BIORAPTOR recallZ-0581-2025 · Smith & Nephew Inc. · initiated Oct 29, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
FDA root cause category: Package design/selection
What the company is doing
An URGENT MEDICAL DEVICE RECALL NOTICE dated 10/29/24 was mailed to consignees. This notification instructed consignees to inspect their inventory for affected devices and quarantine them immediately. If product was further distributed consignees are to notify their customers of this recall and ensure recall actions are completed. Consignees are to complete the provided response form to facilitate the return of devices and email the completed form to [email protected]. Consignees who do not have product on hand are to complete and return the recall response form as well. Any questions about this recall are to be emailed to [email protected].
Recall details
- Recall number
- Z-0581-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 29, 2024
- Posted by the FDA
- Nov 29, 2024
- Recalling firm
- Smith & Nephew Inc.
- Quantity
- 2,156 units
- Product code
- MBI
- FDA submissions
- K071586, K152566
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, AL, IL, IN, KY, MD, MN, NC, NE, NH, NJ, TX and the countries of CA, CL, CN, EC, ES, GB, KZ, NO, PE, PL, RU, TR.
Product description
BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.
Code information (models, lots, serial numbers)
Part No. 72201541, 72201542, 72203280, 72203281; UDI-DI: 03596010595072, 03596010595089, 03596010656452, 03596010656469; Batch No. 2146369, 2151242, 2151690, 2154691, 2153489, 2151692.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- BIORAPTOR 72201541Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall41 reports
- BIORAPTOR 72203281Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall13 reports
- BIORAPTOR 72201542Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall3 reports
- BIORAPTOR 72203280Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall3 reports
Product lines
Other devices under the same FDA submission
10 devices cleared under K071586, K152566 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HEALICOIL 72203378Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE150 reports
- TWINFIX 72200755Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE117 reports
- HEALICOIL 72203380Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE75 reports
- TWINFIX 72200750Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE54 reports
- TWINFIX 72200752Smith & Nephew, Inc. · STAPLE, FIXATION, BONE45 reports
- RAPTORMITE / ULTRABRAID 72201806Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE22 reports
- BIORAPTOR 72201543Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE2 reports
- TWINFIX 72200758Smith & Nephew, Inc. · STAPLE, FIXATION, BONE1 report
- TWINFIX 72200762Smith & Nephew, Inc. · STAPLE, FIXATION, BONE1 report
- TWINFIX 72200796Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page