865, PF1909, PF1924, PF1867, PF1996, PF1957, PF5866, PF2000, PF2188, PF1983, PF1975, PF1984, PF6073, PF2394, PF2384, PF2385, PF2182, PF2022, PF2118, PF2013, PF2026, PF2137, PF6096, PF2139, PF2010, PF2016, PF2128, PF6155, PF2035, PF2040, PF2062, PF2140, PF2064, PF1947, PF0986, PF2383, PF2014, PF2009, PF2120, PF2001, PF2995, PF3001, PF3053, PF3058, PF3070, PF2225, PF0438, PF2370, PF2372, PF2291, PF2314, PF2336, PF3068, PF2403, PF2289, PF2364, PF2288, PF2107, PF2317, PF2209, PF1240, PF2295, PF1238, PF1302, PF1307, PF2464, PF1308, PF2366, PF2396, PF1431, PF2473, PF2405, PF1382, PF1378, PF2373, PF1372, PF1389, PF0656, PF1769, PF1735, PF1776, PF2402, PF2390, PF2208, PF2611, PF1908, PF2460, PF1866, PF2562, PF2323, PF2415, PF2607, PF2511, PF2455, PF2411, PF2892, PF0883, PF3347, PF0795, PF0889, PF2470, PF5336, PF5060, PF2065, PF5136, PF2082, PF2178, PF4347, PF2184, PF3419, PF2172, PF4352, PF2158, PF4357, PF4399, PF2167, PF5647, PF2175, PF2092, PF2176, PF5814, PF5871, PF5961, PF5962, PF2094, PF0963, PF5960, PF0969, PF6101, PF2073, PF2173, PF2179, PF2229, PF2249, PF0112, PF2254, PF2228, PF2232, PF2224, PF2181, PF0226, PF2619, PF2548, PF2545, PF3245, PF2586, PF2602, PF2594, PF2638, PF5432, PF5424, PF2637, PF5496, PF2601, PF5512, PF5537, PF2610, PF5318, PF5308, PF2462, PF2554, PF2625, PF2671, PF2592, PF2633, PF2630, PF1465, PF1466, PF5796, PF5536, PF2677, PF2680, PF5533, PF2672, PF2682, PF2669, PF2702, PF2714, PF2709, PF5826, PF2730, PF2724, PF5831, PF5837, PF5839, PF2725, PF2670, PF2507, PF2765, PF5938, PF2758, PF6105, PF2762, PF6107, PF6179, PF1602, PF1609, PF6181, PF2754, PF6206, PF6186, PF0168, PF0169, PF6193, PF1245, PF2774, PF2773, PF2777, PF2764, PF2768, PF4358, PF2772, PF2779, PF2791, PF1664, PF2800, PF2829, PF2815, PF2793, PF0931, PF2797, PF0933, PF2799, PF2801, PF2810, PF5510, PF2803, PF2805, PF2824, PF2841, PF2045, PF2828, PF6099, PF1101, PF2825, PF2835, PF2842, PF2787, PF2850, PF6279, PF2852, PF1838, PF0895, PF0574, PF2855, PF1852, PF2847, PF0706, PF2757, PF0713, PF2806, PF2807, PF2834, PF2843, PF2861, PF2878, PF2898, PF2895, PF2890, PF2904, PF1900, PF2887, PF1941, PF2913, PF2919, PF2888, PF0702, PF2905, PF2915, PF1509, PF2885, PF2948, PF2954, PF1301, PF2050, PF2938, PF2276, PF1955, PF1985, PF1318, PF2934, PF2945, PF2942, PF1972, PF5258, PF4389, PF1723, PF1902, PF1936, PF3023, PF2319, PF2996, PF3007, PF2453, PF2475, PF3012, PF2476, PF3015, PF3016, PF2551, PF2260, PF3100, PF3040, PF3061, PF3062, PF1763, PF3002, PF2616, PF3078, PF2448, PF2262, PF2259, PF2574, PF3072, PF2261, PF3057, PF2627, PF2728, PF2743, PF2441, PF2430, PF2431, PF3143, PF3074, PF3156, PF3159, PF3171, PF3151, PF3152, PF3199, PF3125, PF3127, PF3327, PF3117, PF2496, PF2447, PF3185, PF2451, PF3348, PF3211, PF2271, PF3181, PF2264, PF3410, PF3124, PF4258, PF4250, PF5298, PF4280, PF2988, PF5514, PF3180, PF3241, PF2160, PF2161, PF3202, PF3095, PF4410, PF2196, PF5035, PF2147, PF2151, PF2152, PF3292, PF3293, PF3150, PF5030, PF3297, PF0856, PF3338, PF3296, PF2572, PF2575, PF3354, PF2578, PF3349, PF3378, PF3380, PF3364, PF3374, PF2536, PF3408, PF5920, PF6002, PF3407, PF5304, PF4010, PF4000, PF4019, PF6348, PF3428, PF4031, PF4020, PF4028, PF4033, PF3430, PF4037, PF4038, PF4077, PF4045, PF4105, PF4079, PF4078, PF0701, PF4107, PF4070, PF4073, PF2311, PF2713, PF2722, PF4120, PF4139, PF4141, PF4142, PF4143, PF5392, PF2520, PF4162, PF5435, PF2937, PF4126, PF2939, PF2935, PF1713, PF4163, PF4072, PF4155, PF4166, PF4174, PF2982, PF2971, PF2973, PF4179, PF4159, PF4168, PF4205, PF4170, PF4210, PF4208, PF3445, PF0535, PF1476, PF4236, PF1514, PF1934, PF0540, PF4211, PF4228, PF0302, PF4213, PF4224, PF4227, PF0542, PF2903, PF5428, PF5441, PF1114, PF5402, PF5410, PF5387, PF1150, PF5393, PF1152, PF5412, PF5394, PF5395, PF5405, PF5411, PF5413, PF1297, PF1249, PF5433, PF1383, PF5439, PF1480, PF5460, PF5400, PF2144, PF2145, PF1906, PF5040, PF5251, PF1049, PF5500, PF5497, PF0310, PF0321, PF1413, PF0592, PF0681, PF0838, PF0395, PF5524, PF0634, PF0790, PF0202, PF0528, PF0779, PF0067, PF3184, PF0901, PF1492, PF5187, PF2909, PF5206, PF5208, PF5210, PF5213, PF2696, PF5170, PF5211, PF5086, PF5406, PF5184, PF5120, PF5321, PF0242, PF5340, PF0269, PF0272, PF4233, PF5355, PF4268, PF5364, PF5385, PF5368, PF0244, PF0495, PF5363, PF5370, PF5399, PF5357, PF3334, PF5362, PF5434, PF4445, PF4088, PF3449, PF4307, PF4296, PF4330, PF4235, PF4391, PF4298, PF4331, PF4310, PF4370, PF4289, PF4329, PF4248, PF4332, PF4277, PF4438, PF4441, PF4446, PF5052, PF5053, PF5023, PF5025, PF5056, PF5054, PF5001, PF4421, PF4363, PF4417, PF5118, PF5122, PF5018, PF5070, PF5133, PF5034, PF5128, PF5131, PF4365, PF5019, PF5081, PF5084, PF5106, PF5102, PF5114, PF5167, PF5057, PF5066, PF5085, PF5099, PF2542, PF5000, PF5036, PF5055, PF5062, PF5119, PF5157, PF5079, PF5083, PF5002, PF5061, PF5077, PF5080, PF5096, PF5155, PF5100, PF5271, PF5274, PF0084, PF5505, PF5523, PF5516, PF5520, PF5526, PF5528, PF4290, PF1108, PF5565, PF3161, PF3191, PF5559, PF5562, PF5566, PF5567, PF5570, PF5574, PF5578, PF2328, PF5580, PF5585, PF5596, PF5600, PF5602, PF5603, PF4382, PF5477, PF5889, PF0622, PF5754, PF5762, PF5699, PF5700, PF0320, PF0605, PF0255, PF0274, PF0312, PF0313, PF0323, PF0326, PF2183, PF0919, PF0798, PF0923, PF1366, PF0911, PF0924, PF0996, PF1449, PF0303, PF0304, PF1155, PF1265, PF0374, PF2237, PF0192, PF0791, PF0508, PF2622, PF0512, PF2647, PF2648, PF2650, PF0793, PF0796, PF2771, PF0936, PF0937, PF0939, PF3435, PF1067, PF1065, PF1270, PF5161, PF1272, PF4447, PF5015, PF5152, PF4273, PF0155, PF1395, PF0029, PF0284, PF4381, PF1327, PF4341, PF4427, PF4428, PF1347, PF6138, PF1418, PF5799, PF5795, PF5854, PF5322, PF1504, PF5326, PF5569, PF1510, PF0613, PF6114, PF0744, PF1564, PF1117, PF1588, PF0186, PF0573, PF1629, PF1751, PF0761, PF0764, PF0763, PF0759, PF0766, PF1753, PF1754, PF0335, PF0334, PF0333, PF0498, PF0456, PF2076, PF0504, PF0568, PF0283, PF2006, PF0757, PF0019, PF1817, PF1989, PF0207, PF0093, PF0104, PF0143, PF0148, PF0145, PF2326, PF0203, PF0311, PF2133, PF2327, PF2329, PF2414, PF2479, PF2534, PF1151, PF2576, PF2579, PF1242, PF1250, PF2155, PF5828, PF5840, PF1578, PF0018, PF0016, PF3422, PF0022, PF0024, PF0040, PF0131, PF0046, PF2153, PF5279, PF0047, PF4348, PF5452, PF5842, PF5843, PF0011, PF0008, PF0384, PF5846, PF5994, PF6045, PF5974, PF5969, PF6330, PF6286, PF1115, PF0141, PF0146, PF0140, PF0662, PF0973, PF0990, PF1129, PF0539, PF1077, PF1227, PF1365, PF1618, PF1647, PF1369, PF1370, PF2700, PF0163, PF0382, PF0394, PF3146, PF0175, PF3122, PF3097, PF0525, PF0524, PF0527, PF3272, PF5178, PF5776, PF5272, PF0909, PF1085, PF1103, PF1111, PF0276, PF1219, PF0289, PF0294, PF0935, PF1701, PF1704, PF0316, PF1715, PF0324, PF1933, PF2017, PF0531, PF0507, PF0589, PF2141, PF2529, PF0339, PF0315, PF0367, PF3223, PF0355, PF5341, PF0611, PF5903, PF5922, PF3209, PF0738, PF1187, PF0732, PF1163, PF2691, PF3165, PF1184, PF1209, PF3352, PF3267, PF0055, PF3308, PF3312, PF3313, PF0006, PF4082, PF4086, PF4091, PF0491, PF1553, PF5263, PF0132, PF0185, PF5543, PF5546, PF5382, PF5041, PF1226, PF3336, PF0862, PF2957, PF6053, PF5959, PF5181, PF4293, PF5936, PF6004, PF5882, PF5239, PF1854, PF3210, PF6134, PF6281, PF3280, PF0377, PF3282, PF5264, PF0894, PF5219, PF5253, PF4324, PF2687, PF4366, PF4080, PF4269, PF4374, PF4262, PF4377, PF4263, PF4264, PF3342, PF4312, PF4089, PF4018, PF4440, PF5745, PF5834, PF5235, PF5465, PF5970, PF5376, PF5593, PF5853, PF5850, PF5910, PF5983, PF5046, PF6005, PF6007, PF6085, PF0843, PF6233, PF6284, PF6315, PF6323, PF0682, PF0788, PF0383, PF0845, PF0440, PF1137, PF1175, PF5823, PF1172, PF1073, PF1076, PF1075, PF1547, PF1457, PF0519, PF1944, PF2377, PF2166, PF2156, PF5691, PF2157, PF5690, PF0039, PF2506, PF0048, PF0052, PF5771, PF0107, PF0295, PF0363, PF0081, PF0079, PF3260, PF4047, PF5912, PF3103, PF3221, PF5973, PF5975, PF5980, PF5985, PF5987, PF0042, PF4408, PF5606, PF5622, PF5626, PF5624, PF5666, PF5674, PF5665, PF5784, PF5785, PF5787, PF5786, PF5794, PF5703, PF5661, PF5706, PF5694, PF5697, PF5728, PF5725, PF0095, PF5743, PF5747, PF5748, PF5750, PF5753, PF5749, PF5730, PF0133, PF6089, PF0369, PF1529, PF6392, PF6078, PF1594, PF1666, PF6111, PF6219, PF6267, PF6273, PF0437, PF6080, PF0814, PF0801, PF0802, PF6302, PF0740, PF6347, PF0755, PF0800, PF0806, PF0562, PF0864, PF0900, PF0557, PF0962, PF0559, PF0560, PF0561, PF0563, PF1016, PF0978, PF0245, PF1464, PF1461, PF1462, PF1463, PF1533, PF1596, PF1468, PF1612, PF1575, PF1574, PF1597, PF2123, PF3013, PF3131, PF2563, PF2417, PF2240, PF2582, PF2474, PF2584, PF2585, PF2525, PF2527, PF2560, PF3026, PF3027, PF3046, PF3043, PF0618, PF3083, PF3084, PF0616, PF3086, PF3087, PF3094, PF0617, PF2689, PF2922, PF0623, PF0666, PF0667, PF0173, PF0280, PF0804, PF0277, PF0300, PF3224, PF0809, PF0811, PF3276, PF0176, PF4103, PF0219, PF4108, PF0220, PF4109, PF0225, PF4411, PF4339, PF0229, PF5220, PF4175, PF3300, PF3305, PF5182, PF5458, PF5194, PF0350, PF0699, PF5809, PF0818, PF0826, PF0432, PF0444, PF5955, PF5953, PF5940, PF0443, PF5651, PF5664, PF5671, PF5707, PF5982, PF5710, PF0449, PF6021, PF0854, PF0373, PF5935, PF6084, PF6071, PF6057, PF0387, PF6094, PF6129, PF0866, PF0391, PF0748, PF0860, PF0736, PF6343, PF6359, PF6300, PF6338, PF6298, PF6297, PF6296, PF0970, PF0184, PF1813, PF0401, PF0402, PF1565, PF2021, PF0407, PF0925, PF1901, PF1905, PF0113, PF2605, PF2718, PF2735, PF2736, PF0473, PF2737, PF0476, PF2738, PF2746, PF2749, PF2751, PF0477, PF0480, PF0221, PF0222, PF0554, PF0548, PF0549, PF0550, PF0036, PF0385, PF4027, PF4026, PF3402, PF4046, PF4147, PF4340, PF5653, PF5662, PF5659, PF5807, PF5670, PF5927, PF5923, PF5904, PF5901, PF5928, PF5963, PF5956, PF6090, PF6025, PF6076, PF6271, PF2533, PF2731, PF0905, PF1470, PF3045, PF2162, PF2164, PF2604, PF2489, PF4180, PF4181, PF4053, PF3259, PF5446, PF5455, PF4060, PF0697, PF0723, PF0829, PF0472, PF0660, PF0824, PF0988, PF0985, PF1525, PF0216, PF4219, PF0062, PF0059, PF0142, PF4369, PF0337, PF4102, PF3092, PF3325, PF5498, PF0625, PF0685, PF0742, PF0745, PF0746, PF0805, PF5300, PF0405, PF0502, PF0847, PF0517, PF0490, PF5147, PF0665, PF0645, PF0810, PF1173, PF1235, PF1225, PF1254, PF1346, PF1520, PF1521, PF1522, PF6087, PF6088, PF1572, PF1757, PF5872, PF1577, PF1625, PF1570, PF1640, PF6276, PF1848, PF1956, PF1896, PF1958, PF2053, PF0760, PF0786, PF3129, PF2436, PF2439, PF2244, PF2255, PF2599, PF2658, PF2505, PF2664, PF2706, PF2923, PF2924, PF2883, PF2427, PF3106, PF3049, PF3107, PF0993, PF0418, PF0057, PF0422, PF0424, PF0308, PF0329, PF0309, PF0354, PF0380, PF3265, PF4221, PF3357, PF3361, PF3385, PF3359, PF0464, PF3403, PF4069, PF4011, PF0514, PF4201, PF5463, PF5467, PF5544, PF1053, PF1059, PF5457, PF3214, PF3220, PF5121, PF5139, PF5166, PF5198, PF1074, PF1082, PF5230, PF5327, PF5335, PF5366, PF5367, PF5378, PF3048, PF1276, PF5028, PF4372, PF4375, PF5037, PF5124, PF5141, PF5212, PF1291, PF5552, PF1319, PF5650, PF5629, PF5630, PF5634, PF5875, PF0555, PF5986, PF6009, PF6040, PF0606, PF0603, PF0604, PF6306, PF0636, PF0642, PF0629, PF0683, PF0733, PF0579, PF0410, PF0474, PF0457, PF0469, PF0516, PF0690, PF0596, PF0964, PF0669, PF1220, PF1120, PF1555, PF1271, PF0708, PF0709, PF0711, PF1667, PF1661, PF0715, PF1986, PF0720, PF2005, PF2896, PF2199, PF2367, PF2371, PF2690, PF2695, PF2002, PF1035, PF1081, PF1090, PF0058, PF0060, PF0090, PF0162, PF0164, PF0215, PF0212, PF0259, PF0260, PF0344, PF2170, PF4007, PF2595, PF2684, PF3193, PF3194, PF5191, PF5172, PF5236, PF0881, PF4385, PF4383, PF5797, PF5803, PF5824, PF5859, PF5688, PF5553, PF6178, PF6196, PF1128, PF6172, PF1127, PF6164, PF6204, PF6177, PF6249, PF6250, PF2425, PF6344, PF1141, PF1142, PF1144, PF1146, PF1145, PF1193, PF1190, PF1198, PF1200, PF1199, PF1204, PF1203, PF1228, PF1259, PF1333, PF1353, PF1356, PF2652, PF1377, PF1419, PF1398, PF1426, PF2361, PF2392, PF2487, PF2450, PF1433, PF1424, PF1428, PF1446, PF1448, PF1435, PF1447, PF3438, PF1459, PF4081, PF4092, PF1542, PF1544, PF1568, PF1563, PF1571, PF1590, PF0661, PF0741, PF1613, PF0813, PF0305, PF0459, PF1304, PF1341, PF2019, PF1968, PF2418, PF2432, PF2216, PF2524, PF1649, PF1645, PF2655, PF2877, PF2899, PF2900, PF1679, PF1684, PF1693, PF1696, PF1702, PF0292, PF3142, PF3415, PF4004, PF4035, PF4032, PF4152, PF0506, PF1745, PF0887, PF0912, PF1760, PF1138, PF1773, PF1775, PF1550, PF1819, PF1784, PF1785, PF2351, PF1949, PF2124, PF2435, PF1799, PF1806, PF2231, PF1809, PF2723, PF2997, PF1822, PF1828, PF1826, PF1833, PF1821, PF1825, PF1831, PF1836, PF1847, PF1857, PFVV0016, PFVV0015, PF1940, PF1966, PF2018, PF2024, PF3028, PF2893, PF3270, PF2049, PF2057, PF3371, PF4009, PF4003, PF3391, PF4015, PF3382, PF3381, PF3417, PF4084, PF4013, PF4194, PF3363, PF3384, PF5244, PF5252, PF5248, PF5188, PF5205, PF5193, PF5268, PF5273, PF4186, PF4096, PF3377, PF5342, PF0678, PF0679, PF1123, PF0620, PF0646, PF0648, PF0651, PF0865, PF0578, PF0966, PF2478, PF2484, PF2203, PF1342, PF1579, PF1580, PF1619, PF1979, PF1939, PF3132, PF3134, PF3135, PF3105, PF2222, PF2454, PF2482, PF2495, PF2565, PF2570, PF2509, PF2514, PF2550, PF2198, PF2956, PF2958, PF3025, PF2504, PF2486, PF2508, PF2233, PF2321, PF2322, PF2325, PF2353, PF2354, PF2649, PF2659, PF0261, PF3446, PF2739, PF2740, PF2741, PF3420, PF4067, PF4114, PF4133, PF5391, PF5397, PF3149, PF3189, PF5176, PF5177, PF5221, PF5224, PF5226, PF5331, PF5381, PF4110, PF4117, PF4135, PF2795, PF2811, PF4059, PF2833, PF5765, PF5767, PF5768, PF5774, PF2809, PF2820, PF2832, PF5907, PF5909, PF5841, PF6119, PF6122, PF2866, PF6215, PF6253, PF2858, PF6295, PF2860, PF6336, PF6289, PF0974, PF1549, PF1856, PF2243, PF2556, PF2946, PF2943, PF2129, PF2927, PF2952, PF2963, PF2972, PF3022, PF3031, PF3033, PF0343, PF0371, PF3315, PF4246, PF3038, PF3090, PF3399, PF0950, PF0955, PF0957, PF3123, PF1168, PF1104, PF0903, PF0899, PF0904, PF0509, PF1105, PF1170, PF1267, PF2099, PF1174, PF3179, PF1102, PF3222, PF1359, PF3228, PF1361, PF1362, PF1420, PF1500, PF1441, PF1440, PF1499, PF1502, PF1690, PF1687, PF1678, PF1660, PF1659, PF1657, PF1656, PF1805, PF1802, PF1765, PF1880, PF1877, PF1875, PF1874, PF1606, PF1971, PF2983, PF2362, PF1897, PF2106, PF2102, PF2103, PF2101, PF2104, PF2583, PF2609, PF2679, PF2568, PF2941, PF3069, PF3247, PF3207, PF4124, PF5427, PF5484, PF5045, PF5065, PF5137, PF5149, PF4378, PF4388, PF4334, PF4320, PF5488, PF6064, PF6197, PF6056, PF0362, PF0378, PF0032, PF1197, PF2194, PF2784, PF2907, PF2865, PF2870, PF2916, PF0078, PF0077, PF0138, PF0157, PF0171, PF0166, PF0249, PF0266, PF0347, PF0351, PF0364, PF0365, PF0396, PF0399, PF3249, PF3257, PF3136, PF3291, PF3288, PF3294, PF3295, PF3314, PF3401, PF4301, PF4313, PF4349, PF4354, PF4333, PF4401, PF4442, PF3311, PF3320, PF3335, PF3441, PF4036, PF4040, PF4123, PF3433, PF4196, PF4214, PF4187, PF4190, PF4229, PF4245, PF4256, PF4257, PF4271, PF4267, PF4270, PF4275, PF4193, PF4281, PF4274, PF5483, PF5486, PF5487, PF5504, PF5507, PF5269, PF5295, PF5324, PF5338, PF5373, PF5415, PF5416, PF5419, PF5453, PF5470, PF5478, PF5502, PF4272, PF4448, PF5009, PF5014, PF5006, PF5010, PF5020, PF5090, PF5098, PF5103, PF5101, PF5132, PF5506, PF5636, PF5638, PF5640, PF5642, PF5643, PF5645, PF5545, PF5564, PF5577, PF5579, PF5583, PF5780, PF5811, PF5881, PF5588, PF5587, PF5683, PF5723, PF5731, PF5540, PF5542, PF5556, PF5738, PF5886, PF5932, PF6044, PF6006, PF6063, PF5284, PF6162, PF6163, PF5375, PF5374, PF6188, PF6194, PF6221, PF6244, PF6278, PF6251, PF6263, PF6269, PF6328, PF6334, PF6332, PF6335, PF6345, PF6356, PF0585, PF0680, PF0002, PF1092, PF1156, PF1421, PF1496, PF2048, PF2270, PF1963, PF2308, PF1974, PF2668, PF2667, PF2863, PF2665, PF3158, PF2265, PF3175, PF3186, PF5319, PF4288, PF1495, PF5515, PF5430, PF5423, PF0898, PF0907, PF1348, PF1518, PF1755, PF1886, PF2052, PF2077, PF2623, PF2639, PF2645, PF3024, PF0139, PF0285, PF0288, PF3274, PF0545, PF0765, PF0252, PF0734, PF0771, PF0781, PF0867, PF0479, PF0481, PF0543, PF0776, PF0868, PF1332, PF2214, PF2191, PF1393, PF1429, PF1401, PF1358, PF1557, PF1404, PF1392, PF1403, PF1556, PF1689, PF1872, PF1858, PF1873, PF1992, PF1997, PF2186, PF2207, PF2387, PF2007, PF2185, PF2117, PF2119, PF2386, PF2039, PF2042, PF2382, PF2011, PF2015, PF2389, PF2378, PF0270, PF3054, PF3177, PF5425, PF5417, PF5301, PF5303, PF5820, PF5531, PF5968, PF5972, PF5911, PF6137, PF6141, PF6176, PF6154, PF3219, PF3215.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD Phoenix™ M50 Automated Microbiology System 443624Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION3 recalls417 reports
Other devices under the same FDA submission
4 devices cleared under K131331, K040099, K020321 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BD Phoenix™ AST Broth 246003Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION182 reports
- BD Phoenix™ AST Broth 4.5 mL 246011Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION42 reports
- BD Phoenix™ AST Indicator Solution 246004Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION15 reports
- BD Phoenix™ AP AST Indicator Solution 246006Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95636).
- Z-0584-2025 · Open, ClassifiedBD EpiCenter Single User Software Version or Model: 441007 Catalog Number: 441007
- Z-0585-2025 · Open, ClassifiedBD EpiCenter Data Management System Version or Model: 440887 Catalog Number: 440887
- Z-0586-2025 · Open, ClassifiedBD EpiCenter Data Management System, Multi User Software Version or Model: 440981 Catalog Number: 440981
- Z-0587-2025 · Open, ClassifiedBD COR PX Instrument Version or Model: 443988 Catalog Number: 443988
- Z-0588-2025 · Open, ClassifiedBD COR MX Instrument Version or Model: 443989 Catalog Number: 443989
- Z-0589-2025 · Open, ClassifiedBD COR GX Instrument Version or Model: 443990 Catalog Number: 443990
- Z-0590-2025 · Open, ClassifiedBD MAX System, BD MAX Instrument Version or Model: 441916 Catalog Number: 441916
- Z-0591-2025 · Open, ClassifiedBD Synapsys Informatics Solution (Laboratory Solution) Version or Model: 444150 Catalog Number: 444150
- Z-0593-2025 · Open, ClassifiedBD BACTEC FX, Instrument Top, Packaged Version or Model: 441385 Catalog Number: 441385
- Z-0594-2025 · Open, ClassifiedBD BACTEC FX40 Instrument Version or Model: 442296 Catalog Number: 442296
- Z-0595-2025 · Open, ClassifiedBACTEC 9120 System Version or Model: 445570 Catalog Number: 445570
- Z-0596-2025 · Open, ClassifiedBACTEC 9240 System Version or Model: 445475 Catalog Number: 445475
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page