BACTEC 9240 System recallZ-0596-2025 · Becton Dickinson & Co. · initiated Oct 31, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.

FDA root cause category: Software design

What the company is doing

On October 31, 2024 URGENT" MEDICAL DEVICE PRODUCT CORRECTION letters were sent to consignees. IT, Safety and Security Actions to be Taken: BD encourages customers to follow best practices for maintaining strong security measures to protect hospital networks and medical devices including: "Ensure access to potentially vulnerable devices is limited to authorized personnel "Inform authorized users of issue, and ensure all relevant passwords are tightly controlled "Monitor and log network traffic attempting to reach medical device managementenvironments for suspicious activity "Where possible, isolate affected devices in a secure VLAN or behind firewalls with restrictedaccess that only permits communication with trusted hosts in other networks when needed "Impacted devices do not require use of RDP ports and these should be disabled or blockedif enabled "Ensure permissions on file shares are appropriately established and enforced, and monitorand log access for evidence of suspicious activity "Disconnect devices from the network if connectivity is not necessary Next Steps to be Taken by Customer: 1.Ensure the contents of this notification are read and understood. 2.Share and post this customer letter within your facility network including IT and safety andsecurity department. Forward to any customers to whom you may have distributed theproduct(s) to ensure awareness. 3.Complete the attached Customer Response Form and return it to the BD contact noted on theForm so that BD may acknowledge your receipt of this notification per FDA requirements. 4.Report any adverse health consequences experienced with the use of these products to BD.Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 If you require further assistance, please contact: Technical Service +1 888 451 5514 Technical Questi

Recall details

Recall number
Z-0596-2025
Status
Open, Classified
Initiated by the company
Oct 31, 2024
Posted by the FDA
Dec 3, 2024
Recalling firm
Becton Dickinson & Co.
Quantity
584 units (313 US, 271 OUS)
Product code
MDB
FDA submissions
K915796
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.

Product description

BACTEC 9240 System Version or Model: 445475 Catalog Number: 445475

Code information (models, lots, serial numbers)

Show the full code information (5k characters)

Catalog No: 445475; UDI-DI: 00382904454753; Serial Numbers: DB5631, DB 6970, DB6857, DB6859, DB6122, DB3043, MD1710, DB4372, DB6043, DB2249, DB5030, DB2882, DB3090, MD2046, DB4927, DB5825, DB2828, DB2835, DB3251, DB3245, DB6893, DB3145, DB3211, DB5876, DB5906, DB4284, DB5948, DB3143, DB3204, DB6891, PM4385R, DB5761, DB4811, DB4718, DB3231, DB2786, DB4793, DB2322, DB6522, DB4820, DB2216, DB6558, DB4113, DB5168, DB2544, DB6674, DB4263, DB4842, DB6369, DB6900, DB4828, DB4418, DB5204, DB5282, DB4286, DB5569, DB5626, FS1472, FS1473, FS1230, FS1234, DB2702, DB4367, DB6655, DB6920, DB4365, DB2219, FT1004, DB6156, DB6961, DB2421, DB5244, DB2530, DB2844, DB4890, DB5146, DB5443, DB6011, DB5940, DB6628, DB3318, DB2417, MD1748, DB6315, DB4134, FS1320, DB3307, DB4304, MD2027, DB4152, MD1851, FS1280, DB2152, DB6243, DB4680, FS1004, FS1094, FS1001, DB6570, DB6391, DB6393, DB3004, DB4706, DB4248, DB6392, DB4260, DB6573, DB6571, DB2840, DB6584, DB2974, DB6634, DB4267, DB4440, DB2548, DB4449, DB6491, DB6903, DB4892, DB5529, DB2925, DB6595, DB5538, DB4271, DB6936, DB4987, DB2337, DB3389, DB3340, DF1540, DB2332, DB7013, DB4492, FS1178, FS1519, DB4565, DB4589, DB4028, DB6430, DB3076, DB6258, MD1730, MD1810, DB4586, DB5340, DB6499, DB5347A, DB5398, DB6603, MD2086, DB6113, DB2884, DB4733, DB4739, DB2804, DB4631, DB2276, DB4506, DB3024, DB6123, DB6590, DB4557, DB6281, 94707, DB6426, DB6657, DB6933, DB6658, DB4936, DB6324, DB2561, DB5865, DB5163, DB5005, DB6745, DB5200, DB5878, DB6709, DB4869, DB6058, DB6646, DB6645, DB6843, DB5094, DB4917, DB6845, DB6863, DB7074, DB7003, DB7071, DB7004, DB6994, DB7041, DB7133, DB7132, DB7167, DB6755, DB7187, DB7144, DB7142, DB6451, DB6689, DB7199, DB7182, DB7203, DB7136, DB5880, DB6896, DB6464, DB6679, FS1426, MD1634, DB1679, DB7050, DB6342, DB6872, DB2657, DB6576, DB6719, DB6938, DB4377, DB7000, DB6733, FS1400, DB2556, MD1750, DB6237, DB5602, DB7007, DB7008, DB6883, DF1563, DB6950, DB6559, DB6951, DF1562, DB6918, DB5858, DB5350, DB6041, DB4998, DB6484, DB6505, DB5363, DB6792, DB7212, DB5381, DB5422, DB5682, DB5653, DB5429, DB5784, FS1494, DB2516, DB4662, DB4665, MD1793, MD1794, DB4763, DB6809, MD1916, DB5905, MD1610, MD1840, DB2160, DB7178, DB6302, DB6165, MD1925, MD2056, MD2122, DB5596, DB5919, DB5861, DB5862, DB5813, DB5937, DB5089, DB6034, DB2403, MD1619, DB7157, DB2453, DB2424, DB5215, DB1059, DB2238, DB4020, DB5740, DB5065, DB6858, DB5955, DB2244, DB5021, DB3086, DB4776, 97R4YC2, DB6296, DB5328, DB2245, DB2483, DB5329, DB2946, DB2945, DB5491, DB4348, DB5319, DB4299, DB4484, MD2033, DB3062, DB3279, DB3349, DB4149, DB 5844, DB 5845, DB5847, DB 5997, DB 5846, DB5888, DB5683, DB4805, DB6802, DF1547, DB6218, DB6685, RC1164, DB2212, DB3212, DB4868, DB4891, DB4478, DB4742, DB5096, DB4855, DB6262, DB5668, DB2384, DB6024, DB6071, DB7101, DB7098, DB7100, DB6018, DB6036, DB2939, DB2967, DB5286, DB2146, FS1073, DB7120, DB5522, DB2977, DB6096, DB6483, DB6604, DB6640, DB6750, DB6670, DB6814, DB6815, DB7099, DB6990, DB7122, DB7110, DB6736, DB7139, DB7140, DB7190, DB7180, DB7191, DB7194, DB7216, DB7204, DB7205, DB7197, DB7173, DB2312, DB6597, DB5077, DB3291, DB4538, DB6155, DB6485, DB6486, DB6487, DB6488, DB7062, DB7185, DB4344, DB4453, DB4952, MD1771, MD1877, DB4138, DB6142, DB6477, MD1847, FS1470, FS1476, DB5084, FS1244, MD2047, DB5302, DB7030, DB2170, DB4816, DB5620, DB5169, DB5507, DB5737, DB5202, DB6811, DB7034, DB2220, DB5525, DB2221, DB6715, DB6708, DB6734, DB6781, DB2588, DB6780, DB7054, DB6268, DB2204, DB5696, DB6403, DB2737, DB5643, DB6717, DB2660, DB6591, DB6911, DB2633, DB5743, DB6852, DB2357, DB2758, DF1552, DB6790, TBD1234, DB6786, DB6323, DB6996, DB6941, DB6850, DB6907, DB6980, DB6952, DB5449, DB5375, DB5373, DB7076, DB4234, DB5608, DB5407, DB5652, DB5430, DB5512, DB4550, DB5594, DB7097, DB6213, DB5550, MD1832, DB2493, DB5083, MD1917, DB5734, DB7175, DB5911, DB5756, DB7176, DB7179, DB5800, DB5914, MD1611, DB2705, DB2377, MD1615, DB2378, DB6032, DB6054, DB6861, DB6494, DB4055, DB2845, DB6357, DB4083, DB4225, DB6121, DB4635, DB4078, DB6383, DB6671, DB6381, DB2819, DB3154, FS1011, DB2966, FS1193, DB6532, DB2818, DB6291, DB6283, MD1791, DB6617, DB6457, DB3124, DB4835, DB6169, DB4682, DB4596, DB6212, DB4649, DB4366, DB7096, DB6929, DB6978, DB6913, DB6319, FS1485, FS1341, DB2178, DB2619, FS1270, DB6240, MD1690, DB6317, DB6239, DB6300, DB4872, DB4441, DB2998, DB6581, DB6540, DB4451, DB6635, DB6699, DB2335, FS1461, DB6160, FS1204, DB5325, DB6161, FS1520, DB4206, DB4024, DB4765, FS1419, DF1596, MD1624, DB4619, DB6242, DB6496, DB4321, DB6331, DB6241, DB6667, DB4759, DB4322, DB4326, DB4752, DB4338, DB6339, DB3054, DB6989, DB4594, DB4938, DB4939, DB7022, DB7023, DB5177, DB5232, DB5205, DB5242, DB5274, DB7040, DB6905, DB2163, DB6065, DB6993, DB2620.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K915796 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95636).