Skip to content

Offset Cup Impactor - 3/8-16 interface. Reusable surgical… recallZ-0549-2025 · Incipio Devices · initiated Oct 30, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Incipio Devices or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Recalled lot does not meet specification and has reported incidents of fracture of the TIG weld.

FDA root cause category: Process change control

What the company is doing

Consignees were mailed a recall notification titled URGENT ADVISORY NOTICE dated 10/30/2024. The notification instructs consignees to examine their inventory and identify all affected devices for quarantine immediately. Consignees are to complete and return the Recall Response Form to Incipio Devices at [email protected] to facilitate the return of recalled devices and issuance of a FedEx return label. Returned devices should be sent back in their original packaging if unused and properly decontaminated and sterilized if being sent back used. Consignees with any questions should contact Incipio Devices by phone at +41 32 754 34 34 or by email at [email protected].

Recall details

Recall number
Z-0549-2025
Status
Open, Classified
Initiated by the company
Oct 30, 2024
Posted by the FDA
Nov 25, 2024
Recalling firm
Incipio Devices
Quantity
100 units
Product code
HWA
Distribution
US Nationwide distribution in the state of TN.

Product description

Offset Cup Impactor - 3/8-16 interface. Reusable surgical instrument used to position and place acetabular sup implants.

Code information (models, lots, serial numbers)

Product Reference No.: 50341105; UDI-DI: 07630055304735; Lot No.: 24A0087.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page