Skip to content

Offset Cup Impactor - 3/8-16 interface 50341105Recalls & reported problems

Incipio Devices SA · Offset Cup Impactor for hip arthroplasty · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630055304735), received between Dec 12, 2024 and Dec 12, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Crack” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 30, 2024Open, Classified

    Recalled lot does not meet specification and has reported incidents of fracture of the TIG weld.

    Z-0549-2025 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Crack3
  • Manufacturing, Packaging or Shipping Problem3
  • Nonstandard Device3
  • Solder Joint Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2024Crack; Nonstandard Device; Solder Joint Fracture; Manufacturing, Packaging or Shipping Problem; Crack; Nonstandard Device; Solder Joint Fracture; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING A THR SURGERY, ONE (1) INCIPIO R3 OFFSET CUP IMPACTOR CRACKED WHILE IMPACTING THE CUP. THE PROCEDURE RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20915083 ↗
  • MalfunctionDec 12, 2024Crack; Nonstandard Device; Solder Joint Fracture; Manufacturing, Packaging or Shipping Problem; Crack; Nonstandard Device; Solder Joint Fracture; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING A THR SURGERY, UPON IMPACTION OF THE CUP, THE STRIKE PLATE OF ONE (1) INCIPIO OFFSET CUP IMPACTOR COMPLETELY BROKE OFF AND FELL ON THE FLOOR. THE PROCEDURE RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20914633 ↗
  • MalfunctionDec 12, 2024Crack; Nonstandard Device; Solder Joint Fracture; Manufacturing, Packaging or Shipping Problem; Crack; Nonstandard Device; Solder Joint Fracture; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING A THR, THE STRIKE PLATE OF A INCIPIO OFFSET CUP IMPACTOR BROKE OFF WHILE IMPACTING THE CUP. IT IS UNKNOWN IF PIECES FELL INTO PATIENT. THERE WAS A NON-SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20914150 ↗

Other Impactor devices

All Impactor devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.